optical coherence tomographyOCT
光学相干断层成像
An imaging method analogous to ultrasound but using light instead of sound: near-infrared light reflected from different depths in tissue is measured by interferometry, giving cross-sections with micrometre resolution. It lets ophthalmologists see the retina layer by layer and is a routine test in macular disease.
Light travels too fast to time its echoes the way ultrasound imaging does, so OCT uses interference. Light from a broadband, low-coherence source is split between the tissue and a reference mirror; the two beams interfere only when their path lengths match to within the coherence length, which pins down the depth each reflection came from. Axial resolution is set by the source bandwidth and is typically under 10 µm, but light penetrates only 1–2 mm into most tissue.
David Huang, James Fujimoto and colleagues at MIT described the method in Science in 1991. Fourier-domain OCT, in clinical use from around 2006, records tens of thousands of depth scans per second, far faster than the early time-domain machines.
The transparent, layered retina is particularly well suited to OCT. Macular edema and thinning of the nerve-fiber layer in glaucoma can be measured in micrometres, and OCT volumes have become a major training source for ophthalmic AI, including the 2023 foundation model RETFound.
- Huang D, Swanson EA, Lin CP, et al. Optical coherence tomography. Science. 1991;254(5035):1178–1181. doi.org/10.1126/science.1957169
Key works · 3
Landmark2018Nature Medicine3,120 citations
Clinically applicable deep learning for diagnosis and referral in retinal disease
The system segments tissues in three-dimensional retinal OCT and uses the tissue maps to predict diagnoses and referral urgency. The retrospective evaluation compares referral decisions with expert judgments. For a second scanner type in the same hospital setting, the authors retrained the segmentation network and tested referral performance with the classifier unchanged.
SignificanceThe intermediate tissue maps made it possible to examine what retinal structures the system detected before assigning a referral decision. By testing a change of scanner at this interface, the study also gave a concrete example of examining transfer across OCT devices.
DOI Full text Jeffrey De FauwPearse Andrew KeaneOlaf Ronneberger optical coherence tomographyimage segmentationreader studydomain shift
Landmark2023Nature1,458 citations
A foundation model for generalizable disease detection from retinal images
RETFound used masked autoencoding to pretrain on about 1.6 million unlabeled color fundus and optical coherence tomography images. After adaptation with fewer labeled examples, it performed better than several comparison models on diagnosis and prognosis of eye disease and prediction of selected systemic events.
SignificanceRETFound released a domain-specific encoder and paired it with evaluations of label-efficient adaptation to ocular and systemic outcomes. This gave retinal imaging research a shared starting point and a benchmark for testing the transferability of visual features across cohorts and tasks.
DOI Full text Yukun ZhouPearse Andrew Keane foundation modelmasked image modelingself-supervised learningfundus photographyoptical coherence tomography
Notable2025Nature Medicine97 citations
An eyecare foundation model for clinical assistance: a randomized controlled trial
EyeFM was pretrained on 14.5 million ocular images from five modalities paired with clinical text. A double-masked randomized trial at one center in China involved 668 high-risk participants and 16 ophthalmologists. Before consultant review, correct diagnosis and referral rates were both 92.2% with assistance, versus 75.4% and 80.5% with standard care.
SignificanceThe randomized study measured screening and referral decisions made by ophthalmologists. It also found higher short-term adherence to self-management and referral advice in the assisted group. The results establish endpoints for this workflow, with longer-term vision outcomes still open.
DOI foundation modelvision–language modelfundus photographyoptical coherence tomographyprospective validation