MG·LB · Researchers

Bernhard A. Herzog

Co-led clinical testing of rapid myocardial perfusion imaging with a dedicated semiconductor camera.

Nuclear medicine and molecular imaging OpenAlex

University Hospital Zurich (2010)

Key works · 1

Key2010European Journal of Nuclear Medicine and Molecular Imaging197 citations

Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation

Ronny R. Buechel, Bernhard A. Herzog … Valerie Treyer

Seventy-five consecutive patients underwent myocardial perfusion imaging with conventional dual-head SPECT and a dedicated cadmium–zinc–telluride camera. The new system acquired stress and rest scans in three and two minutes, respectively, compared with about 15 minutes each for conventional imaging. The patient-level agreement in clinical classification was 96.0%.

SignificanceThe paired clinical study quantified how much a dedicated semiconductor camera could shorten acquisitions while preserving agreement with conventional myocardial perfusion imaging. The EANM subsequently cited it as evidence in its procedure guideline for cardiac-centered gamma cameras.