Modalities

low-field MRI

低场磁共振

MRI at fields well below the clinical 1.5 T, down to a few tens of millitesla. Signal is weaker and images coarser, but the magnet can be cheap, light and permanent, so the scanner can be wheeled to the bedside or installed where no conventional MRI exists.

MRI signal comes from the tiny net magnetisation of nuclear spins, which grows with field strength, and the voltage it induces in the coil grows with frequency too, so signal-to-noise ratio falls steeply as the field drops. Low-field scanners compensate with larger voxels, longer acquisitions and heavier processing; deep-learning reconstruction has proved especially useful here.

In 2020 Hyperfine's Swoop became the first bedside MRI system cleared by the US FDA. Its permanent magnet runs at 64 mT, less than a twentieth of 1.5 T; it plugs into an ordinary wall socket and can be rolled into an intensive-care unit to scan the head of a patient too sick to move.

Low field has advantages of its own: fewer metal artifacts, less radio-frequency heating, no liquid helium, and a fraction of the cost and siting demands. For the many hospitals in the world without any MRI, a coarser image is far more useful than no MRI at all.

Sources
  1. Hyperfine. Regulatory clearances (FDA K192002, February 2020; K201722, August 2020). www.hyperfinemri.com/hyperfine-regulatory-clearances

Key works · 2

Key2021JAMA Neurology311 citations

Assessment of Brain Injury Using Portable, Low-Field Magnetic Resonance Imaging at the Bedside of Critically Ill Patients

Kevin N. Sheth, Mercy H. Mazurek … Adrienne Ward

In this single-center prospective study, the team performed bedside brain MRI for 50 intensive care patients using a portable 0.064 tesla scanner. They obtained several image contrasts, documented abnormalities and the numbers of diagnostic-grade examinations by sequence, and reported no adverse events during deployment or scanning.

SignificanceThe study established feasibility for acquiring brain MRI at the bedside in intensive care. Follow-up work evaluated measurements such as intracranial midline shift, and the American College of Radiology subsequently addressed the use of portable ultralow-field systems in its safety guidance.

Key2025Radiology65 citations

American College of Radiology Manual on MR Safety: 2024 Update and Revisions

Ivan Pedrosa, David A. Altman … Scott B. Reeder

Members of the American College of Radiology committee described the structure and rationale of the 2024 safety manual. They reviewed facility zones, staff roles, patient and equipment screening, and emergency procedures, as well as evidence gaps for remote operation and scanners outside conventional field strengths.

SignificanceThe article documents the technical and clinical reasoning behind the American College of Radiology's consolidated MR safety recommendations. The manual covers institutional controls for risks from static fields, switched gradients, and radio-frequency exposure and addresses portable and ultrahigh-field systems.