Reconstruction · 1967

filtered back-projectionFBP

滤波反投影

The classic CT reconstruction: sharpen each projection with a ramp filter, then back-project. One pass gives a sharp slice, which is why it was the standard way to make CT images for some four decades.

Simple back-projection returns the image blurred by , which in Fourier terms means every spatial frequency component has been scaled down by . Filtered back-projection multiplies each projection by first (or, equivalently, convolves it with the matching kernel), and the two effects cancel. Each projection is filtered once and smeared once, so reconstruction can begin before the scan has finished.

The idea came from radio astronomy: Ronald Bracewell and Anthony Riddle used it in 1967 to recover sky maps from fan-beam scans. In 1971 G. N. Ramachandran and A. V. Lakshminarayanan published a convolution version for radiographs and electron micrographs, and their filter is still called Ram-Lak. Feldkamp, Davis and Kress extended the method to cone beams in 1984.

FBP is fast and predictable, but it assumes complete data and modest noise. At low dose the ramp filter amplifies noise, and metal produces bright and dark streaks: the problems that iterative reconstruction and deep-learning reconstruction were developed to solve.

Sources
  1. Bracewell RN, Riddle AC. Inversion of fan-beam scans in radio astronomy. Astrophys J. 1967;150:427–434. doi.org/10.1086/149346
  2. Ramachandran GN, Lakshminarayanan AV. Three-dimensional reconstruction from radiographs and electron micrographs: application of convolutions instead of Fourier transforms. PNAS. 1971;68(9):2236–2240. doi.org/10.1073/pnas.68.9.2236
  3. Feldkamp LA, Davis LC, Kress JW. Practical cone-beam algorithm. J Opt Soc Am A. 1984;1(6):612–619. doi.org/10.1364/JOSAA.1.000612

Key works · 7

Landmark1971Proceedings of the National Academy of Sciences1,192 citations

Three-dimensional Reconstruction from Radiographs and Electron Micrographs: Application of Convolutions instead of Fourier Transforms

G. N. Ramachandran, A. V. Lakshminarayanan

The paper replaced direct two-dimensional Fourier computation in projection reconstruction with a one-dimensional convolution on each projection. The authors assembled the filtered data from different angles into sections and reported faster computation and greater accuracy in their tests using simulated projections.

SignificanceThe formulation made the filtering stage of CT reconstruction practical with a spatial-domain kernel. The IAEA handbook describes convolution followed by back-projection as a classical CT method and identifies the ramp filter by the authors' names, Ram–Lak.

Landmark1974IEEE Transactions on Nuclear Science2,514 citations

The Fourier reconstruction of a head section

L. A. Shepp, B. F. Logan

Shepp and Logan used a mathematical head phantom built from ellipses to test a Fourier reconstruction against a search algorithm. They proposed a modified weighting function and compared the methods for computational effort, spatial resolution, and sensitivity to noise in the simulated data.

SignificanceThe phantom gave researchers a reproducible object with a known structure and calculable projections. The Shepp–Logan phantom remains available in numerical imaging software and continues to serve as a common test object for tomographic reconstruction.

Landmark1984Journal of the Optical Society of America A7,216 citations

Practical cone-beam algorithm

L. A. Feldkamp, L. C. Davis, J. W. Kress

Feldkamp, Davis, and Kress derived an approximate three-dimensional reconstruction from two-dimensional cone-beam projections along a circular trajectory. Their method uses convolution and back-projection, reduces to fan-beam reconstruction in the central plane, and was demonstrated using projections calculated from a mathematical phantom.

SignificanceThe FDK algorithm extended filtered back-projection to practical cone-beam volumes. Its computational form became a standard analytical baseline for circular cone-beam CT, including comparisons that test artifact suppression and image recovery with limited angular sampling.

Landmark2017IEEE Transactions on Image Processing2,968 citations

Deep Convolutional Neural Network for Inverse Problems in Imaging

Kyong Hwan Jin, Michael T. McCann … Michael Unser

FBPConvNet first applied filtered back-projection to sparse-view CT data, then used a multiresolution residual convolutional network to remove artifacts in the initial image. The authors tested synthetic phantoms and real experimental sinograms. On their more realistic phantoms, it outperformed a total-variation reconstruction baseline.

SignificanceThe paper provided a prominent baseline for combining a conventional inverse with learned artifact correction. The Learned Primal-Dual study later used learned post-processing of filtered back-projection as a comparator for a network that instead placed the measurement operator inside the reconstruction.

Key1979Physics in Medicine & Biology133 citations

Whole-body single-photon emission computed tomography using dual, large-field-of-view scintillation cameras

R. J. Jaszczak, Lee-Tzuu Chang … F. E. Moore

This system mounted two large-area scintillation cameras on a rotating gantry. It reconstructed multiple contiguous transverse sections from the projections collected in one continuous revolution. The implementation used weighted filtered back-projection, offered different collimators, and included a correction for photon attenuation.

SignificanceThe system showed that dual heads, continuous rotation, and simultaneous coverage of several sections could be combined in one instrument. Rotating gamma cameras subsequently became a standard implementation of SPECT and provided the emission component of hybrid SPECT/CT.

Notable2019FDA 510(k) Premarket Notification K183202

Deep Learning Image Reconstruction

GE Medical Systems, LLC

The 510(k) record covers the CT reconstruction option that produces TrueFidelity images. The manufacturer integrated a neural network into the scanner's raw-data reconstruction chain and evaluated image properties including noise texture. In April 2019 the FDA determined that the option was substantially equivalent to legally marketed devices.

SignificanceThe record provides primary documentation of a commercial scanner implementation and its intended use. Subsequent phantom testing measured how TrueFidelity affected noise power spectra, spatial resolution, and task-based detectability over different conditions, supplying an independent comparison with iterative reconstruction.

Notable2021IEEE Transactions on Medical Imaging184 citations

Deep Sinogram Completion With Image Prior for Metal Artifact Reduction in CT Images

Lequan Yu, Zhicheng Zhang … Lei Xing

The method produced an image prior and projected it forward differentiably to guide a second network that repaired metal-corrupted sinogram values. Filtered back-projection yielded the final image. Experiments with simulated and real artifacts examined projection continuity around metal traces and reported reduced artifacts relative to the comparison methods.

SignificanceThe paper showed how an estimated image could guide a projection-space correction and how that correction could be tested on both simulated and real artifacts. The differentiable forward operator allowed the prior-image and sinogram networks to be optimized jointly.