Modalities · 1977

echo-planar imagingEPI

平面回波成像

A way to read an entire MR image after a single excitation, switching the gradients back and forth so the echo train sweeps all of k-space in tens of milliseconds. It is the standard acquisition method for functional MRI and diffusion MRI, at the cost of geometric distortion.

A conventional sequence collects one line of k-space per excitation, so a 256-line image needs 256 repetitions. EPI excites once, then swings the readout gradient rapidly between positive and negative. Each swing forms an echo and reads one line, while small phase-encoding blips step on to the next row, so the trajectory sweeps back and forth through k-space. A whole image takes roughly 50–100 ms.

Peter Mansfield proposed the method in Nottingham in 1977. It demanded gradients that switched faster and harder than the hardware of the day allowed, and only became routine on clinical scanners in the 1990s.

This speed has drawbacks. Because time accumulates from line to line, EPI is acutely sensitive to field inhomogeneity: near the sinuses and ear canals images stretch and lose signal. Small mismatches between odd and even echoes add a faint N/2 ghost, and the rapidly switching gradients are a large part of the scanner's noise.

Sources
  1. Mansfield P. Multi-planar image formation using NMR spin echoes. J Phys C: Solid State Phys. 1977;10(3):L55–L58. doi.org/10.1088/0022-3719/10/3/004
  2. Stehling MK, Turner R, Mansfield P. Echo-planar imaging: magnetic resonance imaging in a fraction of a second. Science. 1991;254(5028):43–50. doi.org/10.1126/science.1925560

Key works · 5

Landmark1977Journal of Physics C: Solid State Physics2,341 citations

Multi-planar image formation using NMR spin echoes

P Mansfield

Mansfield proposed a method for rapidly acquiring two- or three-dimensional spin-density data using nuclear magnetic resonance echoes in time-dependent field gradients. The analysis showed how signals arising from spins across a plane or a set of planes could be distinguished within the readout.

SignificanceThis paper set out the acquisition principle of echo-planar imaging. Once gradient and readout hardware could implement the approach, a single excitation could supply the data for an image, providing the temporal resolution needed to follow functional brain signals on the scale of seconds.

Landmark1992Proceedings of the National Academy of Sciences4,763 citations

Dynamic magnetic resonance imaging of human brain activity during primary sensory stimulation

K. K. Kwong, J. W. Belliveau … B. P. Poncelet

The authors acquired fast echo-planar images through visual and motor tasks, using a gradient-echo sequence sensitive to oxygenation and an inversion-recovery sequence sensitive to flow. During 8 Hz visual stimulation in seven volunteers, both sequences showed a mean signal increase of about 1.8% in primary visual cortex.

SignificanceThis study demonstrated that endogenous hemodynamic contrast could identify task responses in human cortex on the scale of seconds. Together with independent reports in the same year, it established a way to map human task responses with MRI without injecting a contrast agent.

Landmark1992Magnetic Resonance in Medicine2,200 citations

Time course EPI of human brain function during task activation

Peter A. Bandettini, Eric C. Wong … James S. Hyde

Using gradient-echo echo-planar imaging, the authors repeatedly scanned the human brain during a finger movement task. They observed a local signal increase and compared the time course at each location with the task schedule to map regions where the two were correlated.

SignificanceThree independent groups established human task functional MRI with endogenous contrast in the same period. This study showed how a rapid image series and the temporal correspondence between signal and task could be used to localize motor responses.

Landmark1995Magnetic Resonance in Medicine10,980 citations

Functional connectivity in the motor cortex of resting human brain using echo-planar MRI

Bharat Biswal, F. Zerrin Yetkin … James S. Hyde

The investigators first localized sensorimotor regions using hand movement and then recorded a resting echo-planar image series. Low-frequency fluctuations in the resting signal showed strong temporal correlations across the localized regions and with other brain areas related to motor function.

SignificanceThe study introduced the use of correlations in spontaneous resting signals to estimate functional relationships between brain regions. Later whole-brain resting-state imaging studies extended this type of measurement to identify groups of regions forming distributed functional networks.

Notable1991Science746 citations

Echo-Planar Imaging: Magnetic Resonance Imaging in a Fraction of a Second

Michael K. Stehling, Robert Turner, Peter Mansfield

This review described the implementation of echo-planar imaging, which uses rapid gradient changes to acquire an image after a single spin excitation, in a fraction of a second. It discussed applications of that speed to diffusion, perfusion, and dynamic studies of the heart and other organs.

SignificanceThe review documents the transition from the echo-planar imaging proposal to an acquisition that could be implemented on imaging hardware. Subsequent human functional experiments used related rapid readouts to follow task-evoked signals, demonstrating an application discussed in the review.