single-photon emission computed tomographySPECT
单光子发射计算机断层成像
A gamma camera rotated around the patient, recording the tracer from dozens of angles, with slices reconstructed much as in CT. It took nuclear medicine from flat pictures into three dimensions and is used above all to image blood flow to the heart muscle.
One or two camera heads circle the body through 180° or 360°, pausing every few degrees for tens of seconds and collecting some 60–120 projections. Slices are then reconstructed with filtered back-projection or, more often today, iterative reconstruction. Single photon refers to isotopes such as technetium-99m, which emit one gamma photon per decay whose direction only the collimator can fix. positron emission tomography instead locates paired photons electronically, which is why SPECT trails it in sensitivity and resolution.
David Kuhl and Roy Edwards at the University of Pennsylvania published their experiments in radioisotope section scanning in 1963; commercial SPECT systems built on rotating gamma cameras followed in the late 1970s.
Most new systems pair SPECT with CT, which corrects for photon attenuation in the body and anchors each hot spot to anatomy. Beyond cardiology, SPECT images dopamine transporters in suspected Parkinson's disease and localises bone lesions.
- Kuhl DE, Edwards RQ. Image separation radioisotope scanning. Radiology. 1963;80(4):653–662. doi.org/10.1148/80.4.653
Key works · 8
Landmark1963Radiology540 citations
Image Separation Radioisotope Scanning
Kuhl and Edwards sought to separate radioactive sources at different depths that appeared superimposed on a conventional scan. They adapted ideas from section radiography, recording counts from multiple directions and superimposing the resulting signals on film. The experiment separated radioactive sources located in different sections of a test phantom.
SignificanceThe experiments established that measured emissions could form a transverse section. Later histories identify them as early demonstrations of SPECT. Rotating gamma cameras subsequently used the same multiview principle to reconstruct clinical sections with computers.
DOI David E. KuhlRoy Q. Edwards single-photon emission computed tomographytomographyprojection
Landmark1982IEEE Transactions on Medical Imaging5,027 citations
Maximum Likelihood Reconstruction for Emission Tomography
Shepp and Vardi modeled unobserved emissions as Poisson variables and specified the probabilities that emissions from each voxel would be registered by each detector. Their expectation-maximization algorithm iteratively estimated the underlying emission distribution by maximizing the likelihood of the measured counts.
SignificanceThe model brought counting statistics and detection probabilities directly into emission reconstruction. Maximum-likelihood expectation-maximization became a basis for iterative PET and SPECT, including subsequent ordered-subsets methods that sought to reduce the cost of reaching a useful image.
DOI Full text Lawrence A. SheppYehuda Vardi iterative reconstructionpositron emission tomographysingle-photon emission computed tomography
Landmark1994IEEE Transactions on Medical Imaging4,178 citations
Accelerated image reconstruction using ordered subsets of projection data
Hudson and Larkin split projection data into ordered subsets and used each subset to perform an expectation-maximization update before moving to the next. A full iteration visits every subset. This accelerated the early progress of image reconstruction in the experiments they reported.
SignificanceOrdered-subsets expectation-maximization reduced the time required to reach useful image quality, making statistical reconstruction more practical in clinical nuclear medicine. IAEA guidance identifies it as a commonly used reconstruction approach for quantitative PET examinations in clinical practice.
DOI Harold Malcolm HudsonR. S. Larkin iterative reconstructionpositron emission tomographysingle-photon emission computed tomographyordered-subsets expectation maximization
Key1979Physics in Medicine & Biology133 citations
Whole-body single-photon emission computed tomography using dual, large-field-of-view scintillation cameras
This system mounted two large-area scintillation cameras on a rotating gantry. It reconstructed multiple contiguous transverse sections from the projections collected in one continuous revolution. The implementation used weighted filtered back-projection, offered different collimators, and included a correction for photon attenuation.
SignificanceThe system showed that dual heads, continuous rotation, and simultaneous coverage of several sections could be combined in one instrument. Rotating gamma cameras subsequently became a standard implementation of SPECT and provided the emission component of hybrid SPECT/CT.
DOI Ronald J. JaszczakF. E. Moore single-photon emission computed tomographygamma camerafiltered back-projectionattenuation correction
Key1989IEEE Transactions on Medical Imaging198 citations
Three-dimensional display in nuclear medicine
The investigators compared several surface and volume display techniques for nuclear medicine data, including maximum-activity projection, which keeps the highest activity along each viewing ray. Scintigraphic examples illustrated how the different projections represented focal uptake and spatial relationships in the volume.
SignificanceMaximum-activity projection made prominent foci in a volume visible in a single projection. The same maximum-value operation became a display option for CT angiography, where it highlights structures with high attenuation, including vessels opacified with contrast material.
DOI Jerold W. WallisE. C. Kleerup maximum intensity projectionvolume renderingsingle-photon emission computed tomography
Key1996IEEE Transactions on Nuclear Science196 citations
Attenuation correction of SPECT using X-ray CT on an emission-transmission CT system: myocardial perfusion assessment
The researchers registered data from juxtaposed CT and SPECT scanners and converted the CT measurements to attenuation maps at 140 keV. Attenuation correction reduced underestimation of myocardial activity in a torso phantom and in pigs. An additional partial-volume correction improved agreement with tissue samples in pigs but overestimated activity in the phantom.
SignificanceThis study quantified the benefit and limitations of using registered CT data for attenuation correction on a combined emission–transmission system. UCSF traces its human-scale SPECT/CT prototype for clinical research to the group's work on registered emission and transmission imaging.
DOI Stephen C. BlankespoorBruce H. Hasegawa single-photon emission computed tomographycomputed tomographyattenuation correction
Key2004Physics in Medicine & Biology2,510 citations
GATE: a simulation toolkit for PET and SPECT
GATE wrapped Geant4 in a scripted toolkit for simulating PET and SPECT measurements. It let users describe detector or source motion and tracer decay over time. The paper provided benchmark examples for installation and validation and discussed comparison with real scanners.
SignificanceA shared simulation toolkit made it possible for groups to compare detectors, acquisition protocols, and correction algorithms under specified conditions. The OpenGATE collaboration continued to maintain the toolkit and subsequently extended it to additional imaging and dosimetry applications.
DOI arXiv Sébastien JanChristian Morel positron emission tomographysingle-photon emission computed tomographyradiotracer
Key2010European Journal of Nuclear Medicine and Molecular Imaging197 citations
Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation
Seventy-five consecutive patients underwent myocardial perfusion imaging with conventional dual-head SPECT and a dedicated cadmium–zinc–telluride camera. The new system acquired stress and rest scans in three and two minutes, respectively, compared with about 15 minutes each for conventional imaging. The patient-level agreement in clinical classification was 96.0%.
SignificanceThe paired clinical study quantified how much a dedicated semiconductor camera could shorten acquisitions while preserving agreement with conventional myocardial perfusion imaging. The EANM subsequently cited it as evidence in its procedure guideline for cardiac-centered gamma cameras.
DOI Ronny R. BuechelBernhard A. HerzogPhilipp A. Kaufmann single-photon emission computed tomographygamma camera