reader study
读片研究
An experiment in which several doctors read the same set of cases under controlled conditions, so their accuracy alone, with AI assistance, and against the AI can be compared. It is the usual way imaging AI demonstrates clinical value.
The standard design is multi-reader, multi-case (MRMC): every radiologist reads every case, usually once with and once without AI assistance, with a washout of several weeks between sessions, and the results are compared by ROC analysis. Readers and cases are both treated as random samples; the method published by Donald Dorfman, Kevin Berbaum, and Charles Metz in 1992 lets conclusions generalize to other doctors and other patients at once. The FDA routinely asks for such studies when reviewing computer-aided detection devices.
In Google's 2020 Nature study of breast screening, the AI system was compared with six radiologists, and its AUC exceeded the average reader's by an absolute 11.5 percentage points.
The limitation is that reader studies happen in a laboratory. Case sets are enriched with cancers, readers know that no patient depends on their call, and they behave differently from a routine working day. A reader study shows that something can work; it does not replace a prospective clinical trial.
- McKinney SM, Sieniek M, Godbole V, et al. International evaluation of an AI system for breast cancer screening. Nature. 2020;577(7788):89–94. doi.org/10.1038/s41586-019-1799-6
- Dorfman DD, Berbaum KS, Metz CE. Receiver operating characteristic rating analysis: generalization to the population of readers and patients with the jackknife method. Invest Radiol. 1992;27(9):723–731.
Key works · 36
Landmark1982Radiology24,806 citations
The meaning and use of the area under a receiver operating characteristic (ROC) curve.
The paper interpreted the area under an ROC curve from graded readings or continuous scores as the probability that a randomly selected affected case ranks above an unaffected case. Relating the probability to the Wilcoxon statistic yielded practical approximations for standard errors and study sample sizes.
SignificanceThe interpretation gave reader studies a common meaning for a summary across rating thresholds and a way to plan precision. Later clinical evaluation guidance places that summary alongside threshold-specific performance, which is needed when a scoring system will be used at a defined operating point.
DOI James A. HanleyBarbara J. McNeil ROC curve and AUCsensitivity and specificityreader study
Landmark1988Biometrics25,065 citations
Comparing the Areas under Two or More Correlated Receiver Operating Characteristic Curves: A Nonparametric Approach
Diagnostic tests evaluated on the same participants yield correlated estimates of ROC area. DeLong and colleagues used generalized U-statistics to estimate the covariance matrix of the empirical areas. The resulting nonparametric method compares two or more areas while accounting for the paired study design.
SignificanceThe method is a widely used statistical baseline for comparing tests or readers on shared cases, with implementations in statistical software. Studies of chest radiograph classifier generalization have used it to assess differences in ROC area across their planned test comparisons.
DOI Elizabeth R. DeLongDaniel L. Clarke-Pearson ROC curve and AUCreader study
Landmark2017Nature18,648 citations
Dermatologist-level classification of skin cancer with deep neural networks
A network initialized on ImageNet was fine-tuned on 129,450 clinical images. The team compared its predictions with those of 21 dermatologists on biopsy-confirmed images in two tasks, including melanoma versus benign nevi, and reported comparable discrimination under the specified image-reading conditions.
SignificanceThe reader comparison became a widely cited reference for image-based skin lesion classification. The later DDI study carried the biopsy-confirmed evaluation approach into a more diverse setting and showed how performance of dermatology classifiers can vary by skin tone and disease frequency.
DOI Full text Andre EstevaBrett KuprelRoberto A. NovoaSebastian Thrun dermoscopyimage classificationreader study
Landmark2017JAMA4,056 citations
Diagnostic Assessment of Deep Learning Algorithms for Detection of Lymph Node Metastases in Women With Breast Cancer
CAMELYON16 compared algorithms for finding breast cancer metastases in lymph-node whole-slide images. On 129 held-out slides, the best algorithm achieved a slide-level AUC of 0.994. The comparison included eleven time-constrained pathologists and a separate pathologist who read without a time limit.
SignificanceThe evaluation provided a shared task for whole-slide metastasis detection and showed how the reader comparison depended on the conditions of interpretation. CAMELYON17 extended the benchmark from finding metastases on individual slides to estimating nodal status at the patient level.
DOI Full text Babak Ehteshami BejnordiJeroen A. W. M. van der Laak CAMELYONwhole-slide imagereader studyROC curve and AUC
Landmark2018Nature Medicine3,120 citations
Clinically applicable deep learning for diagnosis and referral in retinal disease
The system segments tissues in three-dimensional retinal OCT and uses the tissue maps to predict diagnoses and referral urgency. The retrospective evaluation compares referral decisions with expert judgments. For a second scanner type in the same hospital setting, the authors retrained the segmentation network and tested referral performance with the classifier unchanged.
SignificanceThe intermediate tissue maps made it possible to examine what retinal structures the system detected before assigning a referral decision. By testing a change of scanner at this interface, the study also gave a concrete example of examining transfer across OCT devices.
DOI Full text Jeffrey De FauwPearse Andrew KeaneOlaf Ronneberger optical coherence tomographyimage segmentationreader studydomain shift
Landmark2018The American Journal of Surgical Pathology457 citations
Whole Slide Imaging Versus Microscopy for Primary Diagnosis in Surgical Pathology: A Multicenter Blinded Randomized Noninferiority Study of 1992 Cases (Pivotal Study)
In this blinded randomized study, 16 pathologists at four institutions reviewed slides by digital whole-slide imaging and light microscopy, using the original diagnosis as the reference. In 1,992 cases, the major discordance rates were 4.9% and 4.6%, respectively, meeting the prespecified noninferiority criterion.
SignificanceThe FDA cited this pivotal study when it granted Philips IntelliSite the first US marketing authorization for a whole-slide imaging system for primary surgical pathology diagnosis. The study linked authorization to a prespecified diagnostic endpoint across a broad surgical case mix.
DOI Full text Sanjay MukhopadhyayClive R. Taylor digital pathologywhole-slide imageFDA De Novo classificationreader studynoninferiority trial
Landmark2019Nature Medicine2,876 citations
End-to-end lung cancer screening with three-dimensional deep learning on low-dose chest computed tomography
A volumetric model estimates lung cancer risk from low-dose chest CT, incorporating a prior scan when available. It achieved an AUC of 0.944 on held-out National Lung Screening Trial cases. The study also evaluated 1,139 independent clinical cases and compared readers with and without prior imaging.
SignificanceThe study provided an early end-to-end benchmark for lung screening from volumetric CT and made the availability of prior imaging explicit in the reader comparison. Sybil cites this work and extends the question to multiyear risk estimation from a single scan.
DOI Diego ArdilaDaniel TseShravya Shetty low-dose CTROC curve and AUCreader studyexternal validation
Landmark2020Nature4,409 citations
International evaluation of an AI system for breast cancer screening
The authors evaluated a mammography screening model retrospectively in UK and US cohorts and ran a separate reader study with six radiologists. They reported false-positive and false-negative performance by cohort and simulated how the model might interact with the UK double-reading workflow.
SignificanceThe use of separate national cohorts and a reader study provided evidence for evaluating mammography models under different screening conditions. The later methodological addendum made more of the optimization and augmentation setup available for assessing reproducibility of the retrospective results.
DOI Scott Mayer McKinneyDaniel TseShravya Shetty mammographyreader studyexternal validationROC curve and AUC
Key1983Radiology7,586 citations
A Method of Comparing the Areas under Receiver Operating Characteristic Curves Derived from the Same Cases
The authors compared ROC areas estimated from the same patients under two reading conditions. Using a binormal assumption, they derived a table that converts correlation between paired ratings into correlation between areas. Accounting for that correlation in the standard error of the area difference can increase statistical power.
SignificanceThis paper provided a parametric method for accounting for shared cases in reader or test comparisons. A later methodological study of sample size and power still used the Hanley–McNeil calculation to estimate the variance of differences between correlated ROC areas for continuous diagnostic scores.
DOI Full text James A. HanleyBarbara J. McNeil ROC curve and AUCreader study
Key1989Medical Physics221 citations
Maximum likelihood analysis of free-response receiver operating characteristic (FROC) data
For reader studies that require localization, the paper modeled the scores of signal marks and the production of false-positive marks in free-response ROC data. FROCFIT estimated model parameters and their uncertainties with maximum likelihood. The author reported fits to several experimental data sets.
SignificanceThe paper supplied a statistical model for experiments in which readers can mark several locations per image. Free-response analysis later informed reader-based imaging benchmarks, including the low-dose CT challenge that used a jackknife alternative for evaluating marks of liver metastases.
DOI Dev P. Chakraborty object detectionreader studyROC curve and AUCfree-response ROC
Key1990Investigative Radiology409 citations
Improvement in Radiologists' Detection of Clustered Microcalcifications on Mammograms: The Potential of Computer-Aided Diagnosis
The authors tested whether displaying computer-generated marks improved radiologists' detection of clustered microcalcifications. They compared aided and unaided readings in a controlled setting intended to resemble rapid screening interpretation, finding significantly better detection accuracy with the computerized aid by ROC analysis.
SignificanceThe study evaluated the performance of radiologists using the detector outputs. The history published by Medical Physics identifies it as an early controlled comparison of aided and unaided readings, an approach that preceded evaluations of recall and cancer detection in practice.
DOI Heang‐Ping ChanHeber MacMahon computer-aided detection & diagnosismammographyreader studyROC curve and AUC
Key1990RadioGraphics117 citations
Pulmonary nodules: computer-aided detection in digital chest images.
The method generated candidate nodules in digital chest images using a difference image and features such as growth, slope, and intensity profiles. The investigators showed the computer marks to 12 radiologists reading 60 cases. Their preliminary results indicated that the marks could improve detection performance.
SignificanceThe paper paired candidate generation for chest radiographs with an assessment of whether the marks helped readers. The later medical physics review situates these experiments in the development of thoracic detection methods and documents related evaluations on chest CT.
DOI Maryellen L. GigerF.-F. Yin computer-aided detection & diagnosisradiographyreader study
Key2008New England Journal of Medicine351 citations
Single Reading with Computer-Aided Detection for Screening Mammography
A multicenter equivalence trial in England compared double reading of screening film mammograms with single reading aided by computer detection. In paired comparisons, the workflows detected 87.7% and 87.2% of cancers, respectively. The assisted single-reading workflow had a slightly higher recall rate.
SignificanceThe trial tested a concrete workflow question about replacing a second human read in a film-screening program. It measured recall alongside cancer detection, providing comparative evidence on staffing and screening outcomes under specified reading conditions.
DOI Fiona J. GilbertS.W. Duffy computer-aided detection & diagnosismammographysensitivity and specificityreader study
Key2017arXiv5,269 citations
CheXNet: Radiologist-Level Pneumonia Detection on Chest X-Rays with Deep Learning
CheXNet uses a 121-layer convolutional network trained on ChestX-ray14 to detect pneumonia in frontal chest radiographs. On a set of 420 images annotated by four radiologists, the authors report a higher pneumonia F1 score than the average reader obtained under the study conditions.
SignificanceThe preprint supplied an early reader comparison for a chest radiograph classifier trained with large-scale report-derived labels. The later CheXNeXt study changed the reader and reference-standard setup and reported separate comparisons for a wider set of chest findings.
DOI arXiv Pranav RajpurkarAndrew Y. Ng ChestX-ray14reader studyimage classification
Key2017Medical Physics425 citations
Low-dose CT for the detection and classification of metastatic liver lesions: Results of the 2016 Low Dose CT Grand Challenge
The Low Dose CT Grand Challenge supplied common contrast-enhanced CT cases to groups testing image-space denoising or iterative reconstruction. Participants reconstructed twenty quarter-dose test examinations. Eleven blinded readers marked liver metastases, with substantial differences in reader scores across the submitted reconstruction and denoising methods.
SignificanceThe competition tied reconstruction evaluation to the clinical task of detecting liver metastases. The Cancer Imaging Archive's larger collection includes 13 of the challenge test cases. A separate release provides all 30 training and test cases in the same standardized projection format for further comparisons.
DOI Full text Cynthia H. McColloughJoel G. Fletcher low-dose CTreader studygrand challengefree-response ROC
Key2018PLOS Medicine1,844 citations
Deep learning for chest radiograph diagnosis: A retrospective comparison of the CheXNeXt algorithm to practicing radiologists
CheXNeXt was evaluated on 420 frontal chest radiographs for 14 findings, using majority judgments from three cardiothoracic radiologists as the reference. In the retrospective comparison with nine additional readers, the authors classified performance as comparable for ten findings, higher for one, and lower for three.
SignificanceThe journal study expanded the earlier pneumonia comparison into a per-finding reader evaluation with a separately defined reference panel. Its breakdown makes clear which findings drove the overall agreement and provides a bounded reference for later work on external clinical populations.
DOI Full text Pranav RajpurkarMatthew P. Lungren ChestX-ray14reader studyground truthROC curve and AUC
Key2020Magnetic Resonance in Medicine312 citations
Advancing machine learning for MR image reconstruction with an open competition: Overview of the 2019 fastMRI challenge
The paper reports the single-coil and multicoil tracks of the 2019 fastMRI knee reconstruction challenge. Submissions were first screened by structural similarity and shortlisted reconstructions then went to radiologists for review. Examples included a small osteophyte and a meniscal tear that could be obscured in accelerated reconstructions.
SignificanceThe competition turned the data release into a comparison across teams and exposed the limits of measuring reconstruction quality by image similarity alone. The 2020 brain challenge continued the use of numerical screening and reader review while adding a transfer track and a review focused on pathology.
DOI arXiv Florian KnollMichael P. Recht fastMRIgrand challengereader studydeep-learning reconstructionstructural similarity index
Key2020BMJ1,464 citations
Artificial intelligence versus clinicians: systematic review of design, reporting standards, and claims of deep learning studies
The systematic review examined study design, reporting, and risk of bias for comparisons of diagnostic deep learning with expert clinicians. Of 81 nonrandomized studies, nine were prospective and six tested systems in clinical settings. The authors also searched for registered randomized trials.
SignificanceThe review quantified the gap between retrospective expert comparisons and tests of clinical use, and examined whether published claims were supported by study design. Its separation of study designs provides a reference for assessing how subsequent imaging evaluation has changed.
DOI Full text Myura NagendranMahiben Maruthappu reader studyexternal validationbias and fairness
Key2020American Journal of Roentgenology188 citations
Using Deep Learning to Accelerate Knee MRI at 3 T: Results of an Interchangeability Study
Six readers compared standard 3 T knee MRI with images reconstructed from retrospectively undersampled data. Undersampling simulated a 3.49-fold speedup over fully sampled acquisition and a 1.88-fold speedup over standard parallel acquisition. The images met the study's interchangeability criterion for internal derangement of the knee, and readers favored the accelerated images for image quality.
SignificanceThe study evaluated whether reconstructed scans could substitute for standard scans on a specified diagnostic task. Its reader-based interchangeability design supplied clinical evidence that complemented fastMRI's image similarity metrics and defined the setting in which the replacement was tested.
DOI Full text Michael P. RechtC. Lawrence Zitnick reader studymagnetic resonance imagingdeep-learning reconstruction
Key2021IEEE Transactions on Medical Imaging485 citations
Results of the 2020 fastMRI Challenge for Machine Learning MR Image Reconstruction
The second fastMRI challenge used raw clinical brain MRI and included a transfer track for scanners outside the training set. Radiologists reviewed pathological findings, complementing quantitative image similarity scores. Some high-scoring reconstructions contained a false vessel or a bright line resembling a cerebrospinal fluid cleft.
SignificanceThe challenge incorporated scanner transfer and preservation of pathology into a public benchmark. Its documented failures give concrete examples for understanding when similarity scores and clinical judgments diverge, and for designing evaluations that examine individual structures in reconstructed images.
DOI arXiv Full text Matthew J. MuckleyFlorian Knoll fastMRIhallucinationgrand challengereader studystructural similarity index
Key2022Archives of Pathology & Laboratory Medicine272 citations
Validating Whole Slide Imaging Systems for Diagnostic Purposes in Pathology: Guideline Update From the College of American Pathologists in Collaboration With the American Society for Clinical Pathology and the Association for Pathology Informatics
The expert panel reviewed studies published after the original 2013 guideline and updated the validation recommendations for diagnostic whole-slide imaging. It reaffirmed testing at least 60 cases, assessing within-observer concordance between digital and glass slides, and using a washout of at least two weeks.
SignificanceThe active CAP guideline makes local concordance testing operational and specifies how to handle disagreements between reading modalities. Graded recommendations and practice statements give laboratories a common validation procedure while clarifying which elements are supported by stronger evidence.
DOI Andrew J. EvansNicole Thomas whole-slide imagedigital pathologyreader study
Key2023The Lancet Digital Health487 citations
Artificial intelligence for breast cancer detection in screening mammography in Sweden: a prospective, population-based, paired-reader, non-inferiority study
ScreenTrustCAD enrolled 55,581 women in a Swedish screening program and compared two reading strategies on the same mammograms. One radiologist and software reading independently detected 261 cancers within three months, compared with 250 detected by two radiologists, meeting the prespecified noninferiority criterion.
SignificanceThe paired design tested software as a replacement for one independent reader within a routine screening program. The trial hospital subsequently put this reading arrangement into clinical use, and a follow-up implementation study documented the risks and mitigation measures identified by staff.
DOI Full text Karin DembrowerFredrik Strand mammographyreader studynoninferiority trialprospective validation
Key2023The Lancet Oncology807 citations
Artificial intelligence-supported screen reading versus standard double reading in the Mammography Screening with Artificial Intelligence trial (MASAI): a clinical safety analysis of a randomised, controlled, non-inferiority, single-blinded, screening accuracy study
MASAI randomly assigned screening participants to software-supported mammography reading or standard double reading. The early clinical safety analysis involved about 80,000 participants. Detection rates were 6.1 and 5.1 per 1,000 screened, respectively, and the supported workflow reduced screen-reading workload by 44.3%.
SignificanceThis interim safety analysis counted screen readings alongside cancer detections in an organized screening program. Later reports from the same randomized trial examined screen-detected cancers in the full cohort and evaluated the primary interval-cancer endpoint after follow-up.
DOI Kristina LångAldana Rosso mammographytriagereader studyprospective validation
Key2023Radiology527 citations
Automation Bias in Mammography: The Impact of Artificial Intelligence BI-RADS Suggestions on Reader Performance
Twenty-seven radiologists graded fifty mammograms with BI-RADS suggestions attributed to an automated system. Some suggestions were deliberately incorrect. Correct grading declined across all experience groups when the suggestions were wrong, and less experienced readers followed certain incorrect upward suggestions more often.
SignificanceBy controlling whether the suggested category was correct, the study could directly measure how readers responded when assistance misled them. It provided a reader-study design for evaluating automation bias across experience levels in mammography interpretation.
DOI Thomas DratschDaniel Pinto dos Santos reader studymammographyradiologistautomation bias
Key2023Radiology214 citations
Methods for Clinical Evaluation of Artificial Intelligence Algorithms for Medical Diagnosis
The review covers external testing, performance measures, and reporting for diagnostic algorithms. It discusses paired designs for comparing interpretations with and without assistance on the same cases and parallel designs for evaluating the effect of an intervention. Radiology examples illustrate reader order, washout, and randomized trial design.
SignificanceThe review gives readers a common reference for linking each clinical question to an appropriate design. Its audit of two radiology journals found paired designs in 22 of 24 studies comparing assisted and unassisted diagnosis, illustrating how the published evidence was produced.
DOI Seong Ho ParkJaesoon Choi external validationreader studyROC curve and AUC
Key2024Nature Medicine1,421 citations
Adapted large language models can outperform medical experts in clinical text summarization
The authors adapted eight language models across six datasets on radiology reports, patient questions, progress notes, and patient-clinician dialogue. Ten physicians assessed completeness, correctness, and concision. The best adapted models were favored for completeness and correctness. In the overall comparison, 45% of their summaries were rated equivalent to expert summaries and 36% were rated superior.
SignificanceThe work measured summarization across several sources of clinical text using the same reader framework and checked how automatic metrics agreed with physician judgments. The error analysis made completeness and clinical correctness explicit outcomes for evaluating adapted summarization systems.
DOI arXiv Full text Dave Van VeenCara Van UdenAkshay S. Chaudhari large language modelradiology report generationreader studyclinical natural language processing
Key2024Nature Medicine207 citations
Deep learning-aided decision support for diagnosis of skin disease across skin tones
A simulated teledermatology study enrolled 848 specialist and primary-care physicians to propose differential diagnoses from photographs across skin tones. Model support improved overall diagnostic accuracy for both groups, while increasing the difference in accuracy between lighter- and darker-skin images among generalists.
SignificanceThe experiment evaluates the combined physician and model workflow across reader groups and skin tones. It shows why evaluation of the model alone or of the team's aggregate accuracy cannot substitute for subgroup analysis of the actual decisions made with assistance.
DOI Full text Matthew GrohRosalind Picard bias and fairnessreader studyimage classification
Key2024Nature Medicine295 citations
Heterogeneity and predictors of the effects of AI assistance on radiologists
The study analyzed algorithm-assisted reading by 140 radiologists across fifteen chest radiograph tasks. Experience, subspecialty, and familiarity with such tools did not reliably predict individual effects. In the aggregate analysis and for several individual findings, erroneous model predictions were associated with poorer reader performance.
SignificanceThe analysis separated aggregate reader performance from the effects for individual readers and findings, and examined model correctness explicitly. That design gives clinical evaluations a way to measure heterogeneity of assistance and to identify errors that affect the combined judgment.
DOI Full text Feiyang YuPranav Rajpurkar reader studyradiologistradiography
Key2024JAMA Network Open1,129 citations
Large Language Model Influence on Diagnostic Reasoning: A Randomized Clinical Trial
Fifty physicians were randomized to use GPT-4 alongside conventional resources or to use conventional resources alone when working through clinical vignettes. Blinded graders scored diagnosis and reasoning using a standardized rubric. The study found no statistically significant improvement with access to GPT-4.
SignificanceThe trial made the clinician’s performance the main outcome and measured the model by itself separately. This distinction let researchers test whether access to a strong standalone model improves assisted diagnosis and provided a comparison for a later trial on management reasoning.
DOI Full text Ethan GohJonathan H. Chen large language modelreader study
Key2024IEEE Transactions on Medical Imaging106 citations
MedSyn: Text-Guided Anatomy-Aware Synthesis of High-Fidelity 3-D CT Images
MedSyn trained a hierarchical diffusion model with radiology reports and anatomy channels for lung lobes, airways, and vessels. It generated coarse chest CT volumes and then increased their resolution. In a blinded assessment by ten radiologists, the generated fissures and lobe structures were rated as more anatomically plausible than those of the two generative baselines.
SignificanceThe paper gave text-conditioned chest CT synthesis a reference that tests whether generated anatomy and prompted findings are recognizable to readers. Better Tokens for Better 3D subsequently used MedSyn as a comparator when assessing text alignment and image synthesis in volumetric chest CT.
DOI arXiv Yanwu XuKayhan Batmanghelich diffusion modelsynthetic datacomputed tomographyreader study
Key2025Nature Medicine330 citations
Collaboration between clinicians and vision–language models in radiology report generation
The authors trained Flamingo-CXR to generate chest radiograph reports and convened a radiologist panel to review model and clinician reports retrospectively. For inpatient and outpatient radiographs, clinically significant errors occurred in both reports in 24.8% of cases, only the generated report in 22.8%, and only the clinician report in 14.0%.
SignificanceThe error review also supplied edits for a retrospective proof of concept. Radiologists’ corrections were incorporated into model drafts, and raters compared the edited reports with the original clinical reports. Janus-Pro-CXR subsequently examined report quality and time in a prospective workflow study.
DOI Full text Ryutaro TannoDavid G. T. BarrettAlan KarthikesalingamSofia Ira Ktena vision–language modelradiology report generationreader studyhallucination
Key2025Nature Medicine286 citations
GPT-4 assistance for improvement of physician performance on patient care tasks: a randomized controlled trial
The trial randomized 92 practicing physicians to use GPT-4 with conventional resources or conventional resources alone for five simulated patient-management cases. Investigators scored choices about testing, treatment, and follow-up. Access to GPT-4 improved management reasoning scores, while participants spent more time per case.
SignificanceThe study measured the effect of a language model on clinician management decisions alongside time spent on the task. Together with the earlier diagnostic trial, it supplied separate evidence on collaboration for two different types of clinical reasoning.
DOI Full text Ethan GohJonathan H. ChenAdam M. Rodman large language modelreader study
Key2025Nature1,061 citations
Towards conversational diagnostic artificial intelligence
AMIE used self-play and automated feedback to train a model for diagnostic dialogue. In a randomized, double-blind crossover study, standardized patients completed textual consultations with the system and primary care physicians. The study compared history-taking, diagnosis, and communication across 159 scenarios.
SignificanceThe study assessed diagnostic dialogue as an interaction and obtained ratings from both specialists and standardized patients. Its crossover design provided a reference for testing history-taking and communication as well as diagnostic outcomes in text-based simulated consultations.
DOI arXiv Full text Tao TuAlan KarthikesalingamVivek Natarajan large language modelAI agentreader study
Key2026Nature Medicine16 citations
AI-based triage and decision support in mammography and digital tomosynthesis for breast cancer screening: a paired, noninferiority trial
The prospective paired trial enrolled 31,301 women and compared standard double reading with a workflow that marked low-risk exams normal and sent the rest to software-assisted double reading. Radiologist workload fell by 63.6%, cancer detection rose from 6.3 to 7.3 per 1,000, and recall increased, failing its noninferiority criterion.
SignificanceThe trial quantified how removing low-risk exams from the reading queue affected workload, detection, and recall within one program. Reading reductions were similar for digital mammography and digital breast tomosynthesis, while detection and recall increased in the digital mammography subgroup. This gives readers modality-specific evidence for workflow design.
DOI Full text Esperanza Elías-CabotMarina Álvarez-Benito mammographytomosynthesistriagereader studynoninferiority trialprospective validation
Notable2025Scientific Reports10 citations
Clinical evaluation of medical image synthesis: a case study in wireless capsule endoscopy
The CEMIS protocol was tested on capsule endoscopy images generated with the TIDE-II variational autoencoder. Ten specialists judged single images, pairs, and groups for realism, diversity, or clinical features. In a paired task they identified real images more often than chance, while separate tasks used both real and synthetic images to assess abnormality recognition.
SignificanceThe case study makes clinical reader assessment of synthetic images reproducible across individual, paired, and grouped tasks. The real-image comparator for abnormality judgments gives readers a way to assess preservation of diagnostic features separately from judging whether an image looks authentic.
DOI arXiv Full text Panagiota GatoulaDimitris K. Iakovidis synthetic datareader studyendoscopyautoencoder
Notable2025IMDRF technical document N88
Good machine learning practice for medical device development: Guiding principles
The IMDRF document sets out ten good machine learning practice principles for developing medical devices. They address representative data, independent training and test sets, performance in clinically relevant conditions, information for users, monitoring deployed models, and evaluating the performance of the human-system team.
SignificanceThe document provides shared technical language for regulators and developers, continuing principles previously issued by national regulators. Its lifecycle framing ties deployed model monitoring to the evidence used in design and validation, including how the device is expected to be used.