Modalities · 1998

PET/CT

A positron emission tomography scanner and a computed tomography scanner in one gantry, so a single visit yields both scans already aligned. PET shows where metabolism is abnormal, CT shows exactly where that is, and the CT data also correct the PET image for attenuation.

The patient lies on one table: a CT scan of a few tens of seconds, then the PET acquisition, with the bed position known exactly, so the two image sets overlay without separate image registration. The CT also replaced the slow radioactive-source transmission scan older PET systems needed. Rescaled to 511 keV, its attenuation values correct the PET data for photon attenuation, shortening the whole examination considerably.

David Townsend, engineer Ronald Nutt of the company CTI and colleagues built the first PET/CT prototype at the University of Pittsburgh in 1998 and reported it in 2000. Commercial systems arrived in 2001, and stand-alone PET scanners soon faded from the market.

PET/CT is now standard for staging cancer, judging response to treatment and planning radiotherapy. The costs are the extra radiation dose from the CT and misalignment caused by breathing, most noticeable near the diaphragm.

Sources
  1. Beyer T, Townsend DW, Brun T, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med. 2000;41(8):1369–1379. jnm.snmjournals.org/content/41/8/1369
  2. Townsend DW. Combined positron emission tomography–computed tomography: the historical perspective. Semin Ultrasound CT MR. 2008;29(4):232–235. doi.org/10.1053/j.sult.2008.05.006

Key works · 8

Landmark2000Journal of Nuclear Medicine1,977 citations

A Combined PET/CT Scanner for Clinical Oncology

Thomas Beyer, David W. Townsend … Raymond Roddy

The prototype mounted a spiral CT scanner and a rotating PET scanner on a common gantry and acquired both studies using the same patient support. The CT data supplied localization and correction information for PET. The report described scanner performance and experience imaging more than 110 patients with cancer.

SignificanceThe prototype made it possible to localize metabolic findings against anatomy acquired during the same examination. Clinical evaluation began in 1998. The first commercial PET/CT system entered clinical use in 2001 and the combined modality subsequently became common in oncology.

Landmark2019Journal of Nuclear Medicine778 citations

First Human Imaging Studies with the EXPLORER Total-Body PET Scanner

Ramsey D. Badawi, Hongcheng Shi … Patricia M. Price

This report presented initial human imaging with the EXPLORER PET/CT scanner. Its 194 cm axial field of view allowed activity to be recorded across the body at one bed position. Examples illustrated several ways of using the added sensitivity, including rapid, low-activity, delayed, and dynamic whole-body studies.

SignificanceSimultaneous coverage made it possible to observe tracer kinetics in multiple organs in the same time frame. Subsequent characterization of the instrument measured sensitivity, count-rate performance, and image quality, creating benchmarks for later long-axial-field systems.

Key1998Medical Physics1,081 citations

Attenuation correction for a combined 3D PET/CT scanner

P. E. Kinahan, D. W. Townsend … D. Sashin

The authors tested whether a registered CT image could supply an attenuation map for 511 keV PET photons. Using bone and soft-tissue phantoms and human scans, they compared conversion methods and demonstrated a hybrid mapping from CT attenuation to PET attenuation.

SignificanceThe proof of principle gave CT a quantitative function in hybrid imaging alongside anatomical localization. The first reported clinical PET/CT prototype incorporated CT-derived correction, allowing combined scans to use the CT data in place of a separate PET transmission acquisition.

Key2004Journal of Digital Imaging1,946 citations

OsiriX: An Open-Source Software for Navigating in Multidimensional DICOM Images

Antoine Rosset, Luca Spadola, Osman Ratib

The authors developed an open-source desktop application for navigating large DICOM image series, with multiplanar reformatting, volume rendering, and fusion of PET and CT. The system used ITK and VTK libraries for image processing and rendering and demonstrated navigation with multidimensional, multimodality studies.

SignificanceOsiriX combined DICOM communication and interactive volume display in an extensible open-source desktop application. Follow-up reports described its use for interpreting large PET/CT studies and an extension for sending and retrieving DICOM series between computers over a network.

Key2007Journal of Nuclear Medicine739 citations

Performance of Philips Gemini TF PET/CT Scanner with Special Consideration for Its Time-of-Flight Imaging Capabilities

Suleman Surti, Austin Kuhn … Joel S. Karp

The study benchmarked the Gemini TF PET/CT scanner using standard and larger phantoms. Measurements covered count rates, spatial resolution, and coincidence timing. It also compared image reconstruction with and without timing information to assess contrast and noise for lesion-like objects.

SignificanceThe evaluation documented the first commercially available fully three-dimensional PET system with time-of-flight capability. The scanner provided a platform for subsequent phantom and patient studies that quantified how timing information changes lesion contrast and image noise.

Key2015European Journal of Nuclear Medicine and Molecular Imaging3,675 citations

FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0

Ronald Boellaard, Roberto Delgado-Bolton … Wolfgang Eschner

The EANM updated its guidance for adult oncologic FDG PET/CT, covering preparation, acquisition, reconstruction, quality control, quantitation, and reporting. It specified procedures for repeatability on one scanner and for comparability across instruments and sites, including harmonization of quantitative performance across imaging centers.

SignificanceThe guideline specifies procedures and equipment checks for quantitative comparability. The EANM's EARL accreditation report documented implementation across about 200 scanners at 150 sites. Phantom measurements identified calibration errors, and corrective feedback brought participating scanners into compliance with the calibration specifications.

Key2020The Lancet2,306 citations

Prostate-specific membrane antigen PET-CT in patients with high-risk prostate cancer before curative-intent surgery or radiotherapy (proPSMA): a prospective, randomised, multicentre study

Michael S. Hofman, Nathan Lawrentschuk … Paul Thomas

The Australian multicenter trial randomly assigned men with high-risk prostate cancer to initial staging using PSMA-targeted PET/CT or conventional CT and bone scanning. Against a prespecified reference standard, the targeted examination more accurately identified pelvic nodal or distant metastatic disease.

SignificanceThe trial supplied comparative evidence for a target-specific PET tracer in initial cancer staging. The joint EANM and SNMMI PSMA imaging procedure guideline cites the study in assessing evidence for staging patients with higher-risk disease.

Key2021Journal of Nuclear Medicine449 citations

Performance Evaluation of the uEXPLORER Total-Body PET/CT Scanner Based on NEMA NU 2-2018 with Additional Tests to Characterize PET Scanners with a Long Axial Field of View

Benjamin A. Spencer, Eric Berg … Yasser G. Abdelhafez

The authors evaluated the 194 cm uEXPLORER using standard NEMA tests and extended phantoms for sensitivity and count-rate measurements. They also assessed image quality at several axial positions. Centered-source sensitivity was 174 thousand counts per second per megabecquerel with the standard line source and 147 thousand with a 170 cm line source.

SignificanceThe tests supplied reproducible performance benchmarks for the long-axial-field scanner behind the initial human images. They also documented how standard and human-length phantoms differed in key measurements. Subsequent Monte Carlo work used these data to validate simulated scanner sensitivity and count-rate performance.