data consistencyDC
数据一致性
Data consistency makes a reconstructed image agree with the acquired measurements when passed through a model of the scanner. Agreement can be enforced exactly or weighted to account for noise.
In MRI reconstruction, a forward model predicts what the current image would yield at the acquired k-space locations. A data consistency step compares those values with the actual measurements and updates the image. In a simple single-coil setting, it can replace the predicted entries with acquired values. More general models can account for receive-coil sensitivities and the expected measurement error.
Enforcing exact agreement can carry measurement noise into the reconstructed image. When the scan is undersampled, many candidate images can agree with the recorded values. Among reconstructions with similar measurement residuals, additional constraints guide the estimate of unmeasured signals. A denoising prior, for example, may favor images with smoother local intensities.
- Aggarwal HK, Mani MP, Jacob M. MoDL: Model-Based Deep Learning Architecture for Inverse Problems. IEEE Trans Med Imaging, 2019. arxiv.org/abs/1712.02862
Key works · 10
Landmark2007Magnetic Resonance in Medicine7,800 citations
Sparse MRI: The application of compressed sensing for rapid MR imaging
The authors undersampled phase-encoding lines with a variable-density pseudorandom pattern and reconstructed images by penalizing transform coefficients subject to consistency with the measured data. Examples in multislice fast spin-echo brain images and three-dimensional contrast-enhanced angiography demonstrated how the method could accelerate acquisition.
SignificanceThe paper converted compressed sensing into an acquisition and reconstruction procedure that could be implemented on an MR scanner. Later studies combined sparse recovery and parallel imaging and evaluated the resulting shorter acquisitions in settings such as contrast-enhanced pediatric abdominal MRI.
DOI Full text Michael LustigDavid L. DonohoJohn M. Pauly compressed sensingk-spacesamplingaliasingphase encodingdata consistency
Landmark2018IEEE Transactions on Medical Imaging1,499 citations
A Deep Cascade of Convolutional Neural Networks for Dynamic MR Image Reconstruction
A cascade alternated convolutional corrections with enforcement of consistency with acquired k-space samples. For dynamic cardiac MRI, neighboring frames also provided information for joint reconstruction. In the authors' undersampling experiments the method better preserved anatomy than the reported comparators and reconstructed complete sequences in under ten seconds.
SignificanceThe cascade was an influential example of pairing learned corrections with repeated data consistency steps. It also demonstrated how temporal information could supplement undersampled measurements. Later experimental reviews of MRI reconstruction retained cascaded networks as a distinct comparison class.
DOI arXiv Jo SchlemperDaniel Rueckert magnetic resonance imagingk-spacedeep-learning reconstructiondeep unrollingdata consistency
Landmark2018IEEE Transactions on Medical Imaging1,136 citations
Learned Primal-Dual Reconstruction
The method unrolled a primal-dual optimization algorithm and replaced the proximal operations with convolutional networks while retaining the forward imaging operator. It started from measured projection data. Low-dose CT phantom experiments reported gains over filtered back-projection, total variation, and learned post-processing of a conventional reconstruction.
SignificanceThe method supplied a reference implementation of an end-to-end tomographic reconstructor with explicit forward and adjoint operations. It remains a reference for studying how learned updates interact with measured projections and for comparing unrolled reconstruction with image-domain correction.
DOI arXiv Jonas AdlerOzan Öktem iterative reconstructionsinograminverse problemdeep unrollingdata consistency
Landmark2018Magnetic Resonance in Medicine2,036 citations
Learning a variational network for reconstruction of accelerated MRI data
The network unrolled gradient updates of a variational model for multicoil MRI, learning its filters, activation functions, and weights for the measured-data terms. It was tested on retrospectively and prospectively undersampled knees. Quantitative comparisons and a reader study supported the results for fourfold regular undersampling.
SignificanceThe variational network made the parallel MRI acquisition model and the learned image prior explicit within a single trainable system. The later end-to-end variational network developed this design further by incorporating coil sensitivity estimation in the same reconstruction model.
DOI arXiv Full text Kerstin HammernikFlorian Knoll parallel imagingk-spacecompressed sensingdeep-learning reconstructiondeep unrollingdata consistency
Landmark2019IEEE Transactions on Medical Imaging1,490 citations
MoDL: Model-Based Deep Learning Architecture for Inverse Problems
MoDL alternated a learned convolutional regularizer with steps that fit the measured data, sharing the convolutional weights across iterations. A conjugate gradient block solved the data consistency subproblem. In undersampled MRI experiments, the solver improved convergence per iteration when available memory restricted the network depth.
SignificanceWeight sharing made it possible to change the number of reconstruction steps while holding the number of learned parameters fixed. The subsequent parallel MRI survey uses MoDL to illustrate the combination of an explicit forward model and a learned regularizer.
DOI arXiv Full text Hemant K. AggarwalMathews Jacob inverse problemiterative reconstructiondeep-learning reconstructiondeep unrollingdata consistency
Key2020IEEE Signal Processing Magazine514 citations
Deep-Learning Methods for Parallel Magnetic Resonance Imaging Reconstruction: A Survey of the Current Approaches, Trends, and Issues
The review explains linear and nonlinear parallel MRI reconstruction from image-domain and k-space perspectives. It describes ways to use neural networks as learned image regularizers or as interpolators for missing k-space measurements. It also discusses measurement consistency, public benchmarks, and performance across acquisition conditions.
SignificanceThis widely cited technical reference ties learning methods to the physics of multicoil MRI. It supplies a common vocabulary for comparing image enhancement, direct reconstruction and methods that explicitly enforce data consistency, and documents the early availability of benchmarks for testing those differences.
DOI arXiv Florian KnollMehmet Akçakaya parallel imagingk-spacedeep-learning reconstructiondata consistency
Key2020MICCAI 2020 (LNCS 12262)539 citations
End-to-End Variational Networks for Accelerated MRI Reconstruction
The end-to-end variational network estimated coil sensitivity maps and reconstructed undersampled multicoil MRI in a single trainable architecture. Cascaded refinement used learned image regularization and consistency with measured k-space. The paper reported better reconstruction scores than its comparators on fastMRI knee and brain data.
SignificanceThe architecture brought coil sensitivity estimation into the optimization of the reconstruction system. The fastMRI project distributed a Variational Network implementation, giving researchers an inspectable baseline for experiments on multicoil data and a reproducible reference when comparing alternative reconstruction stages.
DOI arXiv Anuroop SriramPatricia Johnson fastMRIparallel imagingdeep-learning reconstructiondeep unrollingdata consistency
Key2020Magnetic Resonance in Medicine436 citations
Self‐supervised learning of physics‐guided reconstruction neural networks without fully sampled reference data
SSDU split the acquired k-space samples into two disjoint sets. One supplied data consistency within a physics-guided network and the other defined the training loss. On fourfold accelerated knee MRI, results were close to fully supervised reconstruction. The authors also evaluated prospectively undersampled brain data.
SignificanceThe paper established a way to train physics-guided reconstruction from the kind of incomplete measurements used at deployment. The later multi-mask SSDU study directly extended its measurement partitioning scheme to make fuller use of samples available within each scan.
DOI arXiv Full text Burhaneddin YamanMehmet Akçakaya self-supervised learningk-spacefastMRIdata consistency
Key2022Medical Image Analysis671 citations
Score-based diffusion models for accelerated MRI
The method learned a score model from fully sampled MR images and combined diffusion sampling with k-space data consistency during reconstruction. The authors evaluated the same learned prior with single-coil and multicoil acquisitions and different undersampling schemes, reporting improvements over the reconstruction methods they compared.
SignificanceThe paper showed that a single independently learned image prior could be applied to several acquisition settings. Subsequent work on adaptive diffusion priors continued this line of investigation by examining how the learned prior could be adjusted at test time for individual accelerated MRI acquisitions.
DOI arXiv Hyungjin ChungJong Chul Ye diffusion modelk-spacedeep-learning reconstructiondata consistency
Key2022ICLR 2022865 citations
Solving Inverse Problems in Medical Imaging with Score-Based Generative Models
The researchers learned a generative score prior from medical images and introduced the measurements and their forward operator during sampling at test time. Tests included CT and MRI inverse problems. The study reported comparable or improved reconstruction in several tasks and stronger generalization to changed measurement processes.
SignificanceThe study used a common framework to test score priors across tomographic and MR measurement models. By varying the operator used at inference, it provided a direct evaluation of whether a learned prior could support multiple reconstruction tasks without acquisition-specific paired training.
arXiv Yang SongStefano Ermon inverse problemdiffusion modelcomputed tomographymagnetic resonance imagingdata consistency