computed radiographyCR
计算机X线摄影
Radiography on a reusable imaging plate instead of film. A storage phosphor holds the X-ray pattern as trapped electrons until a scanning laser releases it as light, point by point, into a digital image. It let hospitals go digital without replacing their X-ray rooms.
The plate is coated with europium-doped barium fluorohalide. X-rays lift electrons into traps in the crystal, where they stay as a latent image. In the reader a red laser scans the plate line by line; the trapped electrons fall back and emit blue-violet light, which a photomultiplier measures and the reader digitizes (quantization). Bright light then erases the plate for reuse, in principle thousands of times.
Researchers at Fuji Photo Film (Sonoda and colleagues) described the system in Radiology in 1983, the year Fuji brought it into clinical use. Its dynamic range is far wider than film's, so an imperfect exposure can be rescued in processing.
Because the plates fit standard film cassettes, CR slotted into existing X-ray rooms and was the transitional technology through which many hospitals went digital and onto picture archiving and communication system in the 1990s. Plates still have to be carried to a reader, and laser light scattering in the phosphor makes images a little less sharp than digital radiography.
- Sonoda M, Takano M, Miyahara J, Kato H. Computed radiography utilizing scanning laser stimulated luminescence. Radiology. 1983;148(3):833–838. doi.org/10.1148/radiology.148.3.6878707
- IAEA. Avoidance of unnecessary dose to patients while transitioning from analogue to digital radiology. IAEA-TECDOC-1667. 2011. www-pub.iaea.org/MTCD/Publications/PDF/te_1667_web.pdf
Key works · 2
Landmark1983Radiology978 citations
Computed radiography utilizing scanning laser stimulated luminescence.
The Fuji system recorded a latent X-ray image on a storage phosphor plate. A laser subsequently scanned the plate and released light in proportion to the stored signal, which the reader converted to digital values. The plate could then be erased and reused.
SignificanceStorage plates made it possible to digitize projection radiographs while retaining cassette-based acquisition in existing rooms. Digital processing and archiving could therefore be introduced before flat-panel detectors were widely available. The IAEA handbook describes this technology as computed radiography.
DOI Mitumasa SonodaH. Kato computed radiographydigital radiographyradiographic film
Notable1999Radiology212 citations
Principles of Digital Radiography with Large-Area, Electronically Readable Detectors: A Review of the Basics
The review compared large-area electronic X-ray detectors. It explained indirect conversion using a scintillator and light-sensitive elements, direct conversion using charge generated in a photoconductor, and how the detector array reads out the resulting signal. It also discussed implications for image quality.
SignificanceThe article documents the transition from systems requiring separate plate readout to electronically readable area detectors. Its distinction between direct and indirect conversion remains a framework used in medical physics teaching to explain how digital radiographic detectors collect signals.
DOI Harrell G. Chotas digital radiographyflat-panel detectorcomputed radiography