Reconstruction

sinogram

正弦图

The raw data of a CT scan: every projection stacked row by row, one row per angle. Each point in the body traces a sine curve through it, hence the name; the cross-sectional image is reconstructed from it.

The first EMI scanner built its data the plain way. The X-ray beam traversed the head and recorded 160 transmission readings, one projection; the gantry turned 1° and repeated, 180 times, for 28,800 numbers in all. Stacking those projections row by row, with detector position across and angle down, gives a sinogram.

A small object at lands at in each view, so as the angle turns it traces a sine curve; the further it sits from the center of rotation, the wider the swing. The whole sinogram is the sum of such curves, which is to say the object's Radon transform.

The sinogram is also where faults show up. On a modern third-generation scanner a dead detector element draws a straight vertical line through it, which reconstruction turns into a ring artifact; a patient who moves mid-scan makes every curve jump at the same row.

Sources
  1. Hounsfield GN. Computerized transverse axial scanning (tomography): Part 1. Description of system. Br J Radiol. 1973;46(552):1016–1022. doi.org/10.1259/0007-1285-46-552-1016
  2. Goldman LW. Principles of CT and CT technology. J Nucl Med Technol. 2007;35(3):115–128. tech.snmjournals.org/content/35/3/115

Key works · 3

Landmark2018IEEE Transactions on Medical Imaging1,136 citations

Learned Primal-Dual Reconstruction

Jonas Adler, Ozan Öktem

The method unrolled a primal-dual optimization algorithm and replaced the proximal operations with convolutional networks while retaining the forward imaging operator. It started from measured projection data. Low-dose CT phantom experiments reported gains over filtered back-projection, total variation, and learned post-processing of a conventional reconstruction.

SignificanceThe method supplied a reference implementation of an end-to-end tomographic reconstructor with explicit forward and adjoint operations. It remains a reference for studying how learned updates interact with measured projections and for comparing unrolled reconstruction with image-domain correction.

Key2019CVPR 2019353 citations

DuDoNet: Dual Domain Network for CT Metal Artifact Reduction

Wei-An Lin, Haofu Liao … Shaohua Kevin Zhou

DuDoNet corrected metal-corrupted sinograms and refined the resulting CT images within one trainable network. A differentiable Radon inversion layer connected the two stages. In the reported experiments, joint processing improved detail recovery and reduced residual and secondary artifacts relative to single-domain approaches.

SignificanceDuDoNet provided a concrete dual-domain baseline for learned metal artifact reduction. Later work on deep sinogram completion with an image prior compared against it when studying how projection continuity at the metal trace boundary affects artifacts after reconstruction.

Notable2021IEEE Transactions on Medical Imaging184 citations

Deep Sinogram Completion With Image Prior for Metal Artifact Reduction in CT Images

Lequan Yu, Zhicheng Zhang … Lei Xing

The method produced an image prior and projected it forward differentiably to guide a second network that repaired metal-corrupted sinogram values. Filtered back-projection yielded the final image. Experiments with simulated and real artifacts examined projection continuity around metal traces and reported reduced artifacts relative to the comparison methods.

SignificanceThe paper showed how an estimated image could guide a projection-space correction and how that correction could be tested on both simulated and real artifacts. The differentiable forward operator allowed the prior-image and sinogram networks to be optimized jointly.