radiation dose
辐射剂量
How much energy ionizing radiation deposits in the body, and what that means for health. Absorbed dose is measured in grays; weighting by organ sensitivity gives the effective dose in sieverts, the common yardstick for comparing examinations.
Dose comes in layers. Absorbed dose is energy per kilogram of tissue, 1 Gy = 1 J/kg. Multiplying by a radiation weighting factor (1 for X- and gamma rays, 20 for alpha particles) gives equivalent dose, and summing organs weighted by their cancer sensitivity gives effective dose. The ICRP introduced the idea in 1977 as "effective dose equivalent" and renamed it effective dose in its 1990 recommendations.
Typical effective doses: a chest X-ray about 0.1 mSv, a head CT 1.6 mSv, a chest CT about 6 mSv, a low-dose CT for lung screening 1.5 mSv. The average American receives about 3 mSv a year from natural background.
Scanners report dose indices such as CTDIvol and DLP after every CT examination, so exposure can be recorded and audited. Lowering dose raises noise, the problem that iterative reconstruction and deep-learning reconstruction were built to solve.
- RadiologyInfo.org (RSNA/ACR). Radiation dose. www.radiologyinfo.org/en/info/safety-xray
- ICRP. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP. 2007;37(2–4). www.nrc.gov/docs/ml1233/ML12338A682.pdf
Key works · 9
Landmark1981Medical Physics488 citations
A method for describing the doses delivered by transmission x-ray computed tomography
Shope and colleagues proposed deriving a dose descriptor for a series of contiguous CT sections from a single-scan dose profile. They integrated the profile along the scan axis and tested the estimate at different phantom positions and for different numbers of scans using five types of scanner.
SignificanceThe work introduced the CT dose index as a reproducible descriptor of scanner output. The FDA subsequently adopted the quantity, and modern ICRP guidance uses its volumetric form in setting and assessing diagnostic reference levels for CT.
DOI Thomas B. ShopeGordon C. Johnson radiation dosecomputed tomographyCT dose index
Landmark2007Annals of the ICRP
The 2007 Recommendations of the International Commission on Radiological Protection
The Commission revised the radiation and tissue weights used for equivalent and effective dose and defined planned, emergency, and existing exposure situations. It applied justification and optimization across all three situations and specified individual dose limits for occupational and public exposure in planned situations.
SignificanceThe recommendations supplied the framework used in the IAEA International Basic Safety Standards. Those standards apply the principles of justification and optimization to medical exposures and assign responsibilities for implementing them in diagnostic and interventional imaging.
DOI Jack Valentin radiation dosesievertALARAionizing radiation
Key1972Radiology70 citations
X-Ray Exposure Reduction Using Rare-Earth Oxysulfide Intensifying Screens
This study compared terbium-activated gadolinium and lanthanum oxysulfide screens with calcium tungstate. The rare-earth phosphors absorbed more incident radiation over part of the diagnostic energy range and converted it to light more efficiently. A gadolinium oxysulfide screen reduced exposure at comparable resolution even with blue-sensitive film.
SignificanceMatching rare-earth screens to green-sensitive film let radiographic systems use more of the light produced per exposure. UNSCEAR records faster rare-earth screens among the technical advances that reduced patient doses in conventional radiography during the screen-film era.
DOI Full text Robert A. BuchananKenneth A. Wickersheim radiographic filmscintillatorradiation dose
Key1995NIST2,156 citations
Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest
The NIST report compiled photon mass attenuation and mass energy-absorption coefficients over energies from 1 keV to 20 MeV. The tables and graphs cover elements with atomic numbers 1 through 92 and 48 additional substances and mixtures relevant to dosimetry.
SignificanceThe compilation extended Hubbell's earlier reference tables and was subsequently made available as NIST Standard Reference Database 126. It provides consistent material data for comparing tissue attenuation, calculating photon penetration in shielding, and estimating energy deposition in dosimetry.
DOI Full text John H. HubbellStephen M. Seltzer attenuation coefficientattenuationradiation dose
Key2012The Lancet4,240 citations
Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study
The authors followed a retrospective cohort of children and young adults examined by CT in Great Britain. They estimated cumulative absorbed doses to red bone marrow and the brain and used cancer registry data to assess subsequent disease. Estimated doses were positively associated with leukemia and brain tumor incidence.
SignificanceThe study supplied direct epidemiologic evidence linking reconstructed organ doses from childhood CT with later cancer incidence. The National Cancer Institute cites the results in its discussion of pediatric CT radiation, including the need to optimize exposure.
DOI Full text Mark S. PearceAmy Berrington de González radiation doseionizing radiationlow-dose CT
Key2014The New England Journal of Medicine748 citations
Ultrasonography versus Computed Tomography for Suspected Nephrolithiasis
This multicenter trial randomly assigned 2,759 patients with suspected kidney stones to point-of-care ultrasound, radiology ultrasound, or CT as the initial study. The ultrasound groups had lower cumulative radiation exposure over six months, with no significant difference in high-risk diagnoses with complications within 30 days.
SignificanceThe trial compared initial imaging strategies using patient outcomes and cumulative radiation exposure, while allowing additional imaging at the treating physician's discretion. It provided comparative evidence for starting with ultrasound in selected emergency patients with suspected kidney stones.
DOI Rebecca Smith-BindmanSteven R. Cummings point-of-care ultrasoundultrasound imagingcomputed tomographyradiation dose
Key2017Annals of the ICRP1,338 citations
ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging
The Commission clarified which quantities to use for diagnostic reference levels, how to survey practices, and when to update benchmark values. It extended practical guidance to interventional procedures, pediatric imaging, and newer technologies and discussed how automated dose reporting could support periodic comparisons.
SignificanceThe report turns optimization into a process that can be audited across facilities using comparable examinations. The IAEA uses the diagnostic reference level approach in its implementation guidance to identify protocols whose higher typical doses warrant investigation.
DOI Eliseo VañóA. Rogers radiation doseALARAdiagnostic reference levelCT dose index
Key2023Annals of the ICRP59 citations
Optimisation of Radiological Protection in Digital Radiology Techniques for Medical Imaging
The Commission describes how to optimize protection in digital radiography, fluoroscopy, and CT while preserving image quality adequate for the clinical task. It proposes four levels for facilities to assess their practices, covering equipment selection and monitoring, protocol review, staff training, and collaboration across professions.
SignificancePublication 154 gives digital imaging departments a way to assess their implementation of the Commission's optimization principle in stages. The task group's companion strategy describes how departments can use professional roles, technical capability, and management processes to identify their next steps.
DOI Full text Colin J. MartinJenia Vassileva radiation doseALARAdigital radiographyfluoroscopycomputed tomographydiagnostic reference level
Key2023Nature Medicine412 citations
Risk of hematological malignancies from CT radiation exposure in children, adolescents and young adults
EPI-CT followed 948,174 people in nine European countries who underwent CT before age 22. The researchers reconstructed active bone marrow dose from examination and scanner information. Cumulative dose was associated with hematologic malignancies, with an excess relative risk of 1.96 per 100 mGy and a 95% confidence interval of 1.10 to 3.12.
SignificanceThe study provided a multinational reference estimate for hematologic malignancy following CT in young people by applying a common organ-dose method. The subsequent North American RIC cohort study explicitly compared the two cohorts' risk estimates at 100 mGy.
DOI Full text Magda Bosch de BaseaElisabeth Cardis computed tomographyradiation doseionizing radiation