Modalities

cone-beam CTCBCT

锥形束CT

CT with a cone-shaped X-ray beam and a flat-panel detector that captures a whole volume in a single rotation. Compact and low-dose, it is used in dentistry, radiotherapy positioning and on the interventional table, though its soft-tissue contrast falls short of conventional CT.

Conventional CT uses a thin fan of X-rays and gets a few slices per turn. Cone-beam CT opens the beam into a cone: each frame on the flat panel is a full 2D projection, and one rotation of roughly 200° to 360° yields hundreds of them. The volume is reconstructed with the approximate algorithm Lee Feldkamp, Lloyd Davis and James Kress published in 1984, still known as FDK.

In the late 1990s Piero Mozzo's group in Verona introduced the NewTom 9000, the first dental CBCT, while Yoshinori Arai developed a similar unit independently in Japan. A dental CBCT gives about 0.18 mSv, a fraction of a conventional CT. Mounted on a linear accelerator, CBCT checks the tumor's position before each radiotherapy session.

The wide cone admits a lot of scatter, panels have limited dynamic range, and a rotation takes seconds, so soft-tissue contrast and Hounsfield unit accuracy lag behind diagnostic CT, and motion artifacts are common.

Sources
  1. Feldkamp LA, Davis LC, Kress JW. Practical cone-beam algorithm. J Opt Soc Am A. 1984;1(6):612–619. doi.org/10.1364/JOSAA.1.000612
  2. Mozzo P, Procacci C, Tacconi A, Martini PT, Andreis IA. A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results. Eur Radiol. 1998;8(9):1558–1564. doi.org/10.1007/s003300050586
  3. RadiologyInfo.org (RSNA/ACR). Radiation dose in X-ray and CT exams. www.radiologyinfo.org/en/info/safety-xray

Key works · 2

Landmark1984Journal of the Optical Society of America A7,216 citations

Practical cone-beam algorithm

L. A. Feldkamp, L. C. Davis, J. W. Kress

Feldkamp, Davis, and Kress derived an approximate three-dimensional reconstruction from two-dimensional cone-beam projections along a circular trajectory. Their method uses convolution and back-projection, reduces to fan-beam reconstruction in the central plane, and was demonstrated using projections calculated from a mathematical phantom.

SignificanceThe FDK algorithm extended filtered back-projection to practical cone-beam volumes. Its computational form became a standard analytical baseline for circular cone-beam CT, including comparisons that test artifact suppression and image recovery with limited angular sampling.

Key2008Physics in Medicine & Biology2,229 citations

Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimization

Emil Y. Sidky, Xiaochuan Pan

The algorithm minimized image total variation while enforcing nonnegativity and requiring the calculated projections to agree with the data within a tolerance. It combined adaptive steepest descent with projection onto convex sets and was tested on simulated cone-beam data from disk and jaw phantoms.

SignificanceThe paper became a widely cited baseline for sparse-view CT, separating the allowed projection error from the total variation penalty. The same group subsequently extended the framework to measured low-intensity X-ray projections and tested it under reduced-exposure conditions.