X-ray tube
X射线管
The vacuum tube that makes medical X-rays: electrons from a hot filament are accelerated through tens of kilovolts into a tungsten target, where about 1% of their energy becomes X-rays and the rest heat. Voltage sets photon energy; current × time sets how many.
Röntgen used a Crookes tube, which got its electrons by ionizing residual gas and drifted as the gas pressure changed. In 1913 William Coolidge introduced the hot-cathode tube: a heated tungsten filament in high vacuum emits electrons, so current and voltage can be set independently. Every modern tube follows his design.
Two settings matter most. Tube voltage (kVp) fixes the maximum photon energy and penetration: around 120 kVp for a chest radiograph, only 25–35 kVp for mammography. Tube current times exposure time (mAs) fixes the number of photons; doubling it doubles the dose and cuts quantum noise by about 30%.
About 99% of the electrical energy becomes heat, and the focal spot can reach 2500 °C during an exposure, so the anode is a disc spinning at high speed to spread the load around a track. A smaller focal spot gives sharper images but tolerates less heat, the central trade-off of tube design; long helical CT acquisitions make heat capacity a key specification.
- Coolidge WD. A powerful röntgen ray tube with a pure electron discharge. Phys Rev. 1913;2(6):409–430. doi.org/10.1103/PhysRev.2.409
- X-ray tube. Wikipedia. en.wikipedia.org/wiki/X-ray_tube
Key works · 1
Landmark1913Physical Review164 citations
A Powerful Röntgen Ray Tube with a Pure Electron Discharge
Coolidge described an X-ray tube in which a heated tungsten filament supplies electrons in a high vacuum. A separate high voltage accelerates the electrons toward the target. The design let the operator regulate tube current through filament heating and control the energy of the electrons through the accelerating voltage.
SignificanceThe hot cathode and high vacuum made X-ray generation more reproducible and gave operators separate control over tube current and voltage. The IAEA diagnostic radiology physics handbook describes thermionic emission as the operating principle of modern diagnostic X-ray tubes.