digital radiographyDR
数字X线摄影
Radiography captured directly by an electronic detector, usually a flat-panel detector, with the image on screen within seconds and no film or plate to process. It is now the usual way chest and bone X-rays are taken, and the source of the digital archives imaging AI is trained on.
Its core component is a flat-panel detector. Indirect panels use a caesium iodide scintillator to turn X-rays into light and amorphous-silicon photodiodes to turn the light into charge; direct panels convert X-rays straight to charge in a layer of amorphous selenium. A thin-film transistor behind each pixel lets the charge be read out row by row as a digital image.
Amorphous-silicon and amorphous-selenium panels appeared in the mid-1990s and reached clinics by the end of the decade; earlier systems photographed a phosphor screen with a CCD camera. Compared with computed radiography, DR removes the walk to the plate reader, speeds up every exam, and with an efficient panel can lower the dose.
Images go straight into picture archiving and communication system as DICOM files, ready for windowing, measurement and remote reading. One new hazard: unlike film, an overexposed digital image does not turn black, so doses can creep upward unnoticed unless an exposure index is monitored.
- Zhao W, Rowlands JA. X-ray imaging using amorphous selenium: feasibility of a flat panel self-scanned detector for digital radiology. Med Phys. 1995;22(10):1595–1604. doi.org/10.1118/1.597628
- Rowlands JA, et al. Flat-panel digital radiology with amorphous selenium and active-matrix readout. RadioGraphics. 1997;17(3):753–760. doi.org/10.1148/radiographics.17.3.9153709
Key works · 6
Landmark1983Radiology978 citations
Computed radiography utilizing scanning laser stimulated luminescence.
The Fuji system recorded a latent X-ray image on a storage phosphor plate. A laser subsequently scanned the plate and released light in proportion to the stored signal, which the reader converted to digital values. The plate could then be erased and reused.
SignificanceStorage plates made it possible to digitize projection radiographs while retaining cassette-based acquisition in existing rooms. Digital processing and archiving could therefore be introduced before flat-panel detectors were widely available. The IAEA handbook describes this technology as computed radiography.
DOI Mitumasa SonodaH. Kato computed radiographydigital radiographyradiographic film
Key1982Proceedings of SPIE 31820 citations
Filmless Picture Archiving And Communication In Diagnostic Radiology
The authors examined changes in digital acquisition and proposed a sequence of archiving and communication strategies from a research or teaching setting to broad clinical service. Their assessment compared levels of online storage and concluded that a system with partial online digital storage was feasible.
SignificancePresented at the first dedicated PACS meeting, the paper made filmless radiology concrete as a system of archiving and communication capabilities that could be introduced in stages. Later hospital-scale evaluations examined how these capabilities affected clinical access to images.
DOI André J. DuerinckxE. J. Pisa picture archiving and communication systemdigital radiographyteleradiology
Key1995Medical Physics355 citations
X-ray imaging using amorphous selenium: Feasibility of a flat panel self-scanned detector for digital radiology
Zhao and Rowlands assessed a flat-panel concept that used amorphous selenium to convert X-rays directly to electrical charge and a thin-film transistor matrix to read out the resulting image. The analysis indicated feasibility for projection radiography, with electronic noise still limiting performance at typical fluoroscopic exposure rates.
SignificanceThe analysis set design requirements for a direct-conversion radiographic panel. The same team subsequently reported images of a hand phantom obtained with an operating prototype. Later clinical reports document use of amorphous-selenium flat panels for routine projection radiography.
DOI Wei ZhaoJohn A. Rowlands flat-panel detectordigital radiographyfluoroscopy
Key2005New England Journal of Medicine2,187 citations
Diagnostic Performance of Digital versus Film Mammography for Breast-Cancer Screening
In DMIST, 49,528 asymptomatic women underwent both digital and film screening mammography. Readers interpreted the images independently, and complete data were available for 42,760 women. Receiver operating characteristic analysis showed similar overall accuracy, with higher accuracy for digital mammography among women under 50, women with dense breasts, and premenopausal or perimenopausal women.
SignificanceThe paired design established a direct comparison during the transition from film to digital screening. The National Cancer Institute uses DMIST when summarizing the evidence on digital mammography, including the observed differences by age and breast density.
DOI Etta D. PisanoMurray Rebner mammographydigital radiographyROC curve and AUC
Key2023Annals of the ICRP59 citations
Optimisation of Radiological Protection in Digital Radiology Techniques for Medical Imaging
The Commission describes how to optimize protection in digital radiography, fluoroscopy, and CT while preserving image quality adequate for the clinical task. It proposes four levels for facilities to assess their practices, covering equipment selection and monitoring, protocol review, staff training, and collaboration across professions.
SignificancePublication 154 gives digital imaging departments a way to assess their implementation of the Commission's optimization principle in stages. The task group's companion strategy describes how departments can use professional roles, technical capability, and management processes to identify their next steps.
DOI Full text Colin J. MartinJenia Vassileva radiation doseALARAdigital radiographyfluoroscopycomputed tomographydiagnostic reference level
Notable1999Radiology212 citations
Principles of Digital Radiography with Large-Area, Electronically Readable Detectors: A Review of the Basics
The review compared large-area electronic X-ray detectors. It explained indirect conversion using a scintillator and light-sensitive elements, direct conversion using charge generated in a photoconductor, and how the detector array reads out the resulting signal. It also discussed implications for image quality.
SignificanceThe article documents the transition from systems requiring separate plate readout to electronically readable area detectors. Its distinction between direct and indirect conversion remains a framework used in medical physics teaching to explain how digital radiographic detectors collect signals.
DOI Harrell G. Chotas digital radiographyflat-panel detectorcomputed radiography