k-space

k空间

The spatial-frequency domain where MRI raw data live. The scanner measures the Fourier transform of the image, and an inverse transform turns it into a picture. The center carries overall contrast, the edges carry fine detail, and how much of it is sampled sets the scan time.

Gradients make each proton's phase vary linearly with position, so the received signal at any instant is one spatial frequency component of the image, at . Changing the gradients steers a path through k-space. A conventional sequence collects one line per repetition time, so scan time is roughly the number of phase-encoding lines times TR: 256 lines at a TR of 500 ms take over two minutes.

In 1983 Ljunggren and Donald Twieg independently proposed describing every MRI acquisition as a trajectory through k-space, the view still used to design sequences today.

Keep only the center and the image keeps its contrast but turns blurry; keep only the edges and only outlines remain. Skipping lines below the Nyquist rate folds the image over on itself through aliasing, the problem parallel imaging, compressed sensing and deep-learning reconstruction attack by recovering full images from less data. Patient motion during acquisition leaves ghosts along the phase-encoding direction.

Sources
  1. Twieg DB. The k-trajectory formulation of the NMR imaging process with applications in analysis and synthesis of imaging methods. Med Phys. 1983;10(5):610–621. doi.org/10.1118/1.595331
  2. Ljunggren S. A simple graphical representation of Fourier-based imaging methods. J Magn Reson. 1983;54(2):338–343. doi.org/10.1016/0022-2364(83)90060-4

Key works · 19

Landmark1975Journal of Magnetic Resonance881 citations

NMR Fourier zeugmatography

Anil Kumar, Dieter Welti, Richard R. Ernst

The authors applied sequences of pulsed magnetic field gradients during a series of free induction decays to encode the spatial frequency components of spin density. A multidimensional Fourier transform reconstructed the image. The paper set out how the gradient pulses and measurement timing determined which components were acquired.

SignificanceDirect Fourier sampling offered a framework for combining frequency encoding with repeated phase encoding. The subsequent spin-warp method adapted Fourier encoding for routine image acquisition, establishing an approach that remains central to conventional clinical magnetic resonance imaging.

Landmark1977Journal of Physics C: Solid State Physics2,341 citations

Multi-planar image formation using NMR spin echoes

P Mansfield

Mansfield proposed a method for rapidly acquiring two- or three-dimensional spin-density data using nuclear magnetic resonance echoes in time-dependent field gradients. The analysis showed how signals arising from spins across a plane or a set of planes could be distinguished within the readout.

SignificanceThis paper set out the acquisition principle of echo-planar imaging. Once gradient and readout hardware could implement the approach, a single excitation could supply the data for an image, providing the temporal resolution needed to follow functional brain signals on the scale of seconds.

Landmark1980Physics in Medicine and Biology952 citations

Spin warp NMR imaging and applications to human whole-body imaging

W A Edelstein, J M S Hutchison … T Redpath

The Aberdeen group described spin-warp imaging, in which a gradient encodes phase with a different amplitude on successive excitations and a readout gradient encodes frequency. They presented examples of human imaging with the sequence using a 0.04 tesla magnet large enough to accommodate the body.

SignificanceSpin-warp imaging made slice selection and Fourier encoding practical in a sequence for human scanning. Aberdeen subsequently used its system to scan a patient, and the method's line-by-line acquisition of spatial frequency data became a standard approach in clinical MRI.

Landmark1986Magnetic Resonance in Medicine2,345 citations

RARE imaging: A fast imaging method for clinical MR

Jürgen Hennig, Arno Nauerth, Hartmut Friedburg

The team assigned a different phase-encoding step to each of several spin echoes following excitation, collecting enough data for a 256 by 256 image in 2 to 40 seconds. They examined artifacts caused by changing T2 weighting across the echo train and how phase-encoding order could reduce them.

SignificanceAcquiring multiple k-space lines from a train of refocused echoes became the basis of clinical fast or turbo spin-echo imaging. Later implementations retained this approach to accelerating T2-weighted acquisitions while addressing the blurring and radio-frequency demands of long echo trains.

Landmark1999Magnetic Resonance in Medicine7,212 citations

SENSE: Sensitivity encoding for fast MRI

Klaas P. Pruessmann, Markus Weiger … Peter Boesiger

The authors used spatial differences in receiver coil sensitivity as additional encoding and formulated a reconstruction from reduced Fourier sampling. Two coils halved acquisition time for the demonstrated brain images, and five coils gave double-oblique cardiac images in one-third of the conventional time.

SignificanceSENSE established image-domain parallel imaging by showing how receiver sensitivity maps can separate pixels folded together by undersampling. Its formulation also described noise amplification associated with coil geometry, an important consideration when choosing an acceleration factor for routine clinical imaging.

Landmark2002Magnetic Resonance in Medicine5,951 citations

Generalized autocalibrating partially parallel acquisitions (GRAPPA)

Mark A. Griswold, Peter M. Jakob … Axel Haase

The authors used extra calibration lines to estimate local relationships among the receiver coils in k-space. They used those relationships to recover missing data for each coil before reconstructing and combining coil images. The paper described implementation and signal-to-noise behavior and demonstrated in vivo imaging.

SignificanceGRAPPA incorporated calibration into parallel acquisition and reconstructed missing samples directly in k-space. It became a common clinical parallel imaging method and supplied a reference for later approaches that generalized how relationships among coil measurements could be calibrated and enforced.

Landmark2007Magnetic Resonance in Medicine7,800 citations

Sparse MRI: The application of compressed sensing for rapid MR imaging

Michael Lustig, David Donoho, John M. Pauly

The authors undersampled phase-encoding lines with a variable-density pseudorandom pattern and reconstructed images by penalizing transform coefficients subject to consistency with the measured data. Examples in multislice fast spin-echo brain images and three-dimensional contrast-enhanced angiography demonstrated how the method could accelerate acquisition.

SignificanceThe paper converted compressed sensing into an acquisition and reconstruction procedure that could be implemented on an MR scanner. Later studies combined sparse recovery and parallel imaging and evaluated the resulting shorter acquisitions in settings such as contrast-enhanced pediatric abdominal MRI.

Landmark2018IEEE Transactions on Medical Imaging1,499 citations

A Deep Cascade of Convolutional Neural Networks for Dynamic MR Image Reconstruction

Jo Schlemper, Jose Caballero … Daniel Rueckert

A cascade alternated convolutional corrections with enforcement of consistency with acquired k-space samples. For dynamic cardiac MRI, neighboring frames also provided information for joint reconstruction. In the authors' undersampling experiments the method better preserved anatomy than the reported comparators and reconstructed complete sequences in under ten seconds.

SignificanceThe cascade was an influential example of pairing learned corrections with repeated data consistency steps. It also demonstrated how temporal information could supplement undersampled measurements. Later experimental reviews of MRI reconstruction retained cascaded networks as a distinct comparison class.

Landmark2018Nature1,965 citations

Image reconstruction by domain-transform manifold learning

Bo Zhu, Jeremiah Z. Liu … Matthew S. Rosen

AUTOMAP used dense and convolutional layers to learn a mapping from sensor measurements to images. The authors evaluated the design with several MRI sampling trajectories and additional imaging simulations. In the demonstrated settings the learned mapping reduced noise or artifacts, and the paper examined the learned intermediate representations.

SignificanceAUTOMAP became a widely cited instance of learning the sensor-to-image transformation directly. The parallel MRI survey places such mappings alongside architectures that retain the acquisition operator, helping readers identify where different reconstruction designs incorporate physics, use training data, and enforce consistency.

Landmark2018Magnetic Resonance in Medicine2,036 citations

Learning a variational network for reconstruction of accelerated MRI data

Kerstin Hammernik, Teresa Klatzer … Florian Knoll

The network unrolled gradient updates of a variational model for multicoil MRI, learning its filters, activation functions, and weights for the measured-data terms. It was tested on retrospectively and prospectively undersampled knees. Quantitative comparisons and a reader study supported the results for fourfold regular undersampling.

SignificanceThe variational network made the parallel MRI acquisition model and the learned image prior explicit within a single trainable system. The later end-to-end variational network developed this design further by incorporating coil sensitivity estimation in the same reconstruction model.

Landmark2020IEEE Transactions on Pattern Analysis and Machine Intelligence1,028 citations

ADMM-CSNet: A Deep Learning Approach for Image Compressive Sensing

Yan Yang, Jian Sun … Zongben Xu

The authors unrolled alternating-direction optimization into two trainable architectures for recovering images from sparse measurements. They learned the transforms and model and solver parameters together. On complex MRI and real-valued natural images, the trained networks delivered competitive reconstruction accuracy and speed in the reported comparisons.

SignificanceThe paper extended the 2016 Deep ADMM-Net proposal into a broader family of trainable solvers. It is a reference for unfolding the steps of an explicit inverse problem solver into layers, a design also seen in parallel MRI variational networks and later model-based architectures.

Landmark2020Radiology: Artificial Intelligence606 citations

fastMRI: A Publicly Available Raw k-Space and DICOM Dataset of Knee Images for Accelerated MR Image Reconstruction Using Machine Learning

Florian Knoll, Jure Zbontar … Aaron Defazio

The fastMRI knee release provided multicoil raw k-space measurements from 1,594 examinations and a separate collection of clinical DICOM image series. Its vendor-neutral measurement format made retrospective undersampling and comparison against fully sampled reconstructions possible using common data instead of locally acquired raw scans.

SignificanceFastMRI supported reproducible comparisons by making measurements available alongside reference images. The 2019 knee challenge used this release, and the later brain challenge broadened the benchmark to include pathology-centered reader evaluations and a track for scanners outside the training data.

Key1983Medical Physics497 citations

The k-trajectory formulation of the NMR imaging process with applications in analysis and synthesis of imaging methods

Donald B. Twieg

Twieg formulated several nuclear magnetic resonance imaging methods as sampling in the spatial-frequency domain, with the gradient history determining the trajectory of sampled points. He used this common description to compare how projection, spin-warp, Fourier, and echo-planar methods measure different spatial frequencies.

SignificanceA common spatial-frequency description made it possible to compare how apparently different sequences cover the data needed for an image and to design new acquisitions. Gradient-defined trajectories in k-space remain standard in describing MRI acquisition and reconstruction.

Key1997Magnetic Resonance in Medicine2,458 citations

Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with radiofrequency coil arrays

Daniel K. Sodickson, Warren J. Manning

SMASH formed spatial harmonics by linearly combining the sensitivities of array elements. This enabled the reconstruction of skipped phase-encoding lines in k-space before Fourier transformation. Using commercial phased arrays on two MRI systems, the authors demonstrated a twofold reduction in acquisition time.

SignificanceSMASH established a way to recover phase encoding in k-space from the spatial responses of a coil array. The later VD-AUTO-SMASH method used extra acquired data to calibrate the synthesis. The GRAPPA paper identifies that self-calibrating variant as one of its direct predecessors.

Key2018arXiv1,345 citations

fastMRI: An Open Dataset and Benchmarks for Accelerated MRI

Jure Zbontar, Florian Knoll … Aaron Defazio

The initial fastMRI report specified single-coil and multicoil knee reconstruction tasks, with fourfold and eightfold retrospective undersampling and common scoring metrics. It compared total variation reconstruction and U-Net baselines on the same splits. The U-Net baselines obtained higher test-set structural similarity in both tasks.

SignificanceReference implementations and results made the data release usable as a shared reconstruction benchmark. The 2019 fastMRI competition used the resource to organize single-coil and multicoil tracks and added radiologist review to numerical screening, extending the initial experimental protocol.

Key2020IEEE Signal Processing Magazine514 citations

Deep-Learning Methods for Parallel Magnetic Resonance Imaging Reconstruction: A Survey of the Current Approaches, Trends, and Issues

Florian Knoll, Kerstin Hammernik … Mehmet Akçakaya

The review explains linear and nonlinear parallel MRI reconstruction from image-domain and k-space perspectives. It describes ways to use neural networks as learned image regularizers or as interpolators for missing k-space measurements. It also discusses measurement consistency, public benchmarks, and performance across acquisition conditions.

SignificanceThis widely cited technical reference ties learning methods to the physics of multicoil MRI. It supplies a common vocabulary for comparing image enhancement, direct reconstruction and methods that explicitly enforce data consistency, and documents the early availability of benchmarks for testing those differences.

Key2020Magnetic Resonance in Medicine436 citations

Self‐supervised learning of physics‐guided reconstruction neural networks without fully sampled reference data

Burhaneddin Yaman, Seyed Amir Hossein Hosseini … Mehmet Akçakaya

SSDU split the acquired k-space samples into two disjoint sets. One supplied data consistency within a physics-guided network and the other defined the training loss. On fourfold accelerated knee MRI, results were close to fully supervised reconstruction. The authors also evaluated prospectively undersampled brain data.

SignificanceThe paper established a way to train physics-guided reconstruction from the kind of incomplete measurements used at deployment. The later multi-mask SSDU study directly extended its measurement partitioning scheme to make fuller use of samples available within each scan.

Key2022Medical Image Analysis671 citations

Score-based diffusion models for accelerated MRI

Hyungjin Chung, Jong Chul Ye

The method learned a score model from fully sampled MR images and combined diffusion sampling with k-space data consistency during reconstruction. The authors evaluated the same learned prior with single-coil and multicoil acquisitions and different undersampling schemes, reporting improvements over the reconstruction methods they compared.

SignificanceThe paper showed that a single independently learned image prior could be applied to several acquisition settings. Subsequent work on adaptive diffusion priors continued this line of investigation by examining how the learned prior could be adjusted at test time for individual accelerated MRI acquisitions.

Notable1991Science746 citations

Echo-Planar Imaging: Magnetic Resonance Imaging in a Fraction of a Second

Michael K. Stehling, Robert Turner, Peter Mansfield

This review described the implementation of echo-planar imaging, which uses rapid gradient changes to acquire an image after a single spin excitation, in a fraction of a second. It discussed applications of that speed to diffusion, perfusion, and dynamic studies of the heart and other organs.

SignificanceThe review documents the transition from the echo-planar imaging proposal to an acquisition that could be implemented on imaging hardware. Subsequent human functional experiments used related rapid readouts to follow task-evoked signals, demonstrating an application discussed in the review.