Reconstruction

attenuation correction

衰减校正

Attenuation correction accounts for photons lost before detection in emission imaging. It uses an estimate of tissue attenuation to reduce bias in the reconstructed distribution of tracer activity.

Identical tracer activities near the surface and deep in the body can yield different detected counts. In PET, both photons must reach their detectors. The probability that both reach the detectors without attenuation depends on the attenuation coefficient integrated along the full line connecting those detectors. Reconstruction can use that probability when estimating how much activity generated the observed coincidences.

In PET/CT, CT values are converted into attenuation coefficients appropriate to the energy of the emitted photons. The resulting map is aligned with the emission data. If the diaphragm moves between acquisitions, a region of soft tissue may be assigned a correction appropriate to lung. The alignment error can then appear as a local error in estimated uptake.

Sources
  1. Lee WW, K-SPECT Group. Clinical Applications of Technetium-99m Quantitative Single-Photon Emission Computed Tomography/Computed Tomography. Nucl Med Mol Imaging, 2019. pmc.ncbi.nlm.nih.gov/articles/PMC6554391/
  2. Sun T, Mok GSP. Techniques for Respiration-induced Artifacts Reductions in Thoracic PET/CT. Quant Imaging Med Surg, 2012. pmc.ncbi.nlm.nih.gov/articles/PMC3496495/

Key works · 5

Landmark2000Journal of Nuclear Medicine1,977 citations

A Combined PET/CT Scanner for Clinical Oncology

Thomas Beyer, David W. Townsend … Raymond Roddy

The prototype mounted a spiral CT scanner and a rotating PET scanner on a common gantry and acquired both studies using the same patient support. The CT data supplied localization and correction information for PET. The report described scanner performance and experience imaging more than 110 patients with cancer.

SignificanceThe prototype made it possible to localize metabolic findings against anatomy acquired during the same examination. Clinical evaluation began in 1998. The first commercial PET/CT system entered clinical use in 2001 and the combined modality subsequently became common in oncology.

Key1979Physics in Medicine & Biology133 citations

Whole-body single-photon emission computed tomography using dual, large-field-of-view scintillation cameras

R. J. Jaszczak, Lee-Tzuu Chang … F. E. Moore

This system mounted two large-area scintillation cameras on a rotating gantry. It reconstructed multiple contiguous transverse sections from the projections collected in one continuous revolution. The implementation used weighted filtered back-projection, offered different collimators, and included a correction for photon attenuation.

SignificanceThe system showed that dual heads, continuous rotation, and simultaneous coverage of several sections could be combined in one instrument. Rotating gamma cameras subsequently became a standard implementation of SPECT and provided the emission component of hybrid SPECT/CT.

Key1996IEEE Transactions on Nuclear Science196 citations

Attenuation correction of SPECT using X-ray CT on an emission-transmission CT system: myocardial perfusion assessment

S. C. Blankespoor, X. Xu … B. H. Hasegawa

The researchers registered data from juxtaposed CT and SPECT scanners and converted the CT measurements to attenuation maps at 140 keV. Attenuation correction reduced underestimation of myocardial activity in a torso phantom and in pigs. An additional partial-volume correction improved agreement with tissue samples in pigs but overestimated activity in the phantom.

SignificanceThis study quantified the benefit and limitations of using registered CT data for attenuation correction on a combined emission–transmission system. UCSF traces its human-scale SPECT/CT prototype for clinical research to the group's work on registered emission and transmission imaging.

Key1998Medical Physics1,081 citations

Attenuation correction for a combined 3D PET/CT scanner

P. E. Kinahan, D. W. Townsend … D. Sashin

The authors tested whether a registered CT image could supply an attenuation map for 511 keV PET photons. Using bone and soft-tissue phantoms and human scans, they compared conversion methods and demonstrated a hybrid mapping from CT attenuation to PET attenuation.

SignificanceThe proof of principle gave CT a quantitative function in hybrid imaging alongside anatomical localization. The first reported clinical PET/CT prototype incorporated CT-derived correction, allowing combined scans to use the CT data in place of a separate PET transmission acquisition.

Key2011Physics in Medicine & Biology439 citations

Design and performance evaluation of a whole-body Ingenuity TF PET–MRI system

H. Zaidi, N. Ojha … P. Maniawski

This study evaluated a sequential whole-body system combining a time-of-flight PET scanner with 3 T MRI. The PET photomultipliers received magnetic shielding and the electronics were relocated to reduce mutual interference. Measurements characterized PET performance and the use of MR-derived attenuation maps.

SignificanceThe Philips sequential design retained photomultiplier-based PET detectors through shielding and physical separation. In their evaluation of the simultaneous Siemens scanner, Delso and colleagues discussed the Philips design as a contemporaneous clinical alternative. Their standardized phantom measurements characterized the Siemens scanner.