Modalities

time-of-flight PETTOF-PET

飞行时间PET

positron emission tomography that times the arrival of each photon pair to within a few hundred picoseconds, narrowing where along the line of response the annihilation happened. Better timing means less noise, which can be traded for shorter scans or lower doses.

Both photons travel at the speed of light, so if the annihilation sits a distance from the midpoint between two detectors, they arrive apart. Turned around, a timing resolution localises the event to within : 400 ps gives 6 cm, 200 ps gives 3 cm. That is far too coarse to image directly, but confining each event to a short segment rather than the whole line cuts noise, with an effective sensitivity gain of roughly the patient's diameter divided by . Larger patients benefit most.

Time-of-flight scanners were built in the 1980s with caesium fluoride and barium fluoride crystals, but their poor stopping power lost out to BGO. Fast lutetium-based scintillators (LSO, LYSO) brought TOF back to commercial PET/CT around 2006, the Philips Gemini TF timing at about 585 ps, and current digital systems reach about 210 ps.

Sources
  1. Vandenberghe S, Mikhaylova E, D'Hoe E, Mollet P, Karp JS. Recent developments in time-of-flight PET. EJNMMI Phys. 2016;3:3. doi.org/10.1186/s40658-016-0138-3
  2. Surti S, et al. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med. 2007;48(3):471–480. jnm.snmjournals.org/content/48/3/471
  3. van Sluis J, et al. Performance characteristics of the digital Biograph Vision PET/CT system. J Nucl Med. 2019;60(7):1031–1036. doi.org/10.2967/jnumed.118.215418

Key works · 4

Key2007Journal of Nuclear Medicine739 citations

Performance of Philips Gemini TF PET/CT Scanner with Special Consideration for Its Time-of-Flight Imaging Capabilities

Suleman Surti, Austin Kuhn … Joel S. Karp

The study benchmarked the Gemini TF PET/CT scanner using standard and larger phantoms. Measurements covered count rates, spatial resolution, and coincidence timing. It also compared image reconstruction with and without timing information to assess contrast and noise for lesion-like objects.

SignificanceThe evaluation documented the first commercially available fully three-dimensional PET system with time-of-flight capability. The scanner provided a platform for subsequent phantom and patient studies that quantified how timing information changes lesion contrast and image noise.

Key2011Physics in Medicine & Biology439 citations

Design and performance evaluation of a whole-body Ingenuity TF PET–MRI system

H. Zaidi, N. Ojha … P. Maniawski

This study evaluated a sequential whole-body system combining a time-of-flight PET scanner with 3 T MRI. The PET photomultipliers received magnetic shielding and the electronics were relocated to reduce mutual interference. Measurements characterized PET performance and the use of MR-derived attenuation maps.

SignificanceThe Philips sequential design retained photomultiplier-based PET detectors through shielding and physical separation. In their evaluation of the simultaneous Siemens scanner, Delso and colleagues discussed the Philips design as a contemporaneous clinical alternative. Their standardized phantom measurements characterized the Siemens scanner.

Key2018Journal of Nuclear Medicine827 citations

Total-Body PET: Maximizing Sensitivity to Create New Opportunities for Clinical Research and Patient Care

Simon R. Cherry, Terry Jones … Ramsey D. Badawi

The authors reviewed geometric and timing improvements that could increase PET sensitivity. They estimated that extending axial coverage to the whole body could increase effective sensitivity about 40-fold for whole-body imaging and discussed the implications for scan duration, administered activity, and dynamic studies.

SignificanceThe article provided a quantitative rationale for a scanner with an extended axial field and distinguished the gains for whole-body imaging from those for a single organ. The EXPLORER system subsequently tested several of those opportunities in human imaging.

Key2021Journal of Nuclear Medicine449 citations

Performance Evaluation of the uEXPLORER Total-Body PET/CT Scanner Based on NEMA NU 2-2018 with Additional Tests to Characterize PET Scanners with a Long Axial Field of View

Benjamin A. Spencer, Eric Berg … Yasser G. Abdelhafez

The authors evaluated the 194 cm uEXPLORER using standard NEMA tests and extended phantoms for sensitivity and count-rate measurements. They also assessed image quality at several axial positions. Centered-source sensitivity was 174 thousand counts per second per megabecquerel with the standard line source and 147 thousand with a 170 cm line source.

SignificanceThe tests supplied reproducible performance benchmarks for the long-axial-field scanner behind the initial human images. They also documented how standard and human-length phantoms differed in key measurements. Subsequent Monte Carlo work used these data to validate simulated scanner sensitivity and count-rate performance.