Evaluation & clinic · 2018

autonomous AI

自主AI诊断

A system whose output is the clinical decision itself, with no doctor reviewing the images. The first was IDx-DR, authorized in the US in 2018 to screen for diabetic eye disease; the AI moves from assisting a reader to replacing one.

Most imaging AI assists: it marks suspicious regions or reorders a worklist, and a radiologist signs the report. Autonomous AI hands over the whole read. The first example to be authorized was IDx-DR, now called LumineticsCore. A primary-care clinic photographs a diabetic patient's retinas, and the software itself decides: refer to an eye specialist for more-than-mild diabetic retinopathy, or rescreen in 12 months.

It received De Novo authorization on 11 April 2018. The pivotal trial enrolled 900 patients at 10 US primary-care sites and used widefield stereo photographs and OCT graded by an independent reading center as the reference standard: sensitivity was 87.2% and specificity 90.7%.

Radiology's first autonomous product was Oxipit's ChestLink, CE-marked as Class IIb in 2022, which signs off only normal chest X-rays and passes anything abnormal to a doctor. The hard part is accountability: when no doctor checks the result, much of the responsibility for an error shifts to the developer.

Sources
  1. Abràmoff MD, Lavin PT, Birch M, Shah N, Folk JC. Pivotal trial of an autonomous AI-based diagnostic system for detection of diabetic retinopathy in primary care offices. npj Digit Med. 2018;1:39. doi.org/10.1038/s41746-018-0040-6
  2. US FDA De Novo database, DEN180001: IDx-DR, decision 11 April 2018. www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/denovo.cfm?id=D
  3. Oxipit. ChestLink, CE Class IIb autonomous AI for normal chest X-ray reporting (launched April 2022). oxipit.ai/