tomosynthesis
断层合成
The X-ray tube sweeps through a limited arc, taking a dozen or more low-dose projections that are reconstructed into a stack of parallel slices. Somewhere between a radiograph and computed tomography, it is used mainly in the breast, letting readers scroll slice by slice through tissue that would otherwise overlap.
The detector stays put while the tube moves along an arc of less than 60°, taking a low-dose image at each step. Structures at different depths shift by different amounts between views, so lining the views up for one depth and adding them brings that plane into focus and smears the rest. This is shift-and-add; modern systems use filtered back-projection or iterative reconstruction instead. Because the arc is limited, resolution in depth is far coarser than within each slice.
D. G. Grant named the technique in 1972. Loren Niklason and colleagues at Massachusetts General Hospital applied it to the breast in 1997, and on 11 February 2011 Hologic's Selenia Dimensions became the first breast tomosynthesis system approved by the FDA.
Separating overlapping tissue helps find cancers and avoid recalls in dense breasts. Paired at first with a standard 2D mammogram, it roughly doubled the dose; synthetic 2D images reconstructed from the tomosynthesis data later made the separate exposure unnecessary.
- Grant DG. Tomosynthesis: a three-dimensional radiographic imaging technique. IEEE Trans Biomed Eng. 1972;BME-19(1):20–28. doi.org/10.1109/TBME.1972.324154
- Niklason LT, Christian BT, Niklason LE, et al. Digital tomosynthesis in breast imaging. Radiology. 1997;205(2):399–406. doi.org/10.1148/radiology.205.2.9356620
- FDA. Radiology devices; reclassification of digital breast tomosynthesis system (proposed order). Federal Register, 10 Aug 2026. www.federalregister.gov/documents/2026/08/10/2026-16209/radiolog
Key works · 5
Landmark1997Radiology846 citations
Digital Tomosynthesis in Breast Imaging
The investigators acquired multiple low-dose projections on a full-field digital mammography system as the X-ray source moved through an arc. They reconstructed sections from a single arc and tested them on phantoms and mastectomy specimens. Readers generally preferred the sections for lesion visualization at a dose comparable to single-view mammography.
SignificanceThe work demonstrated that a digital breast system could acquire the limited-angle data needed to reduce tissue superposition. The same acquisition principle became the basis of clinical breast tomosynthesis and was subsequently evaluated against conventional digital mammography in screening studies.
DOI Loren T. NiklasonDaniel B. KopansR. F. Wirth tomosynthesismammography
Key2014JAMA927 citations
Breast Cancer Screening Using Tomosynthesis in Combination With Digital Mammography
The investigators retrospectively compared 454,850 examinations across 13 centers before and after adding tomosynthesis to digital mammography. In adjusted estimates per 1,000 screens, recall declined from 107 to 91 and cancer detection increased from 4.2 to 5.4 after the addition.
SignificanceThe study estimated changes in recall and detection across a range of screening practices after clinical adoption. It was included in the Agency for Healthcare Research and Quality evidence review and supplied multicenter data on outcomes later evaluated in prospective trials.
DOI Sarah M. FriedewaldEmily F. Conant mammographytomosynthesis
Key2022The Lancet Oncology139 citations
Digital breast tomosynthesis plus synthesised mammography versus digital screening mammography for the detection of invasive breast cancer (TOSYMA): a multicentre, open-label, randomised, controlled, superiority trial
Across 17 German screening units, TOSYMA randomly assigned women to either digital breast tomosynthesis with synthesized two-dimensional mammography or standard digital mammography. Invasive cancers were detected at rates of 7.1 and 4.8 per 1,000 screened, respectively, in the evaluable groups.
SignificanceTOSYMA provided randomized evidence for higher invasive cancer detection with a configuration that obtains the two-dimensional image from the same acquisition. It therefore evaluated a screening protocol that can be implemented without a separate conventional two-dimensional exposure.
Key2026Nature Medicine16 citations
AI-based triage and decision support in mammography and digital tomosynthesis for breast cancer screening: a paired, noninferiority trial
The prospective paired trial enrolled 31,301 women and compared standard double reading with a workflow that marked low-risk exams normal and sent the rest to software-assisted double reading. Radiologist workload fell by 63.6%, cancer detection rose from 6.3 to 7.3 per 1,000, and recall increased, failing its noninferiority criterion.
SignificanceThe trial quantified how removing low-risk exams from the reading queue affected workload, detection, and recall within one program. Reading reductions were similar for digital mammography and digital breast tomosynthesis, while detection and recall increased in the digital mammography subgroup. This gives readers modality-specific evidence for workflow design.
DOI Full text Esperanza Elías-CabotMarina Álvarez-Benito mammographytomosynthesistriagereader studynoninferiority trialprospective validation
Key2026The Lancet Oncology1 citations
Invasive interval breast cancer rates after digital breast tomosynthesis plus synthesised mammography versus digital mammography screening (TOSYMA): 24-month follow-up of the multicentre, open-label, randomised, controlled, superiority trial
This TOSYMA follow-up linked women with a negative screening result to cancer registries. At 24 months, the tomosynthesis plus synthesized mammography arm had 14.2 invasive interval cancers per 10,000 women, compared with 18.6 with digital mammography. The adjusted odds ratio was 0.76, with a 95% confidence interval of 0.55 to 1.06.
SignificanceThe follow-up reports the second prespecified primary endpoint of the same randomized trial that found higher invasive cancer detection at screening in 2022. The confidence interval leaves uncertainty about whether higher detection at screening translates into fewer invasive interval cancers over the next two years.
DOI Stefanie WeigelWalter Heindel mammographytomosynthesisinterval cancer