FDA De Novo classification
FDA De Novo分类
The US route for a novel, low-to-moderate-risk device that has no predicate to compare with. The first AI stroke-triage software and the first autonomous AI diagnostic both came through it, which allowed later products of the same type to use the 510(k) route.
Under the 1976 law, any new device with no predicate was automatically placed in Class III, the highest-risk class, which requires costly premarket approval. The FDA Modernization Act of 1997 created the De Novo route: if a novel device is low or moderate risk, its maker can ask for it to be classified into Class I or II, and the FDA writes a new classification regulation with special controls. Since 2012 sponsors can file a De Novo request directly, without first being turned down under 510(k).
For imaging AI, De Novo is the route by which new device categories are created. On 13 February 2018 Viz.ai's ContaCT became the first radiological computer-aided triage and notification software, flagging suspected large-vessel-occlusion stroke on CT angiography and texting the specialist. On 11 April 2018 IDx-DR became the first autonomous AI diagnostic system, reading retinal photographs without a clinician.
Once granted, a De Novo device can serve as a predicate, so competitors can follow through the faster 510(k) route.
- US FDA. Evaluation of Automatic Class III Designation (De Novo) Summaries. www.fda.gov/about-fda/cdrh-transparency/evaluation-automatic-cla
- US FDA De Novo database, DEN170073: ContaCT (Viz.ai), decision 13 February 2018, 21 CFR 892.2080. www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/denovo.cfm?id=D
- US FDA De Novo database, DEN180001: IDx-DR (IDx LLC), decision 11 April 2018, 21 CFR 886.1100. www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/denovo.cfm?id=D
Key works · 4
Landmark2018npj Digital Medicine2,097 citations
Pivotal trial of an autonomous AI-based diagnostic system for detection of diabetic retinopathy in primary care offices
The pivotal trial enrolled 900 participants with diabetes at ten primary-care sites and compared the device with reading-center grading of widefield stereoscopic photography and OCT. For detection of more-than-mild diabetic retinopathy, the study reported a sensitivity of 87.2% and a specificity of 90.7%.
SignificanceThe trial supported the FDA's 2018 De Novo authorization of IDx-DR for an autonomous screening decision in primary care. The FDA summary reports its own performance analysis and imageability assessment, providing an example of the evidence used for this intended use.
DOI Full text Michael D. AbràmoffJames C. Folk fundus photographysensitivity and specificityFDA De Novo classificationground truthprospective validation
Landmark2018The American Journal of Surgical Pathology457 citations
Whole Slide Imaging Versus Microscopy for Primary Diagnosis in Surgical Pathology: A Multicenter Blinded Randomized Noninferiority Study of 1992 Cases (Pivotal Study)
In this blinded randomized study, 16 pathologists at four institutions reviewed slides by digital whole-slide imaging and light microscopy, using the original diagnosis as the reference. In 1,992 cases, the major discordance rates were 4.9% and 4.6%, respectively, meeting the prespecified noninferiority criterion.
SignificanceThe FDA cited this pivotal study when it granted Philips IntelliSite the first US marketing authorization for a whole-slide imaging system for primary surgical pathology diagnosis. The study linked authorization to a prespecified diagnostic endpoint across a broad surgical case mix.
DOI Full text Sanjay MukhopadhyayClive R. Taylor digital pathologywhole-slide imageFDA De Novo classificationreader studynoninferiority trial
Key2023JAMA Neurology130 citations
Automated Large Vessel Occlusion Detection Software and Thrombectomy Treatment Times: A Cluster Randomized Clinical Trial
Four stroke centers introduced large-vessel occlusion detection on CT angiograms and secure group messaging in stages. The order was randomized for three centers, with the academic center assigned to go last in advance. The intervention shortened the time from hospital arrival to arterial puncture for thrombectomy. The trial also assessed 90-day disability in an exploratory analysis.
SignificanceThe trial moved beyond specialist notification, an endpoint in the FDA's early ContaCT decision, to an observed treatment workflow measure. It provides a randomized evaluation of imaging triage in hospitals, with the intervention combining detection software and team communication.
DOI Juan Carlos Martínez-GutiérrezLuca GiancardoSunil A. Sheth angiographycomputed tomographytriageFDA De Novo classification
Notable2025FDA final guidance, August 2025
Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions
The FDA final guidance describes predetermined change control plans for eligible device software submitted through the 510(k), De Novo, and premarket approval pathways. It recommends describing the planned changes, the protocol for developing, validating, and implementing them, and an assessment of their impact.
SignificanceThe guidance explains how the FDA evaluates a change plan within a marketing submission. Once authorized, the plan can cover specified future changes, connecting those software updates to pre-established validation methods and impact assessments while preserving continued regulatory oversight.
FDA 510(k) clearanceFDA De Novo classificationUS Food and Drug Administrationpredetermined change control plan