J. Thomas Vaughan

Led a comparison of radio-frequency fields and signal-to-noise in human imaging at 7 and 4 tesla.

Magnetic resonance imaging ORCID OpenAlex

University of Minnesota (2001)

Key works · 1

Key2001Magnetic Resonance in Medicine980 citations

7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images

J. Thomas Vaughan, Michael Garwood … Gregor Adriany

The authors compared human head imaging at 7 and 4 tesla using measurements and electromagnetic modeling. At 7 tesla, average brain signal-to-noise was at least 1.6 times the value at 4 tesla, while the radio-frequency field was less uniform and excitation at the center of the head required more power.

SignificanceThe controlled comparison quantified both the signal gain and the radio-frequency engineering demands of ultrahigh field, providing a reference for subsequent coil and imaging work. The US Food and Drug Administration cleared the first clinical 7 tesla device in 2017.