flat-panel detectorFPD

平板探测器

A thin detector the size of a film cassette, with millions of pixels on a thin-film-transistor array, that reads X-rays out directly as a digital image. It replaced film and image intensifiers in digital radiography, mammography, fluoroscopy and cone-beam CT.

The core is a thin-film-transistor array built on a large glass plate, the same technology used in LCD screens. Indirect panels add a caesium iodide scintillator that turns X-rays into light, which a photodiode at each pixel converts to charge; direct panels use a few hundred micrometres of amorphous selenium in which X-rays free charge directly. Pixels are typically 0.1–0.2 mm across for general radiography.

Through the 1990s John Rowlands's group in Toronto developed amorphous-selenium panels while several companies built amorphous-silicon ones. GE put an amorphous-silicon panel into general radiography in 1998, and in January 2000 into the first full-field digital mammography system approved by the FDA.

Panels read out fast enough for tens of frames per second, so they also displaced the image intensifier in fluoroscopy and let a C-arm spin once around the patient to perform cone-beam CT. They are expensive and fragile, and defective pixels and gain differences need regular calibration.

Sources
  1. Zhao W, Rowlands JA. X-ray imaging using amorphous selenium: feasibility of a flat panel self-scanned detector for digital radiology. Med Phys. 1995;22(10):1595–1604. doi.org/10.1118/1.597628
  2. AuntMinnie. FDA approves GE's full-field digital mammography system. 30 Jan 2000. www.auntminnie.com/clinical-news/womens-imaging/breast/article/1

Key works · 3

Key1995Medical Physics355 citations

X-ray imaging using amorphous selenium: Feasibility of a flat panel self-scanned detector for digital radiology

Wei Zhao, J. A. Rowlands

Zhao and Rowlands assessed a flat-panel concept that used amorphous selenium to convert X-rays directly to electrical charge and a thin-film transistor matrix to read out the resulting image. The analysis indicated feasibility for projection radiography, with electronic noise still limiting performance at typical fluoroscopic exposure rates.

SignificanceThe analysis set design requirements for a direct-conversion radiographic panel. The same team subsequently reported images of a hand phantom obtained with an operating prototype. Later clinical reports document use of amorphous-selenium flat panels for routine projection radiography.

Key1997Medical Physics120 citations

Empirical investigation of the signal performance of a high-resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation

L. E. Antonuk, Y. El-Mohri … R. A. Street

This study measured the behavior of a large-area active-matrix flat panel in radiographic and fluoroscopic modes. An intensifying screen converted X-rays to light, which a matrix of amorphous-silicon photodiodes and thin-film transistors read out. The authors characterized dark signals, temporal response, and sensitivity with different phosphor screens.

SignificanceThe measurements showed which signal and timing properties must be managed to use a single indirect panel for still radiographs and successive fluoroscopy frames. Scintillators coupled to amorphous-silicon readout arrays subsequently became a common flat-panel detector architecture.

Notable1999Radiology212 citations

Principles of Digital Radiography with Large-Area, Electronically Readable Detectors: A Review of the Basics

Harrell G. Chotas, James T. Dobbins III, Carl E. Ravin

The review compared large-area electronic X-ray detectors. It explained indirect conversion using a scintillator and light-sensitive elements, direct conversion using charge generated in a photoconductor, and how the detector array reads out the resulting signal. It also discussed implications for image quality.

SignificanceThe article documents the transition from systems requiring separate plate readout to electronically readable area detectors. Its distinction between direct and indirect conversion remains a framework used in medical physics teaching to explain how digital radiographic detectors collect signals.