pulse–echo principle
脉冲回波
Send a short pulse of sound, time the echo, and convert that time to depth using the speed of sound. All ultrasound imaging imaging rests on this: in soft tissue sound travels about 1540 m/s, so every 13 μs of delay means one more centimeter of depth.
Depth is , halved because the sound travels there and back. The scanner assumes 1540 m/s for all soft tissue and places each echo at the depth its arrival time implies. Echo strength along one beam is an A-mode line; many lines side by side, with strength shown as brightness, make a B-mode image.
The idea came from underwater ranging. Paul Langevin used ultrasound to find submarines in 1917, and sonar and radar located targets by echo throughout the Second World War. Physicians turned it on the body in the 1950s; in 1958 Ian Donald and colleagues described pulsed ultrasound of abdominal masses in The Lancet.
Each echo must return before the next pulse leaves. Imaging to 15 cm takes about 200 μs per line, so a 128-line frame allows at most about 40 frames per second; deeper means slower. The speed assumption can mislead: sound travels about 1450 m/s in fat, so echoes arrive late and structures are drawn too deep, and repeated bounces between interfaces create reverberation artifacts.
- Donald I, MacVicar J, Brown TG. Investigation of abdominal masses by pulsed ultrasound. Lancet. 1958;271(7032):1188–1195. doi.org/10.1016/S0140-6736(58)91905-6
- Medical ultrasound. Wikipedia. en.wikipedia.org/wiki/Medical_ultrasound
Key works · 10
Landmark1952Science216 citations
Application of Echo-Ranging Techniques to the Determination of Structure of Biological Tissues
The investigators used pulsed ultrasound to record echoes from biological tissues and moved the probe to assemble two-dimensional records from successive positions. They compared echo patterns from normal tissue and masses, showing that the reflections could reveal differences in soft tissue structure.
SignificanceThe paper was an early demonstration of using pulse-echo ranging to display soft tissue in two dimensions. The same investigators subsequently extended the approach to imaging lesions in the intact human breast, an early application of diagnostic ultrasound to a specific organ.
DOI Full text John J. WildJohn M. Reid ultrasound imagingpulse–echo principleacoustic impedanceB-mode
Landmark1952Journal of Laboratory and Clinical Medicine229 citations
Ultrasonic visualization of soft tissue structures of the body
The investigators recorded echoes from soft tissue interfaces and assembled the returns according to the probe position and direction to form cross-sectional pictures. They presented ultrasonic sections alongside the corresponding objects to show which boundaries and internal structures the method could depict.
SignificanceThe work established an early approach for assembling echoes from different directions into an anatomical cross section. Howry's group subsequently developed compound scanning arrangements that reduced the need to immerse the subject, making cross-sectional imaging more practical in patients.
Douglass H. HowryW. Roderic Bliss ultrasound imagingpulse–echo principleacoustic impedanceB-modetomography
Landmark1958The Lancet708 citations
Investigation of abdominal masses by pulsed ultrasound
The Glasgow team investigated abdominal masses using pulsed ultrasound and combined echoes obtained from different directions into sectional displays. Clinical cases illustrated differences between the return patterns of fluid-filled cysts and masses containing solid tissue, and the report also included pregnancy-related observations.
SignificanceThe report provided early clinical evidence for pulse-echo ultrasound in obstetrics and gynecology. The contact compound scanner and the diagnostic cases helped establish a program of abdominal, pelvic, and pregnancy imaging that the Glasgow team documented in subsequent clinical publications.
DOI Full text Ian DonaldJohn MacVicarTom Brown ultrasound imagingpulse–echo principleB-modeacoustic impedance
Landmark1968Investigative Radiology1,093 citations
Echocardiography of the Aortic Root
To identify the source of two moving echoes on an echocardiogram, the authors injected saline above the aortic valve while recording continuously. A transient cloud of echoes appeared between the two traces of the aortic root. Blood entering from the left ventricle during systole interrupted the cloud as the valve opened.
SignificanceThe enhanced echoes let the team use flowing blood to verify the anatomical origin of the traces. The group published further contrast experiments in 1969. A University of Rochester retrospective documents the subsequent research that identified bubbles as the source of enhanced echoes.
DOI Raymond GramiakPravin M. Shah ultrasound imagingpulse–echo principlecontrast agentM-mode ultrasound
Landmark1991Ultrasonic Imaging4,232 citations
Elastography: A Quantitative Method for Imaging the Elasticity of Biological Tissues
The authors recorded ultrasound echoes before and after a small compression and used shifts between the signals to estimate displacement at successive depths. Spatial changes in the displacements gave local strain estimates, producing an image that showed regions deforming differently from their surroundings.
SignificanceThe paper established a workflow for estimating local tissue strain from displacement recorded by ultrasound echoes. Later strain elastography used the same principle to display relative tissue deformation alongside grayscale anatomy, adding information on the mechanical response to compression.
DOI Jonathan OphirX. Li elastographypulse–echo principleultrasound imaging
Landmark2004Clinical Physiology and Functional Imaging321 citations
The Use of Ultrasonic Reflectoscope for the Continuous Recording of the Movements of Heart Walls.
Edler and Hertz first published the experiment in the Royal Physiographic Society of Lund’s proceedings in 1954. In the report, reprinted in 2004, they used a reflectoscope to measure echo depths along one beam through the chest wall and recorded the movement of cardiac structures over the cardiac cycle.
SignificanceContinuous pulse-echo recording of cardiac motion developed into motion-mode echocardiography. The work at Lund showed how valve motion could be examined through the chest and led to applications in the evaluation of mitral valve disease.
DOI Inge EdlerCarl Hellmuth Hertz ultrasound imagingtransducerpulse–echo principleM-mode ultrasound
Key1941United States Patent Office
Piezoelectric signaling apparatus
This patent, filed in 1920 and granted in 1941, described a quartz transducer for conversion between electrical oscillations and high-frequency acoustic waves in water. The specification treated transmission and reception as reciprocal functions of the piezoelectric apparatus for acoustic measurements.
SignificanceThe document records an early implementation of quartz piezoelectric transduction for underwater acoustics. Medical probes subsequently used the same conversion between electrical excitation, emitted sound, and returning echoes, a principle also developed for industrial inspection before clinical imaging.
Full text Paul Langevin piezoelectric effecttransducerpulse–echo principle
Key1942United States Patent Office
Flaw detecting device and measuring instrument
The patent described a pulsed ultrasonic instrument for inspecting the interior of materials. It sent short sound pulses into a part and displayed the returned signals on a synchronized oscilloscope. Their time of return gave the location of an interface or the thickness of the part.
SignificanceAn industrial instrument of this kind supplied hardware that early medical investigators could adapt. Teams investigating cardiac echoes in Lund and abdominal imaging in Glasgow both began with flaw detectors before developing equipment for clinical examinations.
Key1970IEEE Transactions on Sonics and Ultrasonics440 citations
Pulsed Ultrasonic Doppler Blood-Flow Sensing
Baker described an instrument for measuring blood flow with repeated ultrasound pulses. The receiver selected returning signals in a chosen time window to define a depth range and used Doppler information across the pulses to characterize flow in that region.
SignificanceCombining time gating with Doppler analysis made it possible to specify where along the beam flow was measured. Later color flow systems retained this localization and used pulsed Doppler to display motion within successive regions of a two-dimensional section.
DOI Donald W. Baker Doppler ultrasoundpulse–echo principleDoppler effect
Key1974IEEE Transactions on Biomedical Engineering244 citations
Ultrasonic Duplex Echo-Doppler Scanner
The authors combined multigate pulsed Doppler with a rotating B-mode scanner, using the same transducer to align flow and tissue images. They demonstrated the system on normal arteries in vivo. Trials in patients with verified occlusive disease led them to propose a two-transducer design.
SignificanceDuplex scanning let clinicians locate Doppler flow signals on an anatomical image with one instrument. The European Association of Cardiovascular Imaging identifies the scanner as a milestone after range-gated Doppler. The group later introduced a digital display superimposing flow information on B-mode images.
DOI Frank E. BarberDonald W. BakerJohn M. Reid B-modeDoppler ultrasoundpulse–echo principle