FDA 510(k) clearance
FDA 510(k)许可
The most common route to the US market for a medical device: show that the new product is substantially equivalent to one already legally sold, its predicate, and the FDA clears it. Most AI imaging software has reached the market by this route.
The name refers to section 510(k) of the US Federal Food, Drug, and Cosmetic Act, put in place by the Medical Device Amendments of 1976. At least 90 days before selling a device, the maker submits evidence that it is substantially equivalent to a legally marketed predicate: the same intended use, and no technological difference that raises new questions of safety or effectiveness. A successful submission is cleared, not approved; approval is reserved for the premarket approval (PMA) route for high-risk devices.
Most AI software falls into moderate-risk Class II, so this is the busiest route. In January 2017 Arterys Cardio DL, which analyses cardiac MRI with deep learning, received 510(k) clearance, one of the first deep-learning products to do so. By 2025 the FDA's list of AI-enabled devices had passed 1,000, most of them in radiology and the great majority cleared through 510(k).
Critics note that chains of predicates can drift far from the original device, and that 510(k) rarely requires prospective clinical trials. Truly new kinds of device go through De Novo instead.
- US FDA. Premarket Notification 510(k). www.fda.gov/medical-devices/premarket-submissions-selecting-and-
- US FDA 510(k) database, K163253: Arterys Cardio DL, decision 5 January 2017. www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K163
- US FDA. Artificial Intelligence-Enabled Medical Devices (list). www.fda.gov/medical-devices/artificial-intelligence-enabled-medi
Key works · 4
Notable2019FDA 510(k) Premarket Notification K183202
Deep Learning Image Reconstruction
The 510(k) record covers the CT reconstruction option that produces TrueFidelity images. The manufacturer integrated a neural network into the scanner's raw-data reconstruction chain and evaluated image properties including noise texture. In April 2019 the FDA determined that the option was substantially equivalent to legally marketed devices.
SignificanceThe record provides primary documentation of a commercial scanner implementation and its intended use. Subsequent phantom testing measured how TrueFidelity affected noise power spectra, spatial resolution, and task-based detectability over different conditions, supplying an independent comparison with iterative reconstruction.
Full text deep-learning reconstructionFDA 510(k) clearancefiltered back-projectioncomputed tomography
Notable2020FDA 510(k) Premarket Notification K193282
SIGNA Premier
On April 10, 2020, the FDA found the SIGNA Premier MRI scanner substantially equivalent with the addition of its AIR Recon DL software feature. The submission described phantom tests and blinded radiologist readings that assessed signal-to-noise ratio, sharpness, lesion conspicuity, and image quality at shorter acquisition times.
SignificanceThe clearance documents scanner integration of learned reconstruction in commercial MRI. A subsequent outpatient implementation study included AIR Recon DL among the reconstruction products deployed and measured examination and turnaround times before and after implementation in routine practice.
Full text deep-learning reconstructionFDA 510(k) clearancemagnetic resonance imaging
Notable2025FDA final guidance, August 2025
Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions
The FDA final guidance describes predetermined change control plans for eligible device software submitted through the 510(k), De Novo, and premarket approval pathways. It recommends describing the planned changes, the protocol for developing, validating, and implementing them, and an assessment of their impact.
SignificanceThe guidance explains how the FDA evaluates a change plan within a marketing submission. Once authorized, the plan can cover specified future changes, connecting those software updates to pre-established validation methods and impact assessments while preserving continued regulatory oversight.
FDA 510(k) clearanceFDA De Novo classificationUS Food and Drug Administrationpredetermined change control plan
Notable2026FDA 510(k) K252970
BriefCase-Triage: CARE Multi-triage CT Body
This software uses a locked algorithm fine-tuned from a foundation model. It processes contrast and noncontrast chest, abdominal, or pelvic CT for people aged 18 and older. It flags 11 specified suspected findings for trained readers to prioritize review of the original exam. The FDA cleared the device in January 2026.
SignificanceThe case shows how a pretrained foundation model can supply several specified triage functions in one device. The public clearance record specifies the scans, the listed findings, and the notification role, defining the scope of the application.
Full text computed tomographyUS Food and Drug Administrationfoundation modelFDA 510(k) clearance