parallel imaging
并行成像
An MRI speed-up that skips lines of k-space and uses an array of receive coils with different spatial sensitivities to unfold the resulting overlap. Accelerations of two to three are routine.
Sample every other line of k-space and the scan time halves, but so does the field of view, and the image aliases into two overlapping copies. With an array of eight, 32 or more small receive coils, each coil is highly sensitive to nearby tissue and weakly sensitive to distant tissue. Knowing those sensitivity maps, one can solve a small system of equations at every pixel and pull the copies apart. That is SENSE, working in image space; GRAPPA instead fits the missing k-space lines from their neighbors.
Daniel Sodickson and Warren Manning demonstrated the first method, SMASH, in 1997. SENSE (Klaas Pruessmann and colleagues, 1999) and GRAPPA (Mark Griswold and colleagues, 2002) made it clinical routine.
The time saved is paid for in signal-to-noise ratio: , where is the acceleration and the geometry factor depends on the coil layout. Push too high and noise increases markedly in the center of the image, with residual fold-over. Compressed sensing and deep learning are often stacked on top to go further.
- Sodickson DK, Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn Reson Med. 1997;38(4):591–603. doi.org/10.1002/mrm.1910380414
- Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42(5):952–962. mriquestions.com/uploads/3/4/5/7/34572113/pruessmann-sense.pdf
- Griswold MA, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47(6):1202–1210. doi.org/10.1002/mrm.10171
Key works · 7
Landmark1990Magnetic Resonance in Medicine2,653 citations
The NMR phased array
The authors reduced coupling among receiving coils by overlapping neighboring coils and using low-input-impedance preamplifiers. They derived a rule for combining coil data to maximize signal-to-noise ratio. Two four-element arrays yielded twofold and threefold gains at the depth of the spine compared with one large coil under matched imaging conditions.
SignificanceThe receive array provided broad coverage while preserving separate measurements from local coils. Both SMASH and SENSE cited the array work and used differences among coil sensitivities as spatial encoding, turning those separate measurements into a means of accelerating acquisition.
DOI Peter B. RoemerWilliam A. EdelsteinO. M. Mueller magnetic resonance imagingsignal-to-noise ratioparallel imaging
Landmark1999Magnetic Resonance in Medicine7,212 citations
SENSE: Sensitivity encoding for fast MRI
The authors used spatial differences in receiver coil sensitivity as additional encoding and formulated a reconstruction from reduced Fourier sampling. Two coils halved acquisition time for the demonstrated brain images, and five coils gave double-oblique cardiac images in one-third of the conventional time.
SignificanceSENSE established image-domain parallel imaging by showing how receiver sensitivity maps can separate pixels folded together by undersampling. Its formulation also described noise amplification associated with coil geometry, an important consideration when choosing an acceleration factor for routine clinical imaging.
DOI Klaas P. PruessmannPeter Boesiger parallel imagingaliasingk-space
Landmark2002Magnetic Resonance in Medicine5,951 citations
Generalized autocalibrating partially parallel acquisitions (GRAPPA)
The authors used extra calibration lines to estimate local relationships among the receiver coils in k-space. They used those relationships to recover missing data for each coil before reconstructing and combining coil images. The paper described implementation and signal-to-noise behavior and demonstrated in vivo imaging.
SignificanceGRAPPA incorporated calibration into parallel acquisition and reconstructed missing samples directly in k-space. It became a common clinical parallel imaging method and supplied a reference for later approaches that generalized how relationships among coil measurements could be calibrated and enforced.
DOI Mark A. GriswoldAxel Haase parallel imagingk-spacealiasing
Landmark2018Magnetic Resonance in Medicine2,036 citations
Learning a variational network for reconstruction of accelerated MRI data
The network unrolled gradient updates of a variational model for multicoil MRI, learning its filters, activation functions, and weights for the measured-data terms. It was tested on retrospectively and prospectively undersampled knees. Quantitative comparisons and a reader study supported the results for fourfold regular undersampling.
SignificanceThe variational network made the parallel MRI acquisition model and the learned image prior explicit within a single trainable system. The later end-to-end variational network developed this design further by incorporating coil sensitivity estimation in the same reconstruction model.
DOI arXiv Full text Kerstin HammernikFlorian Knoll parallel imagingk-spacecompressed sensingdeep-learning reconstructiondeep unrollingdata consistency
Key1997Magnetic Resonance in Medicine2,458 citations
Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with radiofrequency coil arrays
SMASH formed spatial harmonics by linearly combining the sensitivities of array elements. This enabled the reconstruction of skipped phase-encoding lines in k-space before Fourier transformation. Using commercial phased arrays on two MRI systems, the authors demonstrated a twofold reduction in acquisition time.
SignificanceSMASH established a way to recover phase encoding in k-space from the spatial responses of a coil array. The later VD-AUTO-SMASH method used extra acquired data to calibrate the synthesis. The GRAPPA paper identifies that self-calibrating variant as one of its direct predecessors.
DOI Daniel K. SodicksonWarren J. Manning parallel imagingk-spacesignal-to-noise ratiophase encoding
Key2020IEEE Signal Processing Magazine514 citations
Deep-Learning Methods for Parallel Magnetic Resonance Imaging Reconstruction: A Survey of the Current Approaches, Trends, and Issues
The review explains linear and nonlinear parallel MRI reconstruction from image-domain and k-space perspectives. It describes ways to use neural networks as learned image regularizers or as interpolators for missing k-space measurements. It also discusses measurement consistency, public benchmarks, and performance across acquisition conditions.
SignificanceThis widely cited technical reference ties learning methods to the physics of multicoil MRI. It supplies a common vocabulary for comparing image enhancement, direct reconstruction and methods that explicitly enforce data consistency, and documents the early availability of benchmarks for testing those differences.
DOI arXiv Florian KnollMehmet Akçakaya parallel imagingk-spacedeep-learning reconstructiondata consistency
Key2020MICCAI 2020 (LNCS 12262)539 citations
End-to-End Variational Networks for Accelerated MRI Reconstruction
The end-to-end variational network estimated coil sensitivity maps and reconstructed undersampled multicoil MRI in a single trainable architecture. Cascaded refinement used learned image regularization and consistency with measured k-space. The paper reported better reconstruction scores than its comparators on fastMRI knee and brain data.
SignificanceThe architecture brought coil sensitivity estimation into the optimization of the reconstruction system. The fastMRI project distributed a Variational Network implementation, giving researchers an inspectable baseline for experiments on multicoil data and a reproducible reference when comparing alternative reconstruction stages.
DOI arXiv Anuroop SriramPatricia Johnson fastMRIparallel imagingdeep-learning reconstructiondeep unrollingdata consistency