Doppler effect
多普勒效应
When a wave source and receiver move relative to each other, the received frequency shifts: higher when approaching, lower when receding. Ultrasound measures this shift in echoes from moving red blood cells to map the speed and direction of flow, the basis of Doppler ultrasound.
For an ultrasound echo the shift is , where θ is the angle between beam and flow. With a 5 MHz probe, blood at 50 cm/s and a 60° angle, the shift is about 1.6 kHz, squarely in the audible range, which is why a Doppler examination comes with the whooshing sound of flow.
Christian Doppler proposed the effect in 1842, in a paper on the colors of double stars. In 1845 Buys Ballot tested it for sound with musicians playing on a moving train. In the 1950s Shigeo Satomura in Japan applied it to ultrasound echoes from the moving heart and vessels.
The cos θ term is where measurements go wrong. With the beam perpendicular to flow there is no shift at all, and beyond about 60° small angle errors produce large velocity errors. When the shift exceeds half the pulse repetition frequency, fast flow wraps around and appears reversed: aliasing at the Nyquist limit, in ultrasound form.
- Doppler effect. Wikipedia. en.wikipedia.org/wiki/Doppler_effect
- Satomura S. Ultrasonic Doppler method for the inspection of cardiac functions. J Acoust Soc Am. 1957;29(11):1181–1185. doi.org/10.1121/1.1908737
Key works · 3
Landmark1957The Journal of the Acoustical Society of America290 citations
Ultrasonic Doppler Method for the Inspection of Cardiac Functions
Satomura directed continuous ultrasound through the chest and recorded Doppler shifts in waves reflected by moving cardiac structures. Simultaneous electrocardiographic and phonocardiographic recordings let him compare the timing of the signals and use them to characterize motions such as those of heart valves.
SignificanceThe study showed that ultrasound frequency shifts could provide information on cardiac motion in humans. Satomura subsequently applied Doppler to blood flow, and later pulsed and imaging systems developed the ability to localize these measurements within vessels or the heart.
Landmark1985IEEE Transactions on Sonics and Ultrasonics1,395 citations
Real-Time Two-Dimensional Blood Flow Imaging Using an Autocorrelation Technique
The authors combined pulsed Doppler with autocorrelation estimation to display blood flow in a cross section in real time. Color encoded direction and variations in flow. They tested mechanical and electronic scanners in phantoms and also examined normal and diseased hearts.
SignificanceAutocorrelation estimation made it practical to refresh a two-dimensional display while adding Doppler flow information. Color flow imaging subsequently became a means of locating abnormal flow in relation to anatomical structures during echocardiography and vascular ultrasound examinations.
DOI Chihiro KasaiRyozo Omoto Doppler ultrasoundDoppler effectB-mode
Key1970IEEE Transactions on Sonics and Ultrasonics440 citations
Pulsed Ultrasonic Doppler Blood-Flow Sensing
Baker described an instrument for measuring blood flow with repeated ultrasound pulses. The receiver selected returning signals in a chosen time window to define a depth range and used Doppler information across the pulses to characterize flow in that region.
SignificanceCombining time gating with Doppler analysis made it possible to specify where along the beam flow was measured. Later color flow systems retained this localization and used pulsed Doppler to display motion within successive regions of a two-dimensional section.
DOI Donald W. Baker Doppler ultrasoundpulse–echo principleDoppler effect