Image analysis

computer-aided detection & diagnosisCAD

计算机辅助检测/诊断

Software that marks suspicious findings on a scan (detection, CADe) or estimates how likely they are to be malignant (diagnosis, CADx), acting as a second reader for the radiologist. The first mammography CAD system won FDA approval in 1998.

Classic CAD runs a fixed pipeline: preprocess, find candidate regions, compute hand-crafted features, prune false positives with a classifier, and place marks on the image. From the 1980s Kunio Doi's group at the University of Chicago studied CAD for mammograms and chest radiographs and set out its role: the computer's output is a second opinion, and the physician still makes the diagnosis.

In June 1998 the FDA approved R2 Technology's ImageChecker M1000, the first CAD system for mammography, and its use spread quickly in the United States. A 2007 study in the New England Journal of Medicine, covering 43 facilities, found that after CAD was adopted specificity fell and biopsies increased, with no clear gain in the detection of invasive breast cancer.

Deep learning has improved CAD's accuracy and changed its form. Today's AI products also triage worklists and take measurements, and they are expected to show in a reader study that they actually help the people reading the images.

Sources
  1. Doi K. Computer-aided diagnosis in medical imaging: historical review, current status and future potential. Comput Med Imaging Graph. 2007;31(4–5):198–211. doi.org/10.1016/j.compmedimag.2007.02.002
  2. SPIE Newsroom. Computer-aided diagnosis becomes a reality in mammography (ImageChecker M1000, FDA approval June 1998). spie.org/news/computer-aided-diagnosis-becomes-a-reality-in-mamm
  3. Fenton JJ, Taplin SH, Carney PA, et al. Influence of computer-aided detection on performance of screening mammography. N Engl J Med. 2007;356(14):1399–1409. doi.org/10.1056/NEJMoa066099

Key works · 18

Landmark1987Medical Physics391 citations

Image feature analysis and computer-aided diagnosis in digital radiography. I. Automated detection of microcalcifications in mammography

Heang-Ping Chan, Kunio Doi … Peter M. Jokich

The detector subtracted a signal-suppressed image from a signal-enhanced image to reduce structured mammographic background, then selected candidates using physical characteristics of microcalcifications. With simulated clusters placed on normal backgrounds, it detected roughly 80% of clusters at about one false-positive cluster per image.

SignificanceThe study established a candidate-generation and feature-screening procedure with a measurable sensitivity and false-positive tradeoff. The same research program then used controlled reader studies to determine whether providing such candidate marks improved radiologists' recognition of microcalcification clusters.

Landmark1998U.S. Food and Drug Administration

M1000 IMAGECHECKER

R2 Technology, Inc.

The FDA granted premarket approval to the M1000 ImageChecker in June 1998. The device marked suspicious areas on routine screening mammograms for a radiologist to review after completing an initial interpretation, defining a specific order for the aided reading workflow.

SignificanceThe approval established a regulatory precedent for using computer-generated marks during mammographic screening. Subsequent investigations in community screening centers, registry studies, and equivalence trials evaluated how assisted reading affected detection and recall under different workflows.

Key1963Radiology180 citations

Computer Diagnosis of Primary Bone Tumors: A Preliminary Report

Gwilym S. Lodwick, Cosmo L. Haun … Eddie D. Robertson

The preliminary report explored how a computer could support differential diagnosis of bone tumors on radiographs. It described the requirements for encoding observations as diagnostic features and combining them with statistical weights so that the program could evaluate possible disease categories from the recorded findings.

SignificanceThe work documented an early attempt to make radiographic reasoning computational by structuring observed signs. The historical review by Doi lists it among early studies of computerized image analysis, establishing its place in the development of feature-based diagnostic methods.

Key1990Investigative Radiology409 citations

Improvement in Radiologists' Detection of Clustered Microcalcifications on Mammograms: The Potential of Computer-Aided Diagnosis

Heang-Ping Chan, Kunio Doi … Kwok Leung Lam

The authors tested whether displaying computer-generated marks improved radiologists' detection of clustered microcalcifications. They compared aided and unaided readings in a controlled setting intended to resemble rapid screening interpretation, finding significantly better detection accuracy with the computerized aid by ROC analysis.

SignificanceThe study evaluated the performance of radiologists using the detector outputs. The history published by Medical Physics identifies it as an early controlled comparison of aided and unaided readings, an approach that preceded evaluations of recall and cancer detection in practice.

Key1990RadioGraphics117 citations

Pulmonary nodules: computer-aided detection in digital chest images.

M L Giger, K Doi … F F Yin

The method generated candidate nodules in digital chest images using a difference image and features such as growth, slope, and intensity profiles. The investigators showed the computer marks to 12 radiologists reading 60 cases. Their preliminary results indicated that the marks could improve detection performance.

SignificanceThe paper paired candidate generation for chest radiographs with an assessment of whether the marks helped readers. The later medical physics review situates these experiments in the development of thoracic detection methods and documents related evaluations on chest CT.

Key2001Radiology839 citations

Screening Mammography with Computer-aided Detection: Prospective Study of 12,860 Patients in a Community Breast Center

Timothy W. Freer, Michael J. Ulissey

The community breast center prospectively recorded 12,860 screening examinations. Radiologists first interpreted each examination unaided and then reviewed marked locations. The study reported more cancers detected and a higher recall rate after the second step, with the positive predictive value of biopsy unchanged.

SignificanceThis study supplied prospective evidence on adding marks to a community screening workflow and quantified the associated increase in recall. Subsequent large registry studies used different designs to assess whether such changes generalized across facilities and readers.

Key2005Gastroenterology296 citations

Computed Tomographic Virtual Colonoscopy Computer-Aided Polyp Detection in a Screening Population

Ronald M. Summers, Jianhua Yao … J. Richard Choi

The study evaluated a computerized polyp detector on CT colonography from a screening population, using optical colonoscopy findings as the reference. The detector generated candidates from local geometry and classified them, with performance reported at both the polyp and patient levels.

SignificanceThe screening cohort provided evidence on candidate detection outside a small enriched case set. Follow-up investigations compared distinct detection systems and assessed radiologists with and without computer marks, clarifying the difference between standalone detection performance and reader benefit.

Key2007Computerized Medical Imaging and Graphics2,221 citations

Computer-aided diagnosis in medical imaging: Historical review, current status and future potential

Kunio Doi

The review traced computer-aided diagnosis from early radiographic classification to prototype systems for mammography, chest radiography, and CT. It examined candidate detection, characterization, and reader studies, emphasizing the use of computer output as a second opinion within radiologists' interpretation workflow.

SignificanceThe review established a historical reference for the distinct roles of algorithm performance, reader assistance, and clinical evaluation. The later Medical Physics history similarly organizes research around how machine outputs are developed and tested in breast and chest imaging.

Key2007New England Journal of Medicine699 citations

Influence of Computer-Aided Detection on Performance of Screening Mammography

Joshua J. Fenton, Stephen H. Taplin … Carl D'Orsi

The registry study assessed 429,345 screening mammograms from 43 facilities. At the seven facilities that implemented computer-aided detection during the study, specificity fell from 90.2% to 87.2% and the biopsy rate increased by 19.7%. The rise in sensitivity was statistically nonsignificant.

SignificanceThe study measured how assisted reading was associated with specificity and downstream biopsies after adoption in routine care. The subsequent analysis of digital screening revisited performance under a different imaging workflow, extending work beyond controlled reader experiments.

Key2008New England Journal of Medicine351 citations

Single Reading with Computer-Aided Detection for Screening Mammography

Fiona J. Gilbert, Susan M. Astley … Jonathan James

A multicenter equivalence trial in England compared double reading of screening film mammograms with single reading aided by computer detection. In paired comparisons, the workflows detected 87.7% and 87.2% of cancers, respectively. The assisted single-reading workflow had a slightly higher recall rate.

SignificanceThe trial tested a concrete workflow question about replacing a second human read in a film-screening program. It measured recall alongside cancer detection, providing comparative evidence on staffing and screening outcomes under specified reading conditions.

Key2015JAMA Internal Medicine815 citations

Diagnostic Accuracy of Digital Screening Mammography With and Without Computer-Aided Detection

Constance D. Lehman, Robert D. Wellman … for the Breast Cancer Surveillance Consortium

The study compared digital mammograms read with and without detection assistance across 66 community facilities. Registry linkage supplied cancer outcomes over the following year. In roughly 626,000 examinations, the study found no overall accuracy improvement with computer assistance across its performance measures.

SignificanceThis large registry analysis extended clinical evaluation of conventional detection assistance to digital screening. The study included hundreds of readers and accounted for clustering by radiologist, providing evidence on how detector marks performed in routine community practice.

Key2016IEEE Transactions on Medical Imaging6,568 citations

Deep Convolutional Neural Networks for Computer-Aided Detection: CNN Architectures, Dataset Characteristics and Transfer Learning

Hoo-Chang Shin, Holger R. Roth … Isabella Nogues

The authors varied network architecture, dataset size, spatial context, and ImageNet initialization for thoracoabdominal lymph node detection and interstitial lung disease classification on CT. Fine-tuning was effective in the tasks studied. The paper also reported fivefold cross-validation for classifying axial CT slices by interstitial lung disease category.

SignificanceBy varying initialization alongside architecture and training data, this widely cited study showed how to judge transfer alongside other design choices in CT detection and classification. Transfusion later used larger classification datasets and lightweight architectures to test when transfer helped and investigate the source of those gains.

Key2017Scientific Data1,113 citations

A curated mammography data set for use in computer-aided detection and diagnosis research

Rebecca Sawyer Lee, Francisco Gimenez … Daniel L. Rubin

CBIS-DDSM curated a subset of a digitized screen-film mammography collection. The release provided decompressed images, updated mass segmentations and bounding boxes, and pathology information. It included 753 calcification cases and 891 mass cases with documented training and test partitions for evaluation.

SignificanceDocumenting case selection, lesion annotations, and evaluation splits makes it possible to compare mammographic detection and classification methods on a common collection. TCIA maintains the release and provides images and metadata tables that identify the training and test subsets.

Key2017Medical Image Analysis1,681 citations

Validation, comparison, and combination of algorithms for automatic detection of pulmonary nodules in computed tomography images: The LUNA16 challenge

Arnaud Arindra Adiyoso Setio, Alberto Traverso … Piergiorgio Cerello

LUNA16 evaluated complete nodule detection and false-positive reduction on a common subset of 888 LIDC-IDRI CT scans. It compared submissions using shared folds and false-positive criteria. Combining several systems yielded sensitivity above 95% at fewer than one false positive per scan.

SignificanceThe two tracks separated the effects of generating candidates and classifying them. By publishing the scan splits, nodule references, and scoring procedure, LUNA16 provided a reproducible comparison for complete systems and specific stages of CT nodule detection.

Key2019Gut1,021 citations

Real-time automatic detection system increases colonoscopic polyp and adenoma detection rates: a prospective randomised controlled study

Pu Wang, Tyler M. Berzin … Xun Xiao

This prospective randomized trial compared standard colonoscopy with a workflow that issued real-time prompts for suspected polyps. Among 1,058 participants, adenoma detection was 29.1% in the assisted arm and 20.3% in the control arm, with the increase concentrated in diminutive adenomas.

SignificanceThe trial evaluated how live detection prompts affected a procedural endpoint during colonoscopy. Reporting the contribution of small and larger adenomas also clarified which lesions accounted for the overall change, an important distinction when interpreting the clinical value of extra detections.

Notable1994Medical Physics190 citations

Computerized detection of clustered microcalcifications in digital mammograms using a shift‐invariant artificial neural network

Wei Zhang, Kunio Doi … Robert A. Schmidt

The authors trained a locally connected network with shared weights to reclassify microcalcification candidates. In holdout experiments on 168 candidate regions from 34 mammograms, the method removed approximately 55% of the false-positive regions without eliminating any true-positive regions in that sample.

SignificanceThe study is an early medical imaging example of local shared weights trained by backpropagation. Its placement after a candidate generator illustrates how a learned classifier could address a specific source of false-positive detections within an existing screening pipeline.

Notable1995IEEE Transactions on Medical Imaging424 citations

Artificial convolution neural network techniques and applications for lung nodule detection

S.-C.B. Lo, S.-L.A. Lou … S.K. Mun

The system first used a template-matching search to generate possible lung nodules and then classified candidate image blocks with a convolutional network. The network learned two-dimensional convolution kernels through backpropagation. The paper tested the pipeline on lung nodule detection and reported its processing time.

SignificanceThe study documents an early use of learned convolution kernels inside a medical image detector. Template matching generated a sensitive first pass and the network classified a smaller candidate set, a division of work examined in later computer-aided detection research.

Notable1996IEEE Transactions on Medical Imaging471 citations

Classification of mass and normal breast tissue: a convolution neural network classifier with spatial domain and texture images

B. Sahiner, Heang-Ping Chan … M.M. Goodsitt

The authors used a convolutional network to classify manually selected mammogram regions as masses or normal tissue. They compared how downsampled intensities and images of local texture features worked as inputs. In a study of 168 confirmed mass regions and 504 normal regions, the best test AUC was 0.87.

SignificanceThe study documented early convolutional classification of mammographic masses and tested how engineered texture measurements could be used as network inputs. Later work on large-scale mammographic lesion detection cited the study as an early application of convolutional networks to mammography.