Landmark1971Proceedings of the National Academy of Sciences1,192 citations
Three-dimensional Reconstruction from Radiographs and Electron Micrographs: Application of Convolutions instead of Fourier Transforms
The paper replaced direct two-dimensional Fourier computation in projection reconstruction with a one-dimensional convolution on each projection. The authors assembled the filtered data from different angles into sections and reported faster computation and greater accuracy in their tests using simulated projections.
SignificanceThe formulation made the filtering stage of CT reconstruction practical with a spatial-domain kernel. The IAEA handbook describes convolution followed by back-projection as a classical CT method and identifies the ramp filter by the authors' names, Ram–Lak.
DOI Full text G. N. RamachandranA. V. Lakshminarayanan filtered back-projectionconvolutionramp filter