Physics · 1948

scintillator

闪烁晶体

A material that converts an absorbed X- or gamma-ray photon into a brief flash of visible light, which a photomultiplier or silicon photodetector turns into an electrical signal. Gamma cameras, PET, CT and many flat-panel X-ray detectors all rely on one.

A high-energy photon deposits its energy in the crystal as a cascade of excitations, which relax by emitting visible light. The flash is proportional to the energy deposited, so the detector can also measure photon energy and reject scattered photons. A good scintillator is dense, to stop the photons; bright, for energy resolution; and fast, for high count rates and precise timing.

Röntgen discovered X-rays because a fluorescent screen glowed. In 1948 Robert Hofstadter showed that thallium-doped sodium iodide, NaI(Tl), is an excellent scintillator, and it remains the standard crystal of the gamma camera, well matched to the 140 keV photons of technetium-99m.

PET must stop 511 keV photons and time them precisely, so it moved to much denser crystals: bismuth germanate (BGO), then lutetium oxyorthosilicate (LSO/LYSO), introduced in the 1990s and far faster, which made time-of-flight PET practical. CT detectors use cadmium tungstate or rare-earth ceramics, flat panels use needle-like caesium iodide, and photon-counting CT skips the scintillator, turning photons straight into charge in a semiconductor.

Sources
  1. Hofstadter R. Alkali halide scintillation counters. Phys Rev. 1948;74(1):100–101. doi.org/10.1103/PhysRev.74.100
  2. Melcher CL, Schweitzer JS. Cerium-doped lutetium oxyorthosilicate: a fast, efficient new scintillator. IEEE Trans Nucl Sci. 1992;39(4):502–505. doi.org/10.1109/23.159655

Key works · 6

Landmark1948Radiology95 citations

Fluoroscopic Image Brightening by Electronic Means

John W. Coltman

Coltman addressed the low brightness of conventional fluoroscopic screens by converting the X-ray pattern into an electron image. An electric field accelerated the electrons before they struck an output phosphor, adding energy to the image and yielding a brighter display at the same input exposure.

SignificanceElectronic amplification formed the basis of fluoroscopic image intensifiers. The AAPM museum documents how later systems combined it with image minification to increase gain further, eventually allowing the operator to view fluoroscopic images under ordinary room lighting.

Landmark1958Review of Scientific Instruments1,109 citations

Scintillation Camera

Hal O. Anger

Anger combined a pinhole shield with a scintillation crystal viewed by seven photomultiplier tubes. The detector used their signals to estimate where each flash occurred and selected events by pulse height. Accumulating the displayed positions formed an image of the radioactive source.

SignificancePosition-sensitive readout of one large scintillation crystal made practical planar gamma cameras possible. Rotating versions of the camera later acquired projections at multiple angles for SPECT, carrying the design into clinical emission tomography in nuclear medicine.

Landmark1975Radiology1,158 citations

A Positron-Emission Transaxial Tomograph for Nuclear Imaging (PETT)

Michel M. Ter-Pogossian, Michael E. Phelps … Nizar A. Mullani

The PETT combined a hexagonal array of 24 sodium iodide detectors with coincidence electronics and computerized cross-sectional reconstruction. Simulations, phantoms, and animal studies were used to evaluate its ability to recover tracer activity and compare its contrast and resolution with scintillation cameras.

SignificancePETT put multiangle coincidence measurement and computerized emission tomography into a single instrument. Later PET histories identify it as a precursor to clinical scanners that use detector rings and quantitative reconstruction to map tracer distributions.

Landmark1986IEEE Transactions on Nuclear Science553 citations

A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography

M. E. Casey, R. Nutt

The authors coupled an array of bismuth germanate scintillation crystals to four photomultiplier tubes through a light guide. The electronics decoded the position of each event from the shared optical signals. This allowed a PET detector to identify individual crystals in two dimensions with a small number of readout channels.

SignificanceThe block detector let PET scanners use many small crystals without dedicating a photomultiplier to each crystal, making finer spatial sampling practical. The first commercial scanner based on detector blocks was installed at Hammersmith and could read out multiple transaxial planes.

Key1972Radiology70 citations

X-Ray Exposure Reduction Using Rare-Earth Oxysulfide Intensifying Screens

Robert A. Buchanan, Solon I. Finkelstein, Kenneth A. Wickersheim

This study compared terbium-activated gadolinium and lanthanum oxysulfide screens with calcium tungstate. The rare-earth phosphors absorbed more incident radiation over part of the diagnostic energy range and converted it to light more efficiently. A gadolinium oxysulfide screen reduced exposure at comparable resolution even with blue-sensitive film.

SignificanceMatching rare-earth screens to green-sensitive film let radiographic systems use more of the light produced per exposure. UNSCEAR records faster rare-earth screens among the technical advances that reduced patient doses in conventional radiography during the screen-film era.

Key1997Medical Physics120 citations

Empirical investigation of the signal performance of a high-resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation

L. E. Antonuk, Y. El-Mohri … R. A. Street

This study measured the behavior of a large-area active-matrix flat panel in radiographic and fluoroscopic modes. An intensifying screen converted X-rays to light, which a matrix of amorphous-silicon photodiodes and thin-film transistors read out. The authors characterized dark signals, temporal response, and sensitivity with different phosphor screens.

SignificanceThe measurements showed which signal and timing properties must be managed to use a single indirect panel for still radiographs and successive fluoroscopy frames. Scintillators coupled to amorphous-silicon readout arrays subsequently became a common flat-panel detector architecture.