Doppler ultrasound

多普勒超声

Ultrasound that measures blood flow through the Doppler effect: echoes from blood cells moving towards the probe return at a higher frequency, those moving away at a lower one. Colour Doppler paints speed and direction in red and blue over the B-mode image.

The Doppler shift is . With a 5 MHz probe and blood moving at 0.5 m/s straight towards it, the shift is about 3.2 kHz, comfortably within human hearing, which is why Doppler machines can play blood flow as sound. As the angle between beam and vessel approaches 90° the shift vanishes, so velocity readings need angle correction.

In 1957 Shigeo Satomura at Osaka University used continuous-wave Doppler to record the motion of heart walls and valves, then turned to blood flow. In 1985 Chihiro Kasai, Koroku Namekawa and colleagues at the Japanese company Aloka published the autocorrelation method that made real-time two-dimensional color flow imaging practical.

Pulsed Doppler can choose the depth it samples, but fires only so many pulses per second; above a velocity limit it suffers aliasing, showing fast flow as if it ran backwards. Carotid stenosis, deep-vein thrombosis and blood flow in the fetal umbilical cord are all assessed this way.

Sources
  1. Satomura S. Ultrasonic Doppler method for the inspection of cardiac functions. J Acoust Soc Am. 1957;29(11):1181–1185. doi.org/10.1121/1.1908737
  2. Kasai C, Namekawa K, Koyano A, Omoto R. Real-time two-dimensional blood flow imaging using an autocorrelation technique. IEEE Trans Sonics Ultrason. 1985;SU-32(3):458–464. doi.org/10.1109/T-SU.1985.31615

Key works · 4

Landmark1957The Journal of the Acoustical Society of America290 citations

Ultrasonic Doppler Method for the Inspection of Cardiac Functions

Shigeo Satomura

Satomura directed continuous ultrasound through the chest and recorded Doppler shifts in waves reflected by moving cardiac structures. Simultaneous electrocardiographic and phonocardiographic recordings let him compare the timing of the signals and use them to characterize motions such as those of heart valves.

SignificanceThe study showed that ultrasound frequency shifts could provide information on cardiac motion in humans. Satomura subsequently applied Doppler to blood flow, and later pulsed and imaging systems developed the ability to localize these measurements within vessels or the heart.

Landmark1985IEEE Transactions on Sonics and Ultrasonics1,395 citations

Real-Time Two-Dimensional Blood Flow Imaging Using an Autocorrelation Technique

Chihiro Kasai, Kouroku Namekawa … Ryozo Omoto

The authors combined pulsed Doppler with autocorrelation estimation to display blood flow in a cross section in real time. Color encoded direction and variations in flow. They tested mechanical and electronic scanners in phantoms and also examined normal and diseased hearts.

SignificanceAutocorrelation estimation made it practical to refresh a two-dimensional display while adding Doppler flow information. Color flow imaging subsequently became a means of locating abnormal flow in relation to anatomical structures during echocardiography and vascular ultrasound examinations.

Key1970IEEE Transactions on Sonics and Ultrasonics440 citations

Pulsed Ultrasonic Doppler Blood-Flow Sensing

Donald W. Baker

Baker described an instrument for measuring blood flow with repeated ultrasound pulses. The receiver selected returning signals in a chosen time window to define a depth range and used Doppler information across the pulses to characterize flow in that region.

SignificanceCombining time gating with Doppler analysis made it possible to specify where along the beam flow was measured. Later color flow systems retained this localization and used pulsed Doppler to display motion within successive regions of a two-dimensional section.

Key1974IEEE Transactions on Biomedical Engineering244 citations

Ultrasonic Duplex Echo-Doppler Scanner

Frank E. Barber, Donald W. Baker … John M. Reid

The authors combined multigate pulsed Doppler with a rotating B-mode scanner, using the same transducer to align flow and tissue images. They demonstrated the system on normal arteries in vivo. Trials in patients with verified occlusive disease led them to propose a two-transducer design.

SignificanceDuplex scanning let clinicians locate Doppler flow signals on an anatomical image with one instrument. The European Association of Cardiovascular Imaging identifies the scanner as a milestone after range-gated Doppler. The group later introduced a digital display superimposing flow information on B-mode images.