dual-energy CTDECT
双能CT
Scanning the same anatomy with two X-ray spectra. Because elements attenuate differently as energy changes, comparing the two data sets separates iodine, calcium, uric acid and other materials, producing iodine maps and virtual non-contrast images that ordinary CT cannot.
At diagnostic energies attenuation comes mainly from the photoelectric effect and Compton scattering. Photoelectric absorption rises roughly as the cube of atomic number and falls as the cube of photon energy; Compton scattering hardly depends on atomic number. Measurements at two energies give two equations per voxel, enough to decompose it into two basis materials such as water and iodine.
Robert Alvarez and Albert Macovski at Stanford laid out this energy-selective framework in 1976. The hardware takes several forms: Siemens' dual-source scanner of 2005, with two tube–detector pairs mounted roughly 90° apart; fast kilovolt switching on a single tube; and dual-layer detectors that split the spectrum at the detector.
Clinically, a virtual non-contrast image can be computed from a contrast scan, saving an exposure; uric acid crystals in gout can be told apart from calcium, kidney stones typed, and metal artifacts reduced. photon-counting CT carries the same idea further.
- Alvarez RE, Macovski A. Energy-selective reconstructions in X-ray computerised tomography. Phys Med Biol. 1976;21(5):733–744. doi.org/10.1088/0031-9155/21/5/002
- McCollough CH, Leng S, Yu L, Fletcher JG. Dual- and multi-energy CT: principles, technical approaches, and clinical applications. Radiology. 2015;276(3):637–653. doi.org/10.1148/radiol.2015142631
- Siemens Healthineers. Back to the future: 10 years of dual source CT. www.siemens-healthineers.com/en-us/computed-tomography/news/mso-
Key works · 3
Landmark1976Physics in Medicine & Biology2,291 citations
Energy-selective reconstructions in X-ray computerised tomography
Alvarez and Macovski represented the energy dependence of X-ray attenuation with a small number of coefficients. They showed that coarse spectral measurements, combined with conventional tomography, could recover maps of those coefficients and separate contributions that are mixed in an energy-integrated measurement.
SignificanceThe work established the basis representation used for material decomposition in dual- and multienergy CT. Later spectral imaging research uses this representation to distinguish materials whose attenuation can appear similar on a conventional energy-integrating image.
DOI Robert E. AlvarezAlbert Macovski dual-energy CTattenuation coefficientphotoelectric effectCompton scatteringmaterial decomposition
Landmark2006European Radiology1,545 citations
First performance evaluation of a dual-source CT (DSCT) system
This evaluation measured the performance of a scanner with two X-ray tubes and two detector systems offset by approximately a right angle. Electrocardiogram-gated single-segment reconstruction achieved an 83 ms temporal resolution. Acquiring data at two tube voltages also let the authors separate bone from iodine-filled vessels.
SignificanceThe geometry reduced the time interval represented by a cardiac section, addressing motion in coronary imaging. Subsequent commercial systems retained dual sources for fast cardiac acquisitions and also used the two tubes to collect data at different spectra for dual-energy CT.
DOI Thomas G. FlohrBernd M. Ohnesorge computed tomographytemporal resolutiondual-energy CTmaterial decomposition
Key2018Radiology1,499 citations
Photon-counting CT: Technical Principles and Clinical Prospects
The review explains how photon-counting detectors register individual incoming X-rays and assign events using energy thresholds. It relates those detector properties to CT performance, including spatial resolution, image contrast, electronic noise, energy-resolved measurements, and potential clinical uses such as material separation.
SignificanceThis review connected detector design with measurable CT performance before commercial deployment. Rajendran and colleagues cited it in their technical evaluation of the first commercial photon-counting system and assessed spatial resolution, noise, and spectral quantitation on that scanner.
DOI Full text Martin J. WilleminkDominik Fleischmann photon-counting CTdual-energy CTspatial resolutionmaterial decomposition