Performance Evaluation of the uEXPLORER Total-Body PET/CT Scanner Based on NEMA NU 2-2018 with Additional Tests to Characterize PET Scanners with a Long Axial Field of View
The authors evaluated the 194 cm uEXPLORER using standard NEMA tests and extended phantoms for sensitivity and count-rate measurements. They also assessed image quality at several axial positions. Centered-source sensitivity was 174 thousand counts per second per megabecquerel with the standard line source and 147 thousand with a 170 cm line source.
SignificanceThe tests supplied reproducible performance benchmarks for the long-axial-field scanner behind the initial human images. They also documented how standard and human-length phantoms differed in key measurements. Subsequent Monte Carlo work used these data to validate simulated scanner sensitivity and count-rate performance.