Modalities · 1998

multidetector CTMDCT

多排CT

CT with many detector rows side by side along the table axis, so each rotation captures several slices at once. The jump from 4 slices in 1998 to 64 around 2004 made sub-millimeter whole-body scans and coronary CT angiography routine.

A single-row helical scanner gets one slice per rotation, so thin slices and long coverage are hard to achieve together. Multidetector CT divides the detector into rows along the z-axis and widens the beam to match, so each turn yields 4, 16, 64 or even 320 slices: thin and fast at once.

Elscint's CT-Twin of 1992 had two rows. In 1998 the major manufacturers launched four-slice scanners with 0.5 s rotation, and around 2004 64-slice systems arrived; with ECG gating they could freeze the beating coronary arteries within one breath-hold. A 320-row scanner covers 16 cm per turn, enough for the whole heart or brain in a single rotation.

The wider beam is a cone rather than a fan, so reconstruction has to account for cone-beam geometry or show cone-beam artifacts. Studies grew from dozens of images to thousands, pushing picture archiving and communication system storage, multiplanar reconstruction-based reading and, eventually, automated analysis.

Sources
  1. Goldman LW. Principles of CT: multislice CT. J Nucl Med Technol. 2008;36(2):57–68. tech.snmjournals.org/content/36/2/57

Key works · 4

Key2000Radiology561 citations

Four Multidetector-Row Helical CT: Image Quality and Volume Coverage Speed

Hui Hu, H. David He … Stanley H. Fox

This study compared four- and single-detector-row helical CT using phantom measurements, models, and ratings of clinical images. The investigators measured section profiles, noise, and artifacts. The four-row system achieved two to three times the volume coverage speed at comparable diagnostic image quality.

SignificanceThe comparison provided an early quantitative baseline for the advantages of acquiring several detector rows together. Subsequent multidetector systems extended the same approach to faster coverage of organs while preserving thin sections for multiplanar reformats.

Key2005Medical Physics300 citations

Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot

T. G. Flohr, K. Stierstorfer … C. H. McCollough

A CT scanner alternated its focal spot between two longitudinal positions, interleaving the samples from 32 detector rows to acquire 64 overlapping sections per rotation. The study measured longitudinal resolution, section profiles, and noise and evaluated helical image quality using phantom and patient scans.

SignificanceThe scanner demonstrated that geometry could increase longitudinal sampling density beyond the physical detector row count. The technique was implemented in a commercial 64-section system, supporting thin-section volume acquisitions and reformats through the resulting data in multiple planes.

Key2006RadioGraphics396 citations

Volume Rendering versus Maximum Intensity Projection in CT Angiography: What Works Best, When, and Why

Elliot K. Fishman, Derek R. Ney … Pamela T. Johnson

The authors compared maximum intensity projection and volume rendering for CT angiography. Clinical images illustrated what each display preserves and how superimposed structures or rendering choices can affect interpretation. They related the strengths and limitations of the displays to specific clinical tasks.

SignificanceThe review connected algorithmic differences to clinical decisions about displaying CT angiograms. It offered a reference for postprocessing multidetector CT volumes, distinguishing tasks that benefit from emphasizing high-attenuation vessels from those that require surrounding anatomy and depth cues.

Key2007Medical Physics1,028 citations

A three-dimensional statistical approach to improved image quality for multislice helical CT

Jean-Baptiste Thibault, Ken D. Sauer … Jiang Hsieh

The method used a three-dimensional forward model of multislice helical acquisition, statistical weighting of measurements, and a prior on the image. Phantom tests demonstrated improved resolution and lower noise alongside reduced helical cone-beam artifacts. Reconstructions of patient scans illustrated clinical image appearance.

SignificanceThe work demonstrated model-based iterative reconstruction on real multislice helical CT data. Later evaluations of commercial model-based reconstruction cite this study when examining how spatial resolution and image noise vary with exposure and reconstruction settings.