temporal resolution
时间分辨率
The time needed to capture one image, which decides whether motion can be frozen. The heart, blood and breathing never stop; if acquisition is too slow, they blur or leave motion artifacts.
Physics sets the limit. Sound travels through soft tissue at about 1540 m/s, so a single ultrasound imaging scan line to 15 cm depth must wait about 0.2 ms for its echoes to return. A 128-line image takes some 25 ms, about 40 frames per second, and imaging deeper or with more lines lowers the frame rate further.
CT needs at least half a rotation of data to reconstruct a slice. At 60 beats per minute a cardiac cycle lasts a second, and the relatively still diastolic phase is only part of it, so cardiac CT relies on fast gantries, dual X-ray sources and ECG gating. Functional MRI samples the whole brain every 1–2 seconds, but the haemodynamic response it tracks takes several seconds to peak anyway.
Temporal and spatial resolution usually trade against each other: more data gives a sharper image, more slowly. Echo-planar imaging, parallel imaging and compressed sensing are all attempts to break that trade.
Key works · 8
Landmark1986Journal of Magnetic Resonance1,492 citations
FLASH imaging. Rapid NMR imaging using low flip-angle pulses
FLASH used small flip-angle radio-frequency pulses to retain longitudinal magnetization for the next excitation, allowing successive measurements at short intervals. The sequence reversed the readout gradient to generate an echo. Demonstrated acquisition times were about one second for 64 by 128 pixels and six seconds for 256 by 256.
SignificanceCombining small excitations with gradient echoes established a way to speed MRI by shortening the repetition interval. Kose’s technical history traces the development of fast T1-weighted imaging from FLASH to later sequences that used radio-frequency spoiling to control residual transverse magnetization.
DOI Full text Axel HaaseJens FrahmKlaus-Dietmar Merboldt magnetic resonance imagingRF pulsetemporal resolution
Landmark2004IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control2,779 citations
Supersonic shear imaging: a new technique for soft tissue elasticity mapping
The researchers moved the location of focused acoustic excitation so that the resulting shear waves added coherently. An ultrafast scanner tracked propagation and the team reconstructed elasticity maps from the recorded motion, demonstrating the method in heterogeneous phantoms and healthy volunteers.
SignificanceCombining acoustic excitation with ultrafast imaging made it possible to form two-dimensional maps of mechanical properties from propagating shear waves. Later clinical work in the breast evaluated the additional information such maps could provide alongside conventional grayscale findings.
DOI Jérémy BercoffMathias Fink ultrasound imagingelastographytemporal resolution
Landmark2006European Radiology1,545 citations
First performance evaluation of a dual-source CT (DSCT) system
This evaluation measured the performance of a scanner with two X-ray tubes and two detector systems offset by approximately a right angle. Electrocardiogram-gated single-segment reconstruction achieved an 83 ms temporal resolution. Acquiring data at two tube voltages also let the authors separate bone from iodine-filled vessels.
SignificanceThe geometry reduced the time interval represented by a cardiac section, addressing motion in coronary imaging. Subsequent commercial systems retained dual sources for fast cardiac acquisitions and also used the two tubes to collect data at different spectra for dual-energy CT.
DOI Thomas G. FlohrBernd M. Ohnesorge computed tomographytemporal resolutiondual-energy CTmaterial decomposition
Key1968Ultrasonics196 citations
Electronic sector scanning for ultrasonic diagnosis
Somer developed an electronically controlled ultrasound sector scanner to overcome the slow refresh of mechanical B-scanning. An array of 21 elements steered pulses across a 90-degree sector and could repeat the scan across that sector about 30 times per second.
SignificanceThe work showed that an array could sweep a beam across a sector repeatedly without moving the probe. A follow-up report applied this electronic scanner to neurological cases, and phased arrays became a means of acquiring real-time cardiac sections.
DOI Jan C. Somer ultrasound imagingtransducerbeamformingB-modetemporal resolution
Key1990Journal of Digital Imaging271 citations
Real time volumetric ultrasound imaging system
The authors used a 289-element two-dimensional phased array to scan a pyramidal volume with pulse echoes. Parallel processing on reception produced approximately eight volume acquisitions per second and also let the system display sections selected from the acquired three-dimensional data.
SignificanceThe prototype combined a two-dimensional array, parallel reception, and section display from acquired volumes to show how rapidly changing anatomy could be scanned in three dimensions. Later real-time three-dimensional echocardiography developed this approach using matrix-array transducers.
DOI Full text Olaf T. von RammStephen W. Smith ultrasound imagingbeamformingtransducertemporal resolutionvoxel
Key2009IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control2,057 citations
Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography
The authors improved single plane-wave images by coherently summing echoes from transmissions at different angles. Their model predicted image quality comparable to conventional B-mode imaging with roughly one-tenth as many transmissions. Tissue-mimicking phantom experiments supported the prediction, and the paper also demonstrated its application to transient elastography.
SignificanceCoherent compounding provided a way to improve image quality with a limited number of additional transmissions while preserving high frame rates for tasks such as shear wave tracking. Urs and colleagues explicitly credited the method and adapted it to image anatomy and blood flow in the human eye.
DOI Gabriel MontaldoMathias Fink ultrasound imagingB-modebeamformingtemporal resolutionelastography
Key2014IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control1,356 citations
Ultrafast imaging in biomedical ultrasound
This review explained how plane or diverging wave transmission illuminates a broad region and allows many image locations to be reconstructed from the received array signals. It discussed frame rates above a thousand per second and applications to elasticity imaging and blood flow measurements.
SignificanceThe review linked plane- and diverging-wave acquisition to shear wave tracking and blood flow measurement through a shared beamforming framework. Urs and colleagues cited it when introducing the acquisition principles in their subsequent ocular study, which used compounded plane waves to depict anatomy and flow.
DOI Mickaël TanterMathias Fink beamformingultrasound imagingelastographytemporal resolution
Notable1991Science746 citations
Echo-Planar Imaging: Magnetic Resonance Imaging in a Fraction of a Second
This review described the implementation of echo-planar imaging, which uses rapid gradient changes to acquire an image after a single spin excitation, in a fraction of a second. It discussed applications of that speed to diffusion, perfusion, and dynamic studies of the heart and other organs.
SignificanceThe review documents the transition from the echo-planar imaging proposal to an acquisition that could be implemented on imaging hardware. Subsequent human functional experiments used related rapid readouts to follow task-evoked signals, demonstrating an application discussed in the review.
DOI Michael K. StehlingPeter Mansfield echo-planar imagingk-spacetemporal resolution