nuclear spin
核自旋
The intrinsic angular momentum of certain nuclei, such as the hydrogen proton, which makes each act like a tiny bar magnet. In a strong field slightly more of the body's hydrogen nuclei align with the field than against it, and that tiny net magnetization is the signal MRI measures.
Only nuclei with an odd number of protons or neutrons have net spin: ¹H, ¹³C, ¹⁹F and ³¹P do, ¹²C and ¹⁶O do not. A proton has spin ½ and in a magnetic field occupies one of just two energy states, aligned or anti-aligned. About six in ten atoms in the body are hydrogen, mostly in water and fat, so clinical MRI is almost entirely hydrogen imaging.
Wolfgang Pauli proposed in 1924 that nuclei carry spin and a magnetic moment, to explain the hyperfine structure of atomic spectra. In the 1930s Otto Stern measured the proton's magnetic moment and Isidor Rabi first observed nuclear magnetic resonance in molecular beams.
The energy difference between the two states is minute. At 1.5 T and body temperature, aligned protons outnumber anti-aligned ones by only about five in a million. MRI signal is therefore intrinsically weak and is compensated for only by the sheer number of protons, which is why stronger fields, measured in tesla, and repeated averaging are the basic ways to buy signal-to-noise ratio.
- Nuclear magnetic resonance. Wikipedia. en.wikipedia.org/wiki/Nuclear_magnetic_resonance
Key works · 2
Landmark1946Physical Review1,657 citations
Nuclear Induction
The team applied a radio-frequency field to water in a static magnetic field and measured the induced voltage with a pickup coil oriented perpendicular to the excitation coil. At resonance, the nuclear magnetization of the water produced a detectable signal, demonstrating inductive reception in a liquid sample.
SignificanceInductive pickup became a basic means of receiving nuclear magnetic resonance signals in spectroscopy and imaging. This liquid experiment and the independent solid-state work at Harvard established complementary measurement methods for which Bloch and Purcell shared the 1952 Nobel Prize in Physics.
DOI Felix BlochMartin E. Packard nuclear magnetic resonancenuclear spinLarmor frequency
Landmark1946Physical Review7,634 citations
Resonance Absorption by Nuclear Magnetic Moments in a Solid
The authors placed paraffin in a radio-frequency resonant cavity within a static magnetic field and monitored absorption as they varied the field. They detected an absorption signal when the proton precession frequency matched the applied frequency, demonstrating nuclear magnetic resonance in a solid sample.
SignificanceTogether with the independent liquid experiment by Bloch's group, the work established methods for measuring nuclear magnetic resonance in condensed matter. Purcell and Bloch shared the 1952 Nobel Prize in Physics for nuclear magnetic precision measurement methods and associated discoveries.
DOI Edward M. PurcellRobert V. Pound nuclear magnetic resonancenuclear spinLarmor frequency