The SI unit of magnetic flux density, named after Nikola Tesla. Earth's field is about 50 μT; clinical MRI scanners run at 1.5 or 3 T, tens of thousands of times stronger. Higher fields give more signal, along with more artifacts, heating and cost.
One tesla is 10,000 gauss. For scale: Earth's surface field is 25–65 μT, a fridge magnet about 5 mT, clinical MRI 1.5 or 3 T, research ultra-high-field systems 7 T and above. The superconducting coils sit in liquid helium at about −269 °C, where current, once established, keeps circulating almost without loss; an MRI magnet is always on.
The unit was named at the 1960 General Conference on Weights and Measures, the meeting that established the SI, after the inventor Nikola Tesla (1856–1943).
Doubling the field roughly doubles proton polarization and signal, which can be spent on finer resolution or faster scans. But RF heating roughly quadruples, the higher Larmor frequency makes the RF field uneven inside the body, and susceptibility artifacts grow. The fringe field can turn ferromagnetic objects into projectiles, and a 0.5 mT (5 gauss) line around the magnet marks the safe limit for pacemakers and similar devices.
- Tesla (unit). Wikipedia. en.wikipedia.org/wiki/Tesla_(unit)
Key works · 5
Landmark1992Proceedings of the National Academy of Sciences4,224 citations
Intrinsic signal changes accompanying sensory stimulation: functional brain mapping with magnetic resonance imaging
At 4 tesla the authors used gradient-echo imaging to detect transient signal increases in primary visual cortex during visual stimulation. The fractional response fell when the echo time was shortened, consistent with a change in apparent transverse relaxation associated with blood oxygenation.
SignificanceThe study extended BOLD contrast from animal experiments to human cortical mapping and used echo-time dependence to test the proposed mechanism. It was one of three contemporary demonstrations that helped establish endogenous hemodynamic contrast for task-based functional imaging.
DOI Full text Seiji OgawaDavid W. TankKâmil Uğurbil functional MRIBOLD signaltesla
Key2001Magnetic Resonance in Medicine980 citations
7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images
The authors compared human head imaging at 7 and 4 tesla using measurements and electromagnetic modeling. At 7 tesla, average brain signal-to-noise was at least 1.6 times the value at 4 tesla, while the radio-frequency field was less uniform and excitation at the center of the head required more power.
SignificanceThe controlled comparison quantified both the signal gain and the radio-frequency engineering demands of ultrahigh field, providing a reference for subsequent coil and imaging work. The US Food and Drug Administration cleared the first clinical 7 tesla device in 2017.
DOI J. Thomas VaughanKâmil Uğurbil ultra-high-field MRIteslasignal-to-noise ratioRF pulse
Key2021JAMA Neurology311 citations
Assessment of Brain Injury Using Portable, Low-Field Magnetic Resonance Imaging at the Bedside of Critically Ill Patients
In this single-center prospective study, the team performed bedside brain MRI for 50 intensive care patients using a portable 0.064 tesla scanner. They obtained several image contrasts, documented abnormalities and the numbers of diagnostic-grade examinations by sequence, and reported no adverse events during deployment or scanning.
SignificanceThe study established feasibility for acquiring brain MRI at the bedside in intensive care. Follow-up work evaluated measurements such as intracranial midline shift, and the American College of Radiology subsequently addressed the use of portable ultralow-field systems in its safety guidance.
DOI Full text Kevin N. ShethW. Taylor Kimberly low-field MRImagnetic resonance imagingtesla
Key2025Radiology65 citations
American College of Radiology Manual on MR Safety: 2024 Update and Revisions
Members of the American College of Radiology committee described the structure and rationale of the 2024 safety manual. They reviewed facility zones, staff roles, patient and equipment screening, and emergency procedures, as well as evidence gaps for remote operation and scanners outside conventional field strengths.
SignificanceThe article documents the technical and clinical reasoning behind the American College of Radiology's consolidated MR safety recommendations. The manual covers institutional controls for risks from static fields, switched gradients, and radio-frequency exposure and addresses portable and ultrahigh-field systems.
DOI Ivan PedrosaRobert E. Watson magnetic resonance imagingultra-high-field MRIlow-field MRItesla
Notable2003Investigative Radiology247 citations
Magnetic Resonance Imaging at 3.0 Tesla: Challenges and Advantages in Clinical Neurological Imaging
The Calgary group summarized 40 months of clinical neuroradiology at 3 tesla and presented patient images to illustrate potential benefits relative to 1.5 tesla. They described higher signal-to-noise and changes in contrast alongside practical issues involving radio-frequency power, coil design, and implant compatibility.
SignificanceThis report documented early clinical operation at 3 tesla and made the practical demands of higher-field imaging explicit alongside its benefits. Later guidance from the American College of Radiology continues to address risks by field strength and device conditions.
DOI Richard Frayne magnetic resonance imagingteslasignal-to-noise ratioartifact