aliasing

混叠

When sampling is too sparse, a fine pattern appears as a coarse one. In MRI, anatomy outside the field of view wraps onto the opposite side; in Doppler ultrasound, fast flow appears to run backwards. Both are aliasing.

Consider fine stripes repeating every 0.2 mm, sampled only every 0.3 mm. Connecting the samples gives a false stripe pattern with a 0.6 mm period. Wagon wheels turning backwards on film and moiré in photos of screens are the same effect. As the sampling theorem predicts, anything above half the sampling frequency folds down into lower frequencies.

MRI wrap-around is the textbook case: if the field of view in the phase-encoding direction is smaller than the head, the tip of the nose appears at the back of the skull. The cure is a larger field of view or oversampling. In color Doppler ultrasound, flow faster than the velocity matching half the pulse repetition frequency flips from red to blue, as if it had reversed.

Too few projection angles in CT produce fine streaks towards the edge of the image. There are only two real defences: sample more densely, or filter out the high frequencies before sampling.

Key works · 5

Landmark1949Proceedings of the IRE9,337 citations

Communication in the Presence of Noise

Claude E. Shannon

Shannon represented messages and transmitted signals as points in function spaces and expressed a band-limited waveform in terms of uniformly spaced samples. The paper gave a construction for recovering the waveform from sufficiently close samples and analyzed achievable transmission rates in the presence of noise.

SignificanceThe sampling relationship provides a theoretical basis for the spatial spacing of digital image samples. Applying the same reasoning to spatial frequencies explains why detail beyond the range represented by the sampling grid can appear as spurious lower-frequency structure.

Landmark1988ACM SIGGRAPH Computer Graphics1,806 citations

Volume rendering

Robert A. Drebin, Loren Carpenter, Pat Hanrahan

The authors assigned different optical and color properties to materials and their boundaries and computed images by simulating absorption along view rays. The procedures addressed aliasing and quantization artifacts, and the paper illustrated the rendering method on several kinds of volumetric data, including medical images.

SignificanceThis approach made it possible to show several structures and their occlusion relationships from one volume, with optical properties controlling which materials were emphasized. Clinical CT angiography later used related volume-rendering methods to depict vessels alongside relevant surrounding anatomy.

Landmark1999Magnetic Resonance in Medicine7,212 citations

SENSE: Sensitivity encoding for fast MRI

Klaas P. Pruessmann, Markus Weiger … Peter Boesiger

The authors used spatial differences in receiver coil sensitivity as additional encoding and formulated a reconstruction from reduced Fourier sampling. Two coils halved acquisition time for the demonstrated brain images, and five coils gave double-oblique cardiac images in one-third of the conventional time.

SignificanceSENSE established image-domain parallel imaging by showing how receiver sensitivity maps can separate pixels folded together by undersampling. Its formulation also described noise amplification associated with coil geometry, an important consideration when choosing an acceleration factor for routine clinical imaging.

Landmark2002Magnetic Resonance in Medicine5,951 citations

Generalized autocalibrating partially parallel acquisitions (GRAPPA)

Mark A. Griswold, Peter M. Jakob … Axel Haase

The authors used extra calibration lines to estimate local relationships among the receiver coils in k-space. They used those relationships to recover missing data for each coil before reconstructing and combining coil images. The paper described implementation and signal-to-noise behavior and demonstrated in vivo imaging.

SignificanceGRAPPA incorporated calibration into parallel acquisition and reconstructed missing samples directly in k-space. It became a common clinical parallel imaging method and supplied a reference for later approaches that generalized how relationships among coil measurements could be calibrated and enforced.

Landmark2007Magnetic Resonance in Medicine7,800 citations

Sparse MRI: The application of compressed sensing for rapid MR imaging

Michael Lustig, David Donoho, John M. Pauly

The authors undersampled phase-encoding lines with a variable-density pseudorandom pattern and reconstructed images by penalizing transform coefficients subject to consistency with the measured data. Examples in multislice fast spin-echo brain images and three-dimensional contrast-enhanced angiography demonstrated how the method could accelerate acquisition.

SignificanceThe paper converted compressed sensing into an acquisition and reconstruction procedure that could be implemented on an MR scanner. Later studies combined sparse recovery and parallel imaging and evaluated the resulting shorter acquisitions in settings such as contrast-enhanced pediatric abdominal MRI.