Modalities · 1957

gamma camera

γ相机

The standard detector of nuclear medicine. A lead collimator admits only gamma rays traveling straight in, a large scintillator crystal turns each into a flash of light, and an array of photomultiplier tubes works out where it landed, building a flat map of the tracer.

Its core is a single crystal of thallium-doped sodium iodide, tens of centimeters across. In front sits a lead collimator drilled with thousands of parallel holes; only photons traveling along a hole get through, so only about one in ten thousand is used. Each photon that reaches the crystal makes a flash, and dozens of photomultiplier tubes behind it share the light; a weighted average of their signals locates the flash. This is Anger logic.

Hal Anger built the camera at Berkeley's Donner Laboratory in 1957 and published it the following year; the first commercial unit went to Ohio State University in 1962. The rectilinear scanners it replaced had to move a detector point by point across the patient.

Resolution is modest, typically around a centimeter, far coarser than CT. But a gamma camera shows what anatomical imaging cannot: bone metastases, thyroid function, blood flow to the heart muscle. Rotated around the patient, it becomes single-photon emission computed tomography.

Sources
  1. Anger HO. Scintillation camera. Rev Sci Instrum. 1958;29(1):27–33. doi.org/10.1063/1.1715998
  2. Tapscott E. Nuclear medicine pioneer: Hal O. Anger. First scintillation camera is foundation for modern imaging systems. J Nucl Med. 1998;39(3):15N–19N. jnm.snmjournals.org/content/jnumed/39/3/15N.full.pdf

Key works · 3

Landmark1958Review of Scientific Instruments1,109 citations

Scintillation Camera

Hal O. Anger

Anger combined a pinhole shield with a scintillation crystal viewed by seven photomultiplier tubes. The detector used their signals to estimate where each flash occurred and selected events by pulse height. Accumulating the displayed positions formed an image of the radioactive source.

SignificancePosition-sensitive readout of one large scintillation crystal made practical planar gamma cameras possible. Rotating versions of the camera later acquired projections at multiple angles for SPECT, carrying the design into clinical emission tomography in nuclear medicine.

Key1979Physics in Medicine & Biology133 citations

Whole-body single-photon emission computed tomography using dual, large-field-of-view scintillation cameras

R. J. Jaszczak, Lee-Tzuu Chang … F. E. Moore

This system mounted two large-area scintillation cameras on a rotating gantry. It reconstructed multiple contiguous transverse sections from the projections collected in one continuous revolution. The implementation used weighted filtered back-projection, offered different collimators, and included a correction for photon attenuation.

SignificanceThe system showed that dual heads, continuous rotation, and simultaneous coverage of several sections could be combined in one instrument. Rotating gamma cameras subsequently became a standard implementation of SPECT and provided the emission component of hybrid SPECT/CT.

Key2010European Journal of Nuclear Medicine and Molecular Imaging197 citations

Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation

Ronny R. Buechel, Bernhard A. Herzog … Valerie Treyer

Seventy-five consecutive patients underwent myocardial perfusion imaging with conventional dual-head SPECT and a dedicated cadmium–zinc–telluride camera. The new system acquired stress and rest scans in three and two minutes, respectively, compared with about 15 minutes each for conventional imaging. The patient-level agreement in clinical classification was 96.0%.

SignificanceThe paired clinical study quantified how much a dedicated semiconductor camera could shorten acquisitions while preserving agreement with conventional myocardial perfusion imaging. The EANM subsequently cited it as evidence in its procedure guideline for cardiac-centered gamma cameras.