total-body PET

全身PET

positron emission tomography with a detector ring long enough to cover the whole body at once, about two meters. Conventional scanners cover 15–30 cm at a time; collecting photons from everywhere simultaneously raises effective sensitivity dozens of times, allowing faster scans, lower doses or whole-body dynamic imaging.

A conventional PET scanner has an axial field of view of roughly 15–30 cm, so a whole-body study steps the bed through several positions, and at any moment most of the photons leaving the body never reach a detector crystal. The uEXPLORER, developed by Simon Cherry and Ramsey Badawi at the University of California, Davis, with United Imaging Healthcare in Shanghai, stretches the detector ring to 194 cm. Simulations predicted up to a 40-fold gain in effective count rate for total-body imaging, and the team reported its first human studies in 2019.

The extra sensitivity can be spent in different ways: much shorter scans, a much lower radiation dose, or dynamic imaging that follows a tracer through every organ at once from the moment of injection.

The price is the sheer quantity of detector crystal, which makes these scanners very expensive and, for now, rare.

Sources
  1. Badawi RD, et al. First human imaging studies with the EXPLORER total-body PET scanner. J Nucl Med. 2019;60(3):299–303. doi.org/10.2967/jnumed.119.226498
  2. Cherry SR, et al. Total-body PET: maximizing sensitivity to create new opportunities for clinical research and patient care. J Nucl Med. 2018;59(1):3–12. doi.org/10.2967/jnumed.116.184028

Key works · 3

Landmark2019Journal of Nuclear Medicine778 citations

First Human Imaging Studies with the EXPLORER Total-Body PET Scanner

Ramsey D. Badawi, Hongcheng Shi … Patricia M. Price

This report presented initial human imaging with the EXPLORER PET/CT scanner. Its 194 cm axial field of view allowed activity to be recorded across the body at one bed position. Examples illustrated several ways of using the added sensitivity, including rapid, low-activity, delayed, and dynamic whole-body studies.

SignificanceSimultaneous coverage made it possible to observe tracer kinetics in multiple organs in the same time frame. Subsequent characterization of the instrument measured sensitivity, count-rate performance, and image quality, creating benchmarks for later long-axial-field systems.

Key2018Journal of Nuclear Medicine827 citations

Total-Body PET: Maximizing Sensitivity to Create New Opportunities for Clinical Research and Patient Care

Simon R. Cherry, Terry Jones … Ramsey D. Badawi

The authors reviewed geometric and timing improvements that could increase PET sensitivity. They estimated that extending axial coverage to the whole body could increase effective sensitivity about 40-fold for whole-body imaging and discussed the implications for scan duration, administered activity, and dynamic studies.

SignificanceThe article provided a quantitative rationale for a scanner with an extended axial field and distinguished the gains for whole-body imaging from those for a single organ. The EXPLORER system subsequently tested several of those opportunities in human imaging.

Key2021Journal of Nuclear Medicine449 citations

Performance Evaluation of the uEXPLORER Total-Body PET/CT Scanner Based on NEMA NU 2-2018 with Additional Tests to Characterize PET Scanners with a Long Axial Field of View

Benjamin A. Spencer, Eric Berg … Yasser G. Abdelhafez

The authors evaluated the 194 cm uEXPLORER using standard NEMA tests and extended phantoms for sensitivity and count-rate measurements. They also assessed image quality at several axial positions. Centered-source sensitivity was 174 thousand counts per second per megabecquerel with the standard line source and 147 thousand with a 170 cm line source.

SignificanceThe tests supplied reproducible performance benchmarks for the long-axial-field scanner behind the initial human images. They also documented how standard and human-length phantoms differed in key measurements. Subsequent Monte Carlo work used these data to validate simulated scanner sensitivity and count-rate performance.