Key2021IEEE Transactions on Medical Imaging485 citations
Results of the 2020 fastMRI Challenge for Machine Learning MR Image Reconstruction
The second fastMRI challenge used raw clinical brain MRI and included a transfer track for scanners outside the training set. Radiologists reviewed pathological findings, complementing quantitative image similarity scores. Some high-scoring reconstructions contained a false vessel or a bright line resembling a cerebrospinal fluid cleft.
SignificanceThe challenge incorporated scanner transfer and preservation of pathology into a public benchmark. Its documented failures give concrete examples for understanding when similarity scores and clinical judgments diverge, and for designing evaluations that examine individual structures in reconstructed images.
DOI arXiv Full text Matthew J. MuckleyFlorian Knoll fastMRIhallucinationgrand challengereader studystructural similarity index