Modalities

A-mode

A型超声

The simplest ultrasound imaging display: echo strength plotted as spikes against depth along a single beam (A for amplitude). It makes no picture, but it measures distance precisely, and is still used to measure the length of the eye.

An A-mode trace is a single pulse–echo principle record. The horizontal axis is arrival time, converted to depth at about 13 μs per centimeter; the vertical axis is echo amplitude. Every boundary where acoustic impedance changes throws up a spike, and the gap between spikes is the distance between boundaries.

The method came straight from the industrial flaw detectors used to find cracks inside metal. In the 1950s neurologists used it to locate the midline echo inside the skull: a bleed or tumor on one side pushes the midline across, and this echoencephalography was an important emergency test before CT.

Placing many A-lines side by side and converting amplitude into brightness gives B-mode. Today A-mode survives mainly in ophthalmology, where the eye's axial length, about 24 mm in adults, is measured before cataract surgery to choose the power of the replacement lens; errors of a few tenths of a millimeter show up in the patient's vision.

Key works · 1

Key1942United States Patent Office

Flaw detecting device and measuring instrument

Floyd A. Firestone

The patent described a pulsed ultrasonic instrument for inspecting the interior of materials. It sent short sound pulses into a part and displayed the returned signals on a synchronized oscilloscope. Their time of return gave the location of an interface or the thickness of the part.

SignificanceAn industrial instrument of this kind supplied hardware that early medical investigators could adapt. Teams investigating cardiac echoes in Lund and abdominal imaging in Glasgow both began with flaw detectors before developing equipment for clinical examinations.