Basic concepts of medical imaging
Since 1895, medical imaging has allowed physicians to examine the inside of the body without surgery. A medical image is not a photograph. It records the spatial distribution of a physical quantity: how strongly X-rays were absorbed, how strong a signal hydrogen nuclei emitted after a radio pulse, where sound was reflected, or where a tracer accumulated. Interpreting such images requires a small set of concepts — pixels and gray levels; contrast, noise and resolution; and the imaging chain that runs from an energy source to the reader. Each later chapter builds on them.

0.1What a medical image measures1895
An ordinary photograph records how much visible light is reflected from the surfaces of objects. Medical imaging is concerned with structures beneath the skin, and visible light is scattered so strongly in tissue that it cannot form an image of anything deep. Each imaging method therefore uses a different physical signal — one that crosses the body and can still be measured outside it — and arranges the measurements by position. The brightness of each point represents the value of a physical quantity at that location.
For this reason, "white" has a different meaning in each type of image. On a radiograph or a CT scan, white means the X-rays were strongly attenuated there, as they are by bone. On MRI, white means the hydrogen nuclei at that location gave a strong signal at the moment of measurement — it may be fat, it may be fluid. On ultrasound imaging, white indicates a strong echo. On PET, the bright regions are those where an injected radiotracer has accumulated.
| Method | What is measured | Bright usually means | Chapter |
|---|---|---|---|
| Radiography and CT | Attenuation of X-rays crossing tissue | Strong absorption: bone, metal, contrast agent | 1 and 2 |
| MRI | Radio signal from hydrogen nuclei in a magnet, shaped by T1 and T2 relaxation | Depends on the sequence: fat on T1, fluid on T2 | 3 |
| Ultrasound | Echoes of sound from tissue boundaries | Strong reflectors: stones, bone surfaces | 4 |
| Nuclear medicine and PET | Concentration of a radioactive tracer | High uptake: active tumors, the brain | 5 |
Interpreting any medical image therefore begins with identifying the quantity it measures. Two tissues with almost the same X-ray attenuation look alike on CT yet may differ markedly on MRI. A fluid collection that is obvious on ultrasound may not be visible on a radiograph. Physicians often order more than one kind of examination for the same clinical question for exactly this reason: each method measures one physical quantity. The first half of this exhibit describes how each measurement principle was developed and how its measurements are converted into an image.
0.2Pixels, voxels and gray levels
When any digital medical image is magnified far enough, it resolves into a grid of small squares: pixels. Each pixel stores a single number — on CT, the average attenuation of a small piece of tissue; on MRI, a signal strength. CT and MRI actually acquire three-dimensional data, so each number stands for a small block of tissue with thickness, a voxel. A CT slice is usually 512 × 512 pixels1; inside the patient each voxel is roughly 0.5–1 mm across, and slices can be thinner than 1 mm1. Smaller voxels resolve finer detail, but each one collects less signal.
The number of binary digits used to store each value is the bit depth. Eight bits give 256 levels; twelve give 4,096. CT uses at least 12 bits, so that values from −1,024 to +3,071 cover air, soft tissue and bone1, and they are normally stored in 16-bit words1. An ordinary display, however, has 8 bits per channel: 256 levels of gray. Human perception has its own limits — the same step in brightness looks different in dark and bright regions — so medical monitors are calibrated to the DICOM Grayscale Standard Display Function, which makes each gray step perceptually equal1. Mapping 4,096 values onto 256 gray levels requires a selection; chapter 2 describes how windowing makes it.
- 512 × 512
- typical pixel matrix of one CT slice1
- 4096
- levels in 12 bits, the minimum for CT1
- 256
- gray levels on an 8-bit display1
Further detailHow big is a CT scan?
A 512 × 512 CT slice at 2 bytes per pixel is about 0.5 MB. Scanning 30 cm of chest and abdomen in 1 mm slices yields 300 slices, roughly 150 MB — and twice that if the same raw data are reconstructed once with a bone kernel and once with a soft-tissue kernel. These data are also quantitative: every pixel is a measurement that can be compared, quantified and analysed by computer. This is the precondition for the later development of radiomics and deep learning.
0.3Contrast, noise and resolution
Image quality is described chiefly by three quantities. Contrast is the difference in signal between a target and its background, and it is determined first by the physics of the method. On CT, gray matter is near +35 HU and white matter near +25 HU — a difference of about 1% of the attenuation of water1. On MRI the same two tissues differ in T1 by more than a hundred milliseconds, and the contrast is far stronger1.
Noise is the random fluctuation superimposed on the signal. X-ray photons reach a detector at random: if N arrive on average, the count fluctuates by about √N. A hundred photons give 10% noise; ten thousand give 1%. To be detected reliably, a target must differ from its background by roughly five times the noise — the Rose criterion1. Contrast below this level cannot be distinguished from noise.
Signal-to-noise ratio for photon counting. N is the mean number of photons reaching one detector element; √N is its random fluctuation. Halving the noise takes four times the photons — for X-rays and CT, four times the dose.
Spatial resolution is the smallest detail that can still be separated. Pixel size limits it, but blur usually limits it first: the X-ray focal spot has a finite size, detectors spread signal sideways, and patients breathe. The image of an ideal point source is called the point spread function. Mammography must resolve structures as small as 50 μm1, which places strict requirements on every link of the system.
The three quantities constrain one another. Smaller voxels each receive fewer photons, so noise rises; keeping noise down then costs acquisition time or dose. Almost every advance in imaging can be described as a new balance between these three quantities, dose and time.
0.4The imaging chain
Whatever the modality, the formation of a medical image follows the same chain. It starts with a source of energy: an X-ray tube, a radio-frequency coil, an ultrasound probe, or a tracer injected into the patient. The energy interacts with the body — it is absorbed, scattered, reflected, or emitted from within. A detector converts whatever emerges into electrical signals. Reconstruction converts the signals into an image: a radiograph barely needs this step, but CT and MRI must solve an inverse problem, computing the internal distribution from indirect measurements. Display maps the numbers to gray levels. Finally the image is interpreted by a radiologist and, increasingly, by algorithms as well.
The chain obeys one basic rule: no later link can recover information that an earlier one lost. Noise from too few photons can be traded off against other properties by a reconstruction algorithm but not removed; an edge blurred by motion only becomes a more pronounced artifact if the display sharpens it. At the same time, every link can be improved. The period from 1986 to 2026, covered in the second half of this exhibit, is to a large degree a history of computers being introduced into successive links of this chain — first storage and transmission, then reconstruction and denoising, and finally image interpretation itself.
0.5Four families of imaging methods
The body is opaque to most of the electromagnetic spectrum. Visible and infrared light are strongly scattered and absorbed in tissue; ultraviolet barely penetrates the skin. Only two bands, at the two ends of the spectrum, can be used for imaging. At the high-energy end are X-rays and gamma rays: diagnostic X-ray photons carry less than 150 keV1, the gamma photons detected in PET carry 511 keV, and both partly pass through the body — the fraction transmitted carries the information. At the low-energy end are radio waves: MRI at 1.5 T operates near 64 MHz2, and the body is almost transparent to it.
The third approach uses sound rather than electromagnetic waves. Diagnostic ultrasound uses 2–15 MHz and travels through soft tissue at about 1,540 m/s1, giving wavelengths of roughly 0.1–0.8 mm. The fourth approach places the radiation source inside the body: a radioactive tracer is injected, and the gamma photons it emits are detected outside.
| Imaging method | Kind of wave | Typical frequency or energy | Ionizing? |
|---|---|---|---|
| X-rays (radiography, CT) | Electromagnetic | About 20–150 keV | Yes |
| Radio frequency (MRI) | Electromagnetic | About 64 MHz at 1.5 T | No |
| Ultrasound | Mechanical (sound) | 2–15 MHz | No |
| Gamma rays (nuclear medicine, PET) | Electromagnetic, emitted from inside | 511 keV for PET | Yes |
The type of wave determines what can be measured and what risks are involved. X-ray and gamma photons carry tens of thousands of times the energy of a chemical bond, enough to ionize atoms, which is why they involve a radiation dose; radio waves and ultrasound do not ionize. Nor does the wavelength fix the resolution. The radio waves used in MRI are 4.7 m long in vacuum, yet MRI resolves structures smaller than a millimeter, because it localizes its signal not by focusing a beam but with magnetic field gradients, as described in chapter 3.
0.6How to use this exhibit2026
The exhibit has two parts. Chapters 0 to 6 cover the fundamentals: X-rays, CT, MRI, ultrasound and nuclear medicine each have a chapter that begins with the invention of the technique and explains what is measured, how the image is formed and what its limitations and risks are; chapter 6 then summarizes the digital-image principles they all share. Chapters 7 to 13 cover 1986 to 2026: how imaging became digital, how computer methods for image analysis developed, and how artificial intelligence moved from research papers into clinical use and on to foundation models.
Keywords in the text open a short explanation; where a concept has its own section elsewhere, the pop-up offers a link to it. Living figures are interactive — they can be dragged and clicked — and their captions state what to try and what to observe. Boxes headed "Going deeper" contain formulas and more detailed physics; they can be skipped without loss of continuity. Every date and number carries a source, shown by clicking the superscript.
- A medical image represents the spatial distribution of a physical quantity — attenuation, proton signal, echo or tracer concentration — and what "white" means depends on which.
- Each pixel or voxel stores one number; bit depth sets how many levels it can take, and an 8-bit display shows only 256 grays, so a selection is required.
- Contrast, noise and resolution constrain one another and are each linked to dose and time; for photon counting, SNR is √N.
- Every modality follows the same imaging chain, and no later link can recover what an earlier one lost.
- The body is transparent to only a few kinds of wave — X-rays and gamma rays, radio waves, and ultrasound.
References · 2
- Dance DR, Christofides S, Maidment ADA, McLean ID, Ng KH, eds. Diagnostic Radiology Physics: A Handbook for Teachers and Students. Vienna: International Atomic Energy Agency; 2014. (Chapters 4, 9, 11, 12, 14, 16 and 18.) www-pub.iaea.org/MTCD/Publications/PDF/Pub1564webNew-7466642
- National Institute of Standards and Technology. CODATA internationally recommended 2022 values of the fundamental physical constants: proton gyromagnetic ratio in MHz/T (42.577 478 461 MHz/T). physics.nist.gov/cgi-bin/cuu/Value?gammapbar