Physics · 1909

bremsstrahlung

轫致辐射

X-rays emitted when a fast electron is sharply decelerated by the electric field of a nucleus; the German word means "braking radiation". It produces most of the X-rays in an X-ray tube, as a continuous spectrum up to the tube voltage.

An electron passing close to a tungsten nucleus is deflected and slowed, and the lost kinetic energy leaves as a photon. Near misses and close passes give every energy from zero up to a ceiling set by the tube voltage: at 100 kV, no photon exceeds 100 keV (the Duane–Hunt limit). Sharp lines from tungsten's characteristic radiation, near 59 and 67 keV, sit on top of this continuum.

Arnold Sommerfeld introduced the German term in 1909 to explain the continuous part of the X-ray spectrum.

The process is inefficient: at 100 kV only about 1% of the electrons' energy becomes X-rays, a yield roughly proportional to target atomic number and voltage, which is why targets are tungsten (Z = 74). The lowest-energy photons would be absorbed in the skin, adding dose without adding image, so tubes carry aluminum filters that remove them and leave a harder beam.

Sources
  1. Bremsstrahlung. Wikipedia. en.wikipedia.org/wiki/Bremsstrahlung
  2. X-ray tube. Wikipedia. en.wikipedia.org/wiki/X-ray_tube

Key works · 1

Landmark1913Physical Review164 citations

A Powerful Röntgen Ray Tube with a Pure Electron Discharge

W. D. Coolidge

Coolidge described an X-ray tube in which a heated tungsten filament supplies electrons in a high vacuum. A separate high voltage accelerates the electrons toward the target. The design let the operator regulate tube current through filament heating and control the energy of the electrons through the accelerating voltage.

SignificanceThe hot cathode and high vacuum made X-ray generation more reproducible and gave operators separate control over tube current and voltage. The IAEA diagnostic radiology physics handbook describes thermionic emission as the operating principle of modern diagnostic X-ray tubes.