ordered-subsets expectation maximizationOSEM
有序子集期望最大化
Ordered-subsets expectation maximization is a statistical reconstruction method that divides projections into groups and updates the image after each group, accelerating the early stages of emission tomography reconstruction.
In iterative reconstruction of SPECT data, OSEM starts with an estimate of the activity distribution. A system model predicts the photon counts for a group of viewing angles. The ratios of measured to predicted counts are back-projected to calculate the image update. The updated image is immediately used with the next group of projections. Each group typically contains views spread across several angles.
Full-data expectation maximization uses all projections before making an update. With subsets, several updates occur during the same pass through the data. Individual subsets provide somewhat different statistical information, and continued iteration can increase image noise. The number of subsets and the stopping point both help determine the resulting image.
- Hudson HM, Larkin RS. Accelerated Image Reconstruction Using Ordered Subsets of Projection Data. IEEE Trans Med Imaging, 1994. pubmed.ncbi.nlm.nih.gov/18218538/
Key works · 2
Landmark1994IEEE Transactions on Medical Imaging4,178 citations
Accelerated image reconstruction using ordered subsets of projection data
Hudson and Larkin split projection data into ordered subsets and used each subset to perform an expectation-maximization update before moving to the next. A full iteration visits every subset. This accelerated the early progress of image reconstruction in the experiments they reported.
SignificanceOrdered-subsets expectation-maximization reduced the time required to reach useful image quality, making statistical reconstruction more practical in clinical nuclear medicine. IAEA guidance identifies it as a commonly used reconstruction approach for quantitative PET examinations in clinical practice.
DOI Harold Malcolm HudsonR. S. Larkin iterative reconstructionpositron emission tomographysingle-photon emission computed tomographyordered-subsets expectation maximization
Landmark2000Journal of Nuclear Medicine1,977 citations
A Combined PET/CT Scanner for Clinical Oncology
The prototype mounted a spiral CT scanner and a rotating PET scanner on a common gantry and acquired both studies using the same patient support. The CT data supplied localization and correction information for PET. The report described scanner performance and experience imaging more than 110 patients with cancer.
SignificanceThe prototype made it possible to localize metabolic findings against anatomy acquired during the same examination. Clinical evaluation began in 1998. The first commercial PET/CT system entered clinical use in 2001 and the combined modality subsequently became common in oncology.
Full text Thomas BeyerDavid W. TownsendRonald Nutt PET/CTimage registrationattenuationordered-subsets expectation maximizationattenuation correction