Modalities

radiographic film

X线胶片

Silver-halide film that recorded radiographs for most of the 20th century, usually exposed between two fluorescent intensifying screens. One sheet captured, displayed and archived the image at once, but its narrow exposure latitude left no way to rescue a bad exposure.

Film responds poorly to X-rays directly, so radiographic film carries emulsion on both sides and is exposed between two intensifying screens. Phosphor in the screens turns absorbed X-rays into visible light, and it is mostly that light that exposes the film, cutting the dose needed many times over. After development, the more silver deposited, the darker the film; darkness is measured as optical density.

Röntgen's first images were on glass photographic plates, soon replaced by flexible film. Film responds along an S-shaped curve with useful contrast only in the middle, a narrow dynamic range: a slight overexposure or underexposure turns the image black or white, and the only remedy is a retake. Early cellulose-nitrate film was also flammable and gave way to safety base.

Films were read on lightboxes, filed in jackets and fetched by hand. From the 1990s computed radiography, digital radiography and picture archiving and communication system moved diagnosis off film.

Key works · 2

Landmark1983Radiology978 citations

Computed radiography utilizing scanning laser stimulated luminescence.

M. Sonoda, M. Takano … H. Kato

The Fuji system recorded a latent X-ray image on a storage phosphor plate. A laser subsequently scanned the plate and released light in proportion to the stored signal, which the reader converted to digital values. The plate could then be erased and reused.

SignificanceStorage plates made it possible to digitize projection radiographs while retaining cassette-based acquisition in existing rooms. Digital processing and archiving could therefore be introduced before flat-panel detectors were widely available. The IAEA handbook describes this technology as computed radiography.

Key1972Radiology70 citations

X-Ray Exposure Reduction Using Rare-Earth Oxysulfide Intensifying Screens

Robert A. Buchanan, Solon I. Finkelstein, Kenneth A. Wickersheim

This study compared terbium-activated gadolinium and lanthanum oxysulfide screens with calcium tungstate. The rare-earth phosphors absorbed more incident radiation over part of the diagnostic energy range and converted it to light more efficiently. A gadolinium oxysulfide screen reduced exposure at comparable resolution even with blue-sensitive film.

SignificanceMatching rare-earth screens to green-sensitive film let radiographic systems use more of the light produced per exposure. UNSCEAR records faster rare-earth screens among the technical advances that reduced patient doses in conventional radiography during the screen-film era.