spin echoSE

自旋回波

A spin echo is a renewed magnetic resonance signal produced by refocusing spins whose phases spread because of static differences in magnetic field. It appears at a predictable time after excitation.

After a radiofrequency pulse excites the spins, locations with a slightly stronger magnetic field precess faster. Their phases spread apart. In a conventional spin-echo sequence, a 180° refocusing pulse reverses the phase ordering of the transverse magnetization. The local precession rates remain the same. The spins that had been ahead are now behind, so they return to the same phase after an interval equal to the delay between excitation and refocusing and the measured signal rises.

Refocusing compensates for phase spread caused by static field differences. The echo amplitude decays through T2 relaxation. A longer delay before the echo leaves less signal from the same tissue and can change how strongly differences in relaxation affect image contrast.

Sources
  1. Hahn EL. Spin Echoes. Phys Rev, 1950. journals.aps.org/pr/abstract/10.1103/PhysRev.80.580
  2. Plewes DB, Kucharczyk W. Physics of MRI: A Primer. J Magn Reson Imaging, 2012. doi.org/10.1002/jmri.23642

Key works · 5

Landmark1950Physical Review7,007 citations

Spin Echoes

E. L. Hahn

Hahn triggered nuclear magnetic resonance in liquids with sequences of radio-frequency pulses. Magnetization that had dephased produced observable echoes at times set by the pulse intervals. Hahn explained this refocusing, measured relaxation from echo amplitudes, and examined how molecular self-diffusion affected those measurements.

SignificanceSpin echoes supplied a way to refocus dephasing caused by static field inhomogeneity and made tissue-dependent transverse relaxation useful for image contrast. Fast spin-echo imaging subsequently used trains of refocused echoes to collect more image data per excitation.

Landmark1965The Journal of Chemical Physics9,555 citations

Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field Gradient

E. O. Stejskal, J. E. Tanner

The authors derived how molecular diffusion attenuates spin echoes in a time-dependent field gradient and verified the expression with several gradient waveforms. With pulsed gradients they could define the interval over which diffusion was probed, and they measured the self-diffusion coefficient of dry glycerol.

SignificanceThe work linked a controlled gradient waveform to measurable attenuation caused by random molecular motion. Diffusion MRI later applied this measurement voxel by voxel, and diffusion tensor imaging extended the model to measurements across gradient directions to estimate anisotropic diffusion.

Landmark1986Radiology4,375 citations

MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders

D. Le Bihan, E. Breton … M. Laval-Jeantet

The authors used gradient pulses to measure spin-echo attenuation associated with diffusion and microcirculation within individual voxels. Measurements in water and acetone phantoms reproduced known diffusion values, and estimates of apparent diffusion in neurologic images differed between several normal and pathologic tissues.

SignificanceThe work established a way to map apparent diffusion in human tissue and introduced intravoxel incoherent motion as an imaging measurement. Subsequent tensor imaging used diffusion encoding in multiple directions to characterize anisotropy as well as overall diffusion magnitude.

Landmark1986Magnetic Resonance in Medicine2,345 citations

RARE imaging: A fast imaging method for clinical MR

Jürgen Hennig, Arno Nauerth, Hartmut Friedburg

The team assigned a different phase-encoding step to each of several spin echoes following excitation, collecting enough data for a 256 by 256 image in 2 to 40 seconds. They examined artifacts caused by changing T2 weighting across the echo train and how phase-encoding order could reduce them.

SignificanceAcquiring multiple k-space lines from a train of refocused echoes became the basis of clinical fast or turbo spin-echo imaging. Later implementations retained this approach to accelerating T2-weighted acquisitions while addressing the blurring and radio-frequency demands of long echo trains.

Key1971Science1,911 citations

Tumor Detection by Nuclear Magnetic Resonance

Raymond Damadian

Damadian used spin-echo measurements to compare longitudinal and transverse relaxation in excised samples of six normal rat tissues and two malignant tumors. The two tumor types had relaxation values outside the range of the normal tissues tested, suggesting that the measurements might help distinguish tissue specimens.

SignificanceDifferences among the excised rat tissues gave an early reason to investigate relaxation as a source of contrast. In his Nobel lecture, Lauterbur recounted how Damadian’s result prompted follow-up experiments and how learning about that research preceded his idea of locating signals with a magnetic field gradient.