ramp filter
斜坡滤波器
The filter applied to each projection in filtered back-projection. Its gain rises in proportion to spatial frequency, , cancelling the blur of back-projection and amplifying high-frequency noise along the way.
Projections sample Fourier space along radial lines, so the center (low frequencies) is crowded and the periphery (high frequencies) sparse. Back-projection inherits that imbalance and comes out blurred. The ramp filter re-weights each frequency by , a straight ramp when plotted, to even it out. In image space it is a kernel with a central positive peak and small negative side lobes, so neighboring rays partly cancel.
A pure ramp rises forever, so in practice it is cut off at the Nyquist–Shannon sampling theorem frequency: the Ram-Lak filter of 1971. The filter Larry Shepp and Benjamin Logan proposed in 1974 multiplies the ramp by a smooth window so that it rolls off before the cut-off.
Most noise lives at high frequencies, so the steeper the ramp, the grainier the image. The soft-tissue, bone and lung kernels on a CT console are, at heart, ramps tapered by different amounts: a sharp kernel shows fine trabecular bone and covers soft tissue in grain; a smooth kernel does the opposite. It is the most direct trade-off between spatial resolution and noise in CT.
- Ramachandran GN, Lakshminarayanan AV. Three-dimensional reconstruction from radiographs and electron micrographs: application of convolutions instead of Fourier transforms. PNAS. 1971;68(9):2236–2240. doi.org/10.1073/pnas.68.9.2236
- Shepp LA, Logan BF. The Fourier reconstruction of a head section. IEEE Trans Nucl Sci. 1974;21(3):21–43. doi.org/10.1109/TNS.1974.6499235
Key works · 1
Landmark1971Proceedings of the National Academy of Sciences1,192 citations
Three-dimensional Reconstruction from Radiographs and Electron Micrographs: Application of Convolutions instead of Fourier Transforms
The paper replaced direct two-dimensional Fourier computation in projection reconstruction with a one-dimensional convolution on each projection. The authors assembled the filtered data from different angles into sections and reported faster computation and greater accuracy in their tests using simulated projections.
SignificanceThe formulation made the filtering stage of CT reconstruction practical with a spatial-domain kernel. The IAEA handbook describes convolution followed by back-projection as a classical CT method and identifies the ramp filter by the authors' names, Ram–Lak.
DOI Full text G. N. RamachandranA. V. Lakshminarayanan filtered back-projectionconvolutionramp filter