Image basics

spatial resolution

空间分辨率

How close two small objects can be and still appear separate. Measured in line pairs per millimeter or as the width of the point spread function, it depends on detector, focal spot, motion and reconstruction, not on pixel size alone.

The classic test is a pattern of black and white bars: the number of bar pairs per millimeter that can still be told apart is the resolution in lp/mm. A fuller description is the modulation transfer function (MTF), which gives the fraction of contrast preserved at each spatial frequency. Another intuitive measure is the full width at half maximum of the point spread function: how wide a single point is smeared.

Modalities differ widely. Clinical CT resolves around 0.5 mm in-plane, MRI typically about 1 mm, PET several millimeters, because a positron travels a short distance before annihilating and detector crystals have finite width. Making pixels smaller than the system blur does not make the image sharper.

Resolution is always bought with something else. Finer CT detail needs a sharper reconstruction kernel and brings more noise, unless dose goes up to compensate. Smaller MRI voxels cost signal-to-noise ratio or scan time.

Key works · 9

Landmark1986IEEE Transactions on Nuclear Science553 citations

A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography

M. E. Casey, R. Nutt

The authors coupled an array of bismuth germanate scintillation crystals to four photomultiplier tubes through a light guide. The electronics decoded the position of each event from the shared optical signals. This allowed a PET detector to identify individual crystals in two dimensions with a small number of readout channels.

SignificanceThe block detector let PET scanners use many small crystals without dedicating a photomultiplier to each crystal, making finer spatial sampling practical. The first commercial scanner based on detector blocks was installed at Hammersmith and could read out multiple transaxial planes.

Landmark2022Radiology652 citations

First Clinical Photon-counting Detector CT System: Technical Evaluation

Kishore Rajendran, Martin Petersilka … Thomas G. Flohr

The authors characterized a clinical dual-source photon-counting CT scanner with phantom measurements and example examinations from four participants. They measured spatial resolution, noise, section response, and iodine quantitation. The high-resolution configuration achieved a maximum measured in-plane resolution of 125 micrometers.

SignificanceThe evaluation documented performance at the clinical introduction of a commercial photon-counting CT system. A contemporaneous technical baseline for high-resolution imaging and spectral quantitation allowed later assessments, including the Radiology milestones review, to compare these capabilities with earlier CT technology.

Key1999Physics in Medicine & Biology716 citations

Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution

C. S. Levin, E. J. Hoffman

The authors simulated positron trajectories for several radionuclides and estimated how displacement before annihilation contributes to image blur. They combined that contribution with detector element size and photon noncollinearity to assess achievable resolution for PET systems with different detector dimensions.

SignificanceThe analysis quantified the contributions of isotope physics and scanner geometry to PET resolution. It gives designers and evaluators a reference for judging how much finer detectors can improve resolution before other physical effects dominate.

Key2000Radiology561 citations

Four Multidetector-Row Helical CT: Image Quality and Volume Coverage Speed

Hui Hu, H. David He … Stanley H. Fox

This study compared four- and single-detector-row helical CT using phantom measurements, models, and ratings of clinical images. The investigators measured section profiles, noise, and artifacts. The four-row system achieved two to three times the volume coverage speed at comparable diagnostic image quality.

SignificanceThe comparison provided an early quantitative baseline for the advantages of acquiring several detector rows together. Subsequent multidetector systems extended the same approach to faster coverage of organs while preserving thin sections for multiplanar reformats.

Key2005Medical Physics300 citations

Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot

T. G. Flohr, K. Stierstorfer … C. H. McCollough

A CT scanner alternated its focal spot between two longitudinal positions, interleaving the samples from 32 detector rows to acquire 64 overlapping sections per rotation. The study measured longitudinal resolution, section profiles, and noise and evaluated helical image quality using phantom and patient scans.

SignificanceThe scanner demonstrated that geometry could increase longitudinal sampling density beyond the physical detector row count. The technique was implemented in a commercial 64-section system, supporting thin-section volume acquisitions and reformats through the resulting data in multiple planes.

Key2007Journal of Nuclear Medicine739 citations

Performance of Philips Gemini TF PET/CT Scanner with Special Consideration for Its Time-of-Flight Imaging Capabilities

Suleman Surti, Austin Kuhn … Joel S. Karp

The study benchmarked the Gemini TF PET/CT scanner using standard and larger phantoms. Measurements covered count rates, spatial resolution, and coincidence timing. It also compared image reconstruction with and without timing information to assess contrast and noise for lesion-like objects.

SignificanceThe evaluation documented the first commercially available fully three-dimensional PET system with time-of-flight capability. The scanner provided a platform for subsequent phantom and patient studies that quantified how timing information changes lesion contrast and image noise.

Key2011Journal of Nuclear Medicine960 citations

Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner

Gaspar Delso, Sebastian Fürst … Sibylle I. Ziegler

Delso and colleagues measured an integrated whole-body PET/MR scanner while each subsystem operated independently and while both acquired data together. PET measurements included resolution, sensitivity, scatter, and count losses. MRI quality-control tests covered geometric accuracy and signal quality. The authors reported no perceptible effect on overall performance from integration under the conditions they tested.

SignificanceThe evaluation documented the first commercial integrated whole-body PET/MR scanner with simultaneous acquisition. Collecting emissions and MR data over the same period made it possible to investigate tracer behavior alongside MR measures of anatomy and function.

Key2018Radiology1,499 citations

Photon-counting CT: Technical Principles and Clinical Prospects

Martin J. Willemink, Mats Persson … Dominik Fleischmann

The review explains how photon-counting detectors register individual incoming X-rays and assign events using energy thresholds. It relates those detector properties to CT performance, including spatial resolution, image contrast, electronic noise, energy-resolved measurements, and potential clinical uses such as material separation.

SignificanceThis review connected detector design with measurable CT performance before commercial deployment. Rajendran and colleagues cited it in their technical evaluation of the first commercial photon-counting system and assessed spatial resolution, noise, and spectral quantitation on that scanner.

Key2021Journal of Nuclear Medicine449 citations

Performance Evaluation of the uEXPLORER Total-Body PET/CT Scanner Based on NEMA NU 2-2018 with Additional Tests to Characterize PET Scanners with a Long Axial Field of View

Benjamin A. Spencer, Eric Berg … Yasser G. Abdelhafez

The authors evaluated the 194 cm uEXPLORER using standard NEMA tests and extended phantoms for sensitivity and count-rate measurements. They also assessed image quality at several axial positions. Centered-source sensitivity was 174 thousand counts per second per megabecquerel with the standard line source and 147 thousand with a 170 cm line source.

SignificanceThe tests supplied reproducible performance benchmarks for the long-axial-field scanner behind the initial human images. They also documented how standard and human-length phantoms differed in key measurements. Subsequent Monte Carlo work used these data to validate simulated scanner sensitivity and count-rate performance.