CT whose detector counts individual X-ray photons and measures each one's energy, instead of integrating all the energy into one signal. It gives sharper images with almost no electronic noise and spectral information in every scan. The first clinical system was cleared by the FDA in 2021.

A conventional CT detector integrates: a scintillator turns X-rays into light and the total light is measured, so low-energy photons count for little and electronic noise is added in. A photon-counting detector uses a semiconductor such as cadmium telluride, in which each X-ray photon produces a charge pulse whose height reflects its energy. Energy thresholds sort photons into bins, and noise below the lowest threshold is simply rejected.

With no light spreading in a scintillator, pixels can be much smaller, improving spatial resolution, and the energy bins give every scan the material-separating power of dual-energy CT. On 30 September 2021 Siemens Healthineers announced FDA 510(k) clearance of the NAEOTOM Alpha, the first clinical photon-counting CT.

The hard parts are pulse pile-up at the very high photon rates of clinical CT and charge sharing between neighboring pixels, problems that kept the technology in research prototypes for years.

Sources
  1. Mayo Clinic CT Clinical Innovation Center. FDA approval for Siemens NAEOTOM. 30 Sep 2021. ctcicblog.mayo.edu/2021/09/30/fda-approval-for-siemens-naeotom/
  2. Imaging Technology News. First photon-counting CT system cleared by the FDA. 2021. www.itnonline.com/article/first-photon-counting-ct-system-cleare

Key works · 3

Landmark2022Radiology652 citations

First Clinical Photon-counting Detector CT System: Technical Evaluation

Kishore Rajendran, Martin Petersilka … Thomas G. Flohr

The authors characterized a clinical dual-source photon-counting CT scanner with phantom measurements and example examinations from four participants. They measured spatial resolution, noise, section response, and iodine quantitation. The high-resolution configuration achieved a maximum measured in-plane resolution of 125 micrometers.

SignificanceThe evaluation documented performance at the clinical introduction of a commercial photon-counting CT system. A contemporaneous technical baseline for high-resolution imaging and spectral quantitation allowed later assessments, including the Radiology milestones review, to compare these capabilities with earlier CT technology.

Key2016Radiology223 citations

Abdominal Imaging with Contrast-enhanced Photon-counting CT: First Human Experience

Amir Pourmorteza, Rolf Symons … Gregory Henderson

Fifteen asymptomatic volunteers underwent paired, dose-matched contrast-enhanced abdominal CT scans with a photon-counting prototype and an energy-integrating detector. Readers gave the two methods similar scores for image quality, noise, and artifacts. Iodine maps generated from the photon-counting data improved the contrast-to-noise ratio for enhanced kidney tissue in the reported comparison.

SignificanceThis paired study established the feasibility of conventional contrast-enhanced abdominal imaging with a photon-counting prototype while demonstrating spectral iodine mapping from the same acquisition. The group subsequently extended abdominal material decomposition to distinguish two contrast agents in an in vivo animal study.

Key2018Radiology1,499 citations

Photon-counting CT: Technical Principles and Clinical Prospects

Martin J. Willemink, Mats Persson … Dominik Fleischmann

The review explains how photon-counting detectors register individual incoming X-rays and assign events using energy thresholds. It relates those detector properties to CT performance, including spatial resolution, image contrast, electronic noise, energy-resolved measurements, and potential clinical uses such as material separation.

SignificanceThis review connected detector design with measurable CT performance before commercial deployment. Rajendran and colleagues cited it in their technical evaluation of the first commercial photon-counting system and assessed spatial resolution, noise, and spectral quantitation on that scanner.